Estimation of Rectal Dose Using Daily Megavoltage Cone-Beam Computed Tomography and Deformable Image Registration

被引:17
作者
Akino, Yuichi [1 ,2 ]
Yoshioka, Yasuo [1 ]
Fukuda, Shoichi [4 ]
Maruoka, Shintaroh [1 ]
Takahashi, Yutaka [1 ,5 ]
Yagi, Masashi [1 ]
Mizuno, Hirokazu [2 ]
Isohashi, Fumiaki [3 ]
Ogawa, Kazuhiko [1 ]
机构
[1] Osaka Univ, Grad Sch Med, Dept Radiat Oncol, Suita, Osaka 5650871, Japan
[2] Osaka Univ Hosp, Dept Radiol, Suita, Osaka, Japan
[3] Osaka Univ Hosp, Ctr Oncol, Suita, Osaka, Japan
[4] Osaka Gen Med Ctr, Dept Radiat Oncol, Osaka, Japan
[5] Univ Minnesota, Dept Radiat Oncol, Minneapolis, MN USA
来源
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS | 2013年 / 87卷 / 03期
基金
日本学术振兴会;
关键词
FREE-FORM DEFORMATIONS; PROSTATE-CANCER; RADIATION-THERAPY; RADIOTHERAPY; QUALITY; CT; ACCUMULATION; VALIDATION;
D O I
10.1016/j.ijrobp.2013.06.2054
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: The actual dose delivered to critical organs will differ from the simulated dose because of interfractional organ motion and deformation. Here, we developed a method to estimate the rectal dose in prostate intensity modulated radiation therapy with consideration to interfractional organ motion using daily megavoltage cone-beam computed tomography (MVCBCT). Methods and Materials: Under exemption status from our institutional review board, we retrospectively reviewed 231 series of MVCBCT of 8 patients with prostate cancer. On both planning CT (pCT) and MVCBCT images, the rectal contours were delineated and the CT value within the contours was replaced by the mean CT value within the pelvis, with the addition of 100 Hounsfield units. MVCBCT images were rigidly registered to pCT and then nonrigidly registered using B-Spline deformable image registration (DIR) with Velocity AI software. The concordance between the rectal contours on MVCBCTand pCT was evaluated using the Dice similarity coefficient (DSC). The dose distributions normalized for 1 fraction were also deformed and summed to estimate the actual total dose. Results: The DSC of all treatment fractions of 8 patients was improved from 0.75 +/- 0.04(mean +/- SD) to 0.90 +/- 0.02 by DIR. Six patients showed a decrease of the generalized equivalent uniform dose (gEUD) from total dose compared with treatment plans. Although the rectal volume of each treatment fraction did not show any correlation with the change in gEUD(R-2=0.18 +/- 0.13), the displacement of the center of gravity of rectal contours in the anterior-posterior (AP) direction showed an intermediate relationship (R-2=0.61 +/- 0.16). Conclusion: We developed a method for evaluation of rectal dose using DIR and MVCBCT images and showed the necessity of DIR for the evaluation of total dose. Displacement of the rectum in the AP direction showed a greater effect on the change in rectal dose compared with the rectal volume. (C) 2013 Elsevier Inc.
引用
收藏
页码:602 / 608
页数:7
相关论文
共 28 条
[1]   An on-line replanning method for head and neck adaptive radiotherapy [J].
Ahunbay, Ergun E. ;
Peng, Cheng ;
Godley, Andrew ;
Schultz, Christopher ;
Li, X. Allen .
MEDICAL PHYSICS, 2009, 36 (10) :4776-4790
[2]   MEGAVOLTAGE CONE BEAM COMPUTED TOMOGRAPHY DOSE AND THE NECESSITY OF REOPTIMIZATION FOR IMAGING DOSE-INTEGRATED INTENSITY-MODULATED RADIOTHERAPY FOR PROSTATE CANCER [J].
Akino, Yuichi ;
Koizumi, Masahiko ;
Sumida, Iori ;
Takahashi, Yutaka ;
Ogata, Toshiyuki ;
Ota, Seiichi ;
Isohashi, Fumiaki ;
Konishi, Koji ;
Yoshioka, Yasuo .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2012, 82 (05) :1715-1722
[3]   On the dosimetric effect and reduction of inverse consistency and transitivity errors in deformable image registration for dose accumulation [J].
Bender, Edward T. ;
Hardcastle, Nicholas ;
Tome, Wolfgang A. .
MEDICAL PHYSICS, 2012, 39 (01) :272-280
[4]   RESULTS OF A MULTI-INSTITUTION DEFORMABLE REGISTRATION ACCURACY STUDY (MIDRAS) [J].
Brock, Kristy K. .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2010, 76 (02) :583-596
[5]   FITTING OF NORMAL TISSUE TOLERANCE DATA TO AN ANALYTIC-FUNCTION [J].
BURMAN, C ;
KUTCHER, GJ ;
EMAMI, B ;
GOITEIN, M .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1991, 21 (01) :123-135
[6]   Accumulating daily-varied dose distributions of prostate radiation therapy with soft-tissue-based kV CT guidance [J].
Godley, Andrew ;
Ahunbay, Ergun ;
Peng, Cheng ;
Li, X. Allen .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2012, 13 (03) :98-107
[7]   Automated registration of large deformations for adaptive radiation therapy of prostate cancer [J].
Godley, Andrew ;
Ahunbay, Ergun ;
Peng, Cheng ;
Li, X. Allen .
MEDICAL PHYSICS, 2009, 36 (04) :1433-1441
[8]   Late rectal bleeding after conformal radiotherapy of prostate cancer (II): Volume effects and dose-volume histograms [J].
Jackson, A ;
Skwarchuk, MW ;
Zelefsky, MJ ;
Cowen, DM ;
Venkatraman, ES ;
Levegrun, S ;
Burman, CM ;
Kutcher, GJ ;
Fuks, Z ;
Liebel, SA ;
Ling, CC .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2001, 49 (03) :685-698
[9]   Evaluation of nonrigid registration models for interfraction dose accumulation in radiotherapy [J].
Janssens, Guillaume ;
de Xivry, Jonathan Orban ;
Fekkes, Stein ;
Dekker, Andre ;
Macq, Benoit ;
Lambin, Philippe ;
van Elmpt, Wouter .
MEDICAL PHYSICS, 2009, 36 (09) :4268-4276
[10]   Dosimetric and deformation effects of image-guided interventions during stereotactic body radiation therapy of the prostate using an endorectal balloon [J].
Jones, Bernard L. ;
Gan, Gregory ;
Diot, Quentin ;
Kavanagh, Brian ;
Timmerman, Robert D. ;
Miften, Moyed .
MEDICAL PHYSICS, 2012, 39 (06) :3080-3088