Quantitative Assessment of Induced Errors In Attenuation Corrected Myocardial SPECT Images Due to Misregistration

被引:0
作者
Babaeifar, Hamidreza [1 ]
Rajabi, Hossein [1 ]
Sen, Anando [2 ]
Kalantari, Faraz [2 ]
机构
[1] Tabirat Modares Univ, Dept Med Phys, Tehran, Iran
[2] Univ Houston, Dept Biomed Engn, Houston, TX USA
来源
2014 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC) | 2014年
关键词
SPECT; SPECT-CT; Attenuation correction; Misregistration; Myocardial perfusion; PERFUSION SPECT; TOMOGRAPHY/COMPUTED TOMOGRAPHY; EMISSION-TOMOGRAPHY; MOTION; RECONSTRUCTION; REREGISTRATION; SIMULATION; ARTIFACTS; SCATTER; IMPACT;
D O I
暂无
中图分类号
TL [原子能技术]; O571 [原子核物理学];
学科分类号
0827 ; 082701 ;
摘要
Accurate attenuation correction can significantly improve the diagnostic accuracy of SPECT images. However, misregistration between the emission and attenuation maps can cause alignment induced artifacts in the attenuation corrected images. In this study we present a quantitative assessment of such errors in myocardial perfusion SPECT. A family of XCAT phantoms were imaged using the SIMIND Monte Carlo projector. To study the effects of misregistration, patient motion and respiratory motion were simulated. The differences between the test image and a reference image (perfectly aligned attenuation corrected image) were analyzed. The types of errors considered were lesion location errors, pixel density errors and change in contrast. Results indicate that horizontal translation causes the highest quantitative error and lesions in the lateral wall are most sensitive to misregistration induced errors.
引用
收藏
页数:6
相关论文
共 25 条
[1]  
Azencott R., 2011, P 5 INT ICST C PERF, P602
[2]  
Blankespoor S., 1996, NUCL SCI IEEE T, P2263
[3]  
BOTVINICK EH, 1993, J NUCL MED, V34, P303
[4]   Correction of the respiratory motion of the heart by tracking of the center of mass of thresholded projections: A simulation study using the dynamic MCAT phantom [J].
Bruyant, PP ;
King, MA ;
Pretorius, PH .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2002, 49 (05) :2159-2166
[5]  
Fricke H, 2004, J NUCL MED, V45, P1619
[6]   Attenuation correction in myocardial perfusion SPECT/CT: Effects of misregistration and value of reregistration [J].
Goetze, Sibyll ;
Brown, Tracy L. ;
Lavely, William C. ;
Zhang, Zhe ;
Bengel, Frank M. .
JOURNAL OF NUCLEAR MEDICINE, 2007, 48 (07) :1090-1095
[7]   Multicenter clinical trial to evaluate the efficacy of correction for photon attenuation and scatter in SPECT myocardial perfusion imaging [J].
Hendel, RC ;
Berman, DS ;
Cullom, SJ ;
Follansbee, W ;
Heller, GV ;
Kiat, H ;
Groch, MW ;
Mahmarian, JJ .
CIRCULATION, 1999, 99 (21) :2742-2749
[8]  
Hudson H. M., 1994, MED IMAGING IEEE T, V13, P601
[9]  
Kalantari F., 2014, MED IMAGING 2014 IMA, V9037
[10]  
Kalantari F., 2011, HELLENIC J NUCL MED, V14