A Comparison between Scheimpflug Imaging and Optical Coherence Tomography in Measuring Corneal Thickness

被引:78
作者
Huang, Jinhai [1 ,2 ,3 ]
Ding, Xixia [1 ,2 ,3 ]
Savini, Giacomo [4 ]
Pan, Chao [1 ,2 ]
Feng, Yifan [1 ,2 ]
Cheng, Dan [1 ,2 ]
Hua, Yanjun [1 ,2 ]
Hu, Xiuli [1 ,2 ]
Wang, Qinmei [1 ,2 ,3 ]
机构
[1] Wenzhou Med Univ, Sch Optometry & Ophthalmol, Wenzhou 325027, Zhejiang, Peoples R China
[2] Wenzhou Med Univ, Hosp Eye, Wenzhou 325027, Zhejiang, Peoples R China
[3] Minist Hlth PR China, Key Lab Vis Sci, Wenzhou, Zhejiang, Peoples R China
[4] GB Bietti Fdn IRCCS, Rome, Italy
关键词
IN-SITU KERATOMILEUSIS; ULTRASOUND PACHYMETRY; AUTOMATIC MEASUREMENTS; SPECULAR MICROSCOPE; REFRACTIVE SURGERY; ORBSCAN PACHYMETRY; EXTENDED WEAR; CONTACT-LENS; REPEATABILITY; CAMERA;
D O I
10.1016/j.ophtha.2013.02.022
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose: To assess the repeatability and reproducibility of 3 rotating Scheimpflug cameras, the Pentacam (Oculus, Wetzlar, Germany), Sirius (Costruzione Strumenti Oftalmici, Florence, Italy), and Galilei (Ziemer, Biel, Switzerland), and 1 Fourier-domain optical coherence tomography (FD-OCT) system, the RTvue-100 OCT (Optovue Inc., Fremont, CA), in measuring corneal thickness. Design: Evaluation of diagnostic test. Participants: Sixty-six right eyes of 66 healthy volunteers, whose mean age +/- standard deviation (SD) was 35.39 +/- 10.06 years (range, 18-55 years). Methods: Corneal thickness measurements obtained by each system included central corneal thickness (CCT), thinnest corneal thickness (TCT), and midperipheral corneal thickness (MPCT), measured at superior, inferior, nasal, and temporal locations at a distance of 1 and 2.5 mm from the corneal apex. In the first session, 3 consecutive measurements were performed by the same operator to assess intraobserver repeatability and by a second operator to assess interobserver reproducibility. Measurements were repeated in the second session scheduled 1 day to 1 week later. The mean values obtained in the 2 sessions by the first operator were used to investigate the intersession reproducibility. Main Outcome Measures: Intraobserver repeatability and interobserver and intersession repeatability of corneal thickness measurements, as calculated by means of within-subject SD, testeretest repeatability, coefficient of variation (COV), and intraclass correlation coefficients. Results: The precision of CCT, TCT, CT2 mm (midperipheral corneal thickness [MPCT] with a distance of 1 mm from the corneal apex), and CT5 mm (midperipheral corneal thickness [MPCT] with a distance of 2.5 mm from the corneal apex) measurements was high with all 4 systems. The COV was <= 1.16%, 0.94%, and 1.10% for repeatability, interobserver reproducibility, and intersession reproducibility, respectively. The 4 devices offered better interobserver reproducibility than intersession reproducibility for all measurements. The CTsuperior-5 mm (midperipheral corneal thickness [MPCT], measured at superior locations with a distance of 2.5 mm from the corneal apex) measurements showed the poorest repeatability and reproducibility. The Galilei revealed the best precision of CCT, TCT, and CT2 mm measurements. Conclusions: Both Scheimpflug imaging and FD-OCT offer highly repeatable and reproducible measurements of CCT and MPCT. The precision was lower in the midperipheral superior quadrant.
引用
收藏
页码:1951 / 1958
页数:8
相关论文
共 43 条
[1]   Central corneal thickness measurement with the Pentacam Scheimpflug system, optical low-coherence reflectometry pachymeter, and ultrasound pachymetry [J].
Barkana, Y ;
Gerber, Y ;
Elbaz, U ;
Schwartz, S ;
Ken-Dror, G ;
Avni, I ;
Zadok, D .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2005, 31 (09) :1729-1735
[2]   Central and Peripheral Corneal Thinning in Keratoconus [J].
Brautaset, Rune L. ;
Nilsson, Maria ;
Miller, William L. ;
Leach, Norman E. ;
Tukler, Johanna H. ;
Bergmanson, Jan P. G. .
CORNEA, 2013, 32 (03) :257-261
[3]   Measurement of central corneal thickness by high-resolution Scheimpflug imaging, Fourier-domain optical coherence tomography and ultrasound pachymetry [J].
Chen, Shihao ;
Huang, Jinhai ;
Wen, Daizong ;
Chen, Wuhe ;
Huang, David ;
Wang, Qinmei .
ACTA OPHTHALMOLOGICA, 2012, 90 (05) :449-455
[4]   Central and peripheral corneal thickness measured with the TOPCON specular microscope SP-2000P [J].
Cho, P ;
Cheung, SW .
CURRENT EYE RESEARCH, 2000, 21 (04) :799-807
[5]   Comparison of central corneal thickness measurements by ultrasound pachymetry, Konan noncontact optical pachymetry, and Orbscan pachymetry [J].
Christensen, Amarilda ;
Narvaez, Julio ;
Zimmerman, Grenith .
CORNEA, 2008, 27 (08) :862-865
[6]   Reproductivity and repeatability of central corneal thickness measurement in keratoconus using the rotating scheimpflug camera and ultrasound pachymetry [J].
De Sanctis, Ugo ;
Missolungi, Alessandro ;
Mutani, Bernardo ;
Richiardi, Lorenzo ;
Grignolo, Federico M. .
AMERICAN JOURNAL OF OPHTHALMOLOGY, 2007, 144 (05) :712-718
[7]  
Gordon MO, 2002, ARCH OPHTHALMOL-CHIC, V120, P714
[8]   Reliability of RTVue, Visante, and Slit-Lamp Adapted Ultrasonic Pachymetry for Central Corneal Thickness Measurement [J].
Hong, Jin Pyo ;
Nam, Sang Min ;
Kim, Tae-Im ;
Seo, Kyoung Yul ;
Lee, Sang Yeop ;
Meduri, Alessandro ;
Kim, Eung Kweon .
YONSEI MEDICAL JOURNAL, 2012, 53 (03) :634-641
[9]   Intra-examiner repeatability and agreement of corneal pachymetry map measurement by time-domain and Fourier-domain optical coherence tomography [J].
Huang, Jehn-Yu ;
Pekmezci, Melike ;
Yaplee, Stephanie ;
Lin, Shan .
GRAEFES ARCHIVE FOR CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2010, 248 (11) :1647-1656
[10]   A Comprehensive Comparison of Central Corneal Thickness Measurement [J].
Huang, Jinhai ;
Pesudovs, Konrad ;
Yu, Ayong ;
Wright, Thomas ;
Wen, Daizong ;
Li, Ming ;
Yu, Ye ;
Wang, Qinmei .
OPTOMETRY AND VISION SCIENCE, 2011, 88 (08) :940-949