Thermal effects of focused ultrasound energy on bone tissue

被引:54
作者
Smith, NB
Temkin, JM
Shapiro, F
Hynynen, K
机构
[1] Harvard Univ, Brigham & Womens Hosp, Sch Med, Dept Radiol,Div MRI, Boston, MA 02115 USA
[2] Tufts Univ, Dept Biol, Medford, MA 02155 USA
[3] Harvard Univ, Sch Med, Childrens Hosp, Dept Orthopaed Surg, Boston, MA USA
关键词
HIFU; focused ultrasound; bone; muscle; femur; necrosis; coagulation; thermal;
D O I
10.1016/S0301-5629(01)00454-9
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
The effects of focused ultrasound (US) at therapeutic acoustic power levels were studied in vivo on the bone-muscle interface in rabbit thighs. The purpose of this study was to provide direction in establishing safety guidelines for treating tissue masses using focused US on or near bone. A positioning device was used to manipulate a focused US transducer (1.5 MHz) in a magnetic resonance imaging (MRI) scanner. This system was used to sonicate the femurs of 10 rabbits at acoustic power levels of 26, 39, 52 and 65 W for 10 s. The rabbits were euthanized either 4 h or 28 days after the sonications and the bone samples were harvested for histology examinations. In the femurs studied, acoustic power levels from 39 to 65 W resulted in soft tissue damage characterized grossly by coagulated tissue and bone damage depleted by yellow discoloration. Histologic examination of lesions from sonications from 39 to 65 W demonstrated that osteocyte damage and necrosis, characterized by pyknotic cells and empty lacunae, occurred within the ablation area extending through the bone. The follow-up MR images demonstrated an increase in the amount of damage in the femurs at 28 days posttreatment in comparison to images taken immediately after treatment. Focused US directed at the femur caused immediate significant thermal damage to bone in the form of osteocyte necrosis extending through the (approximately) 1 cm bone in this study. The results suggest that, when focused US energy is directed at or near bone-muscle interfaces, precautions should be taken to avoid thermal damage to the bone that can compromise its strength for extended periods. (C) 2001 World Federation for Ultrasound in Medicine & Biology.
引用
收藏
页码:1427 / 1433
页数:7
相关论文
共 30 条
[1]   ERB-YAG AND HOL-YAG LASER OSTEOTOMY - THE EFFECT OF LASER-ABLATION ON BONE HEALING [J].
BUCHELT, M ;
KUTSCHERA, HP ;
KATTERSCHAFKA, T ;
KISS, H ;
LANG, S ;
BEER, R ;
LOSERT, U .
LASERS IN SURGERY AND MEDICINE, 1994, 15 (04) :373-381
[2]   Optimization of spoiled gradient-echo phase imaging for in vivo localization of a focused ultrasound beam [J].
Chung, AH ;
Hynynen, K ;
Colucci, V ;
Oshio, K ;
Cline, HE ;
Jolesz, FA .
MAGNETIC RESONANCE IN MEDICINE, 1996, 36 (05) :745-752
[3]   EXPERIMENTAL EVALUATION OF A PROTOTYPE CYLINDRICAL SECTION ULTRASOUND HYPERTHERMIA PHASED-ARRAY APPLICATOR [J].
EBBINI, ES ;
CAIN, CA .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 1991, 38 (05) :510-520
[4]  
FRY FJ, 1993, EUR UROL, V23, P2
[5]   Study of the cause of the temperature rise at the muscle-bone interface during ultrasound hyperthermia [J].
Fujii, M ;
Sakamoto, K ;
Toda, Y ;
Negishi, A ;
Kanai, H .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1999, 46 (05) :494-504
[6]   Sparse random ultrasound phased array for focal surgery [J].
Goss, SA ;
Frizzell, LA ;
Kouzmanoff, JT ;
Barich, JM ;
Yang, JM .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 1996, 43 (06) :1111-1121
[7]   CALIBRATION OF MINIATURE ULTRASONIC RECEIVERS USING A PLANAR SCANNING TECHNIQUE [J].
HERMAN, BA ;
HARRIS, GR .
JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 1982, 72 (05) :1357-1363
[8]   Feasibility and patient tolerance of a novel transrectal ultrasound hyperthermia system for treatment of prostate cancer [J].
Hurwitz, MD ;
Kaplan, ID ;
Svensson, GK ;
Hynynen, K ;
Hansen, MS .
INTERNATIONAL JOURNAL OF HYPERTHERMIA, 2001, 17 (01) :31-37
[9]   THE ROLE OF NONLINEAR ULTRASOUND PROPAGATION DURING HYPERTHERMIA TREATMENTS [J].
HYNYNEN, K .
MEDICAL PHYSICS, 1991, 18 (06) :1156-1163
[10]   TEMPERATURE ELEVATION AT MUSCLE BONE INTERFACE DURING SCANNED, FOCUSED ULTRASOUND HYPERTHERMIA [J].
HYNYNEN, K ;
DEYOUNG, D .
INTERNATIONAL JOURNAL OF HYPERTHERMIA, 1988, 4 (03) :267-279