Investigation of absolute and relative response for three different liquid chromatography/tandem mass spectrometry systems; the impact of ionization and detection saturation

被引:20
作者
Nilsson, Lars B. [1 ]
Skansen, Patrik [1 ]
机构
[1] AstraZeneca R&D, Regulatory Bioanaly, Global DMPK, SE-43183 Molndal, Sweden
关键词
LABELED INTERNAL STANDARDS; ION SUPPRESSION; ELECTROSPRAY-IONIZATION; HUMAN PLASMA; MATRIX; CALIBRATION; LC/MS/MS; DRUGS; SUPPRESSION/ENHANCEMENT; QUANTIFICATION;
D O I
10.1002/rcm.6239
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
RATIONALE The investigations in this article were triggered by two observations in the laboratory; for some liquid chromatography/tandem mass spectrometry (LC/MS/MS) systems it was possible to obtain linear calibration curves for extreme concentration ranges and for some systems seemingly linear calibration curves gave good accuracy at low concentrations only when using a quadratic regression function. METHODS The absolute and relative responses were tested for three different LC/MS/MS systems by injecting solutions of a model compound and a stable isotope labeled internal standard. The analyte concentration range for the solutions was 0.00391 to 500 mu M (128 000 x ), giving overload of the chromatographic column at the highest concentrations. The stable isotope labeled internal standard concentration was 0.667 mu M in all samples. RESULTS The absolute response per concentration unit decreased rapidly as higher concentrations were injected. The relative response, the ratio for the analyte peak area to the internal standard peak area, per concentration unit was calculated. For system 1, the ionization process was found to limit the response and the relative response per concentration unit was constant. For systems 2 and 3, the ion detection process was the limiting factor resulting in decreasing relative response at increasing concentrations. CONCLUSIONS For systems behaving like system 1, simple linear regression can be used for any concentration range while, for systems behaving like systems 2 and 3, non-linear regression is recommended for all concentration ranges. Another consequence is that the ionization capacity limited systems will be insensitive to matrix ion suppression when an ideal internal standard is used while the detection capacity limited systems are at risk of giving erroneous results at high concentrations if the matrix ion suppression varies for different samples in a run. Copyright (C) 2012 John Wiley & Sons, Ltd.
引用
收藏
页码:1399 / 1406
页数:8
相关论文
共 23 条
[1]   13C labelled internal standards-A solution to minimize ion suppression effects in liquid chromatography-tandem mass spectrometry analyses of drugs in biological samples? [J].
Berg, Thomas ;
Strand, Dag Helge .
JOURNAL OF CHROMATOGRAPHY A, 2011, 1218 (52) :9366-9374
[2]   Systematic and comprehensive strategy for reducing matrix effects in LC/MS/MS analyses [J].
Chambers, Erin ;
Wagrowski-Diehl, Diane M. ;
Lu, Ziling ;
Mazzeo, Jeffrey R. .
JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2007, 852 (1-2) :22-34
[3]   A predictive model for matrix and analyte effects in electrospray ionization of singly-charged ionic analytes [J].
Enke, CG .
ANALYTICAL CHEMISTRY, 1997, 69 (23) :4885-4893
[4]   Matrix matching in liquid chromatography-mass spectrometry with stable isotope labelled internal standards-Is it necessary? [J].
Hewavitharana, Amitha K. .
JOURNAL OF CHROMATOGRAPHY A, 2011, 1218 (02) :359-361
[5]   Monitoring phospholipids for assessment of ion enhancement and ion suppression in ESI and APCI LC/MS/MS for chlorpheniramine in human plasma and the importance of multiple source matrix effect evaluations [J].
Ismaiel, Omnia A. ;
Halquist, Matthew S. ;
Elmamly, Magda Y. ;
Shalaby, Abdalla ;
Karnes, H. Thomas .
JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2008, 875 (02) :333-343
[6]   The use of human plasma as matrix for calibration standards in pre-clinical LC-MS/MS methods-A way to reduce animal use [J].
Jacobson, Britt-Marie ;
Olsson, Anders ;
Fakt, Christina ;
Ohman, Daniel .
JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 2011, 54 (04) :826-829
[7]   Identifying, Evaluating, and Controlling Bioanalytical Risks Resulting from Nonuniform Matrix Ion Suppression/Enhancement and Nonlinear Liquid Chromatography-Mass Spectrometry Assay Response [J].
Liu, Guowen ;
Ji, Qin C. ;
Arnold, Mark E. .
ANALYTICAL CHEMISTRY, 2010, 82 (23) :9671-9677
[8]   Strategies for the assessment of matrix effect in quantitative bioanalytical methods based on HPLC-MS/MS [J].
Matuszewski, BK ;
Constanzer, ML ;
Chavez-Eng, CM .
ANALYTICAL CHEMISTRY, 2003, 75 (13) :3019-3030
[9]   Standard line slopes as a measure of a relative matrix effect in quantitative HPLC-MS bioanalysis [J].
Matuszewski, BK .
JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2006, 830 (02) :293-300
[10]   Matrix effect in quantitative LC/MS/MS analyses of biological fluids: A method for determination of finasteride in human plasma at picogram per milliliter concentrations [J].
Matuszewski, BK ;
Constanzer, ML ;
Chavez-Eng, CM .
ANALYTICAL CHEMISTRY, 1998, 70 (05) :882-889