Analysis of factors important for transurethral ultrasound prostate heating using MR temperature feedback

被引:40
作者
Chopra, R
Wachsmuth, J
Burtnyk, M
Haider, MA
Bronskill, MJ
机构
[1] Sunnybrook & Womens Coll, Hlth Sci Ctr, Toronto, ON M4N 3M5, Canada
[2] Univ Toronto, Princess Margaret Hosp, Hlth Network, Dept Med Imaging, Toronto, ON M5G 2M9, Canada
关键词
D O I
10.1088/0031-9155/51/4/005
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The feasibility of using MR thermometry for temperature feedback to control a transurethral ultrasound heating applicator with planar transducers was investigated. The sensitivity of a temperature-based feedback algorithm to spatial (control point area, slice thickness, angular alignment) and non-spatial (imaging time, temperature uncertainty) parameters was evaluated through numerical simulations. The angular alignment of the control point with the ultrasound beam was an important parameter affecting the average spatial error in beat delivery. The other spatial parameters were less influential, thus providing an opportunity to reduce spatial resolution for increased SNR in the MR imaging. The update time was the most important non-spatial parameter determining the performance of the control algorithm. Combined non-spatial and spatial parameters achieved acceptable performance with it voxel size of 3 mm x 3 mm, it 10 mm slice thickness and a 5 s update time. Temperature uncertainty of up to 2 degrees C had little effect on the performance of the control algorithm but did reduce the average error slightly Clue to a systematic, noise-induced overestimation of the boundary temperature. These simulations imply that MR thermometry performed on clinical 1.5 T imaging systems is of sufficient quality for use as thermal feedback for conformal prostate thermal therapy with transurethral ultrasound heating applicators incorporating planar transducers.
引用
收藏
页码:827 / 844
页数:18
相关论文
共 49 条
[21]   Design and preliminary results of an ultrasound applicator for interstitial thermal coagulation [J].
Lafon, C ;
Chapelon, JY ;
Prat, F ;
Gorry, F ;
Margonari, J ;
Theillere, Y ;
Cathignol, D .
ULTRASOUND IN MEDICINE AND BIOLOGY, 1998, 24 (01) :113-122
[22]   MRI monitoring of the thermal ablation of tissue: Effects of long exposure times [J].
McDannold, N ;
Hynynen, K ;
Jolesz, F .
JOURNAL OF MAGNETIC RESONANCE IMAGING, 2001, 13 (03) :421-427
[23]   Usefulness of MR imaging-derived thermometry and dosimetry in determining the threshold for tissue damage induced by thermal surgery in rabbits [J].
McDannold, NJ ;
King, RL ;
Jolesz, FA ;
Hynynen, KH .
RADIOLOGY, 2000, 216 (02) :517-523
[24]   Multi-modality tissue-mimicking phantom for thermal therapy [J].
McDonald, M ;
Lochhead, S ;
Chopra, R ;
Bronskill, MJ .
PHYSICS IN MEDICINE AND BIOLOGY, 2004, 49 (13) :2767-2778
[25]   MR thermometry-based feedback control of laser interstitial thermal therapy at 980 nm [J].
McNichols, RJ ;
Gowda, A ;
Kangasniemi, M ;
Bankson, JA ;
Price, RE ;
Hazle, JD .
LASERS IN SURGERY AND MEDICINE, 2004, 34 (01) :48-55
[26]   Automatic spatial and temporal temperature control for MR-guided focused ultrasound using fast 3D MR thermometry and multispiral trajectory of the focal point [J].
Mougenot, C ;
Salomir, R ;
Palussière, J ;
Grenier, N ;
Moonen, CTW .
MAGNETIC RESONANCE IN MEDICINE, 2004, 52 (05) :1005-1015
[27]   SOUND FIELD CALCULATION FOR RECTANGULAR SOURCES [J].
OCHELTREE, KB ;
FRIZZELL, LA .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 1989, 36 (02) :242-248
[28]   Feasibility of MR-guided focused ultrasound with real-time temperature mapping and continuous sonication for ablation of VX2 carcinoma in rabbit thigh [J].
Palussière, J ;
Salomir, R ;
Le Bail, B ;
Fawaz, R ;
Quesson, B ;
Grenier, N ;
Moonen, CTW .
MAGNETIC RESONANCE IN MEDICINE, 2003, 49 (01) :89-98
[30]  
Peters RD, 2000, MAGN RESON MED, V44, P873, DOI 10.1002/1522-2594(200012)44:6<873::AID-MRM8>3.0.CO