A novel photoacoustic tomography based on a time-resolved technique and an acoustic lens imaging system

被引:24
作者
He, Yongheng [1 ]
Tang, Zhilie [1 ]
Chen, Zhanxu [1 ]
Wan, Wei [1 ]
Li, Jianghua [1 ]
机构
[1] S China Normal Univ, Sch Phys & Telecommun Engn, Guangzhou 510006, Peoples R China
关键词
D O I
10.1088/0031-9155/51/10/019
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
A novel photoacoustic (PA) tornographic method, which is based on a time-resolved technique and an acoustic lens imaging system, is presented in this paper. A YAG laser operating at 532 nm with a 7 ns pulse width and 10 mJ optical pulse is employed as the excitation source to irradiate the tissue. PA signals generated from the tissue are imaged onto a multi-element linear array transducer with an acoustic lens. A 64 electronic switch is efficiently used for changing the parallel PA signals into a series. The proposed method directly provides PA images without any complex reconstruction algorithms. With the time-resolved technique, tomographic imaging can be achieved successfully. The results show that the images agree well with the original samples.
引用
收藏
页码:2671 / 2680
页数:10
相关论文
共 20 条
[1]  
ALFANO RR, 1996, OSA TRENDS OPTICS PH, V2
[2]  
Cox BA, 1998, AM SURGEON, V64, P934
[3]   Laser optoacoustic imaging for breast cancer diagnostics: Limit of detection and comparison with X-ray and ultrasound imaging [J].
Esenaliev, RO ;
Karabutov, AA ;
Tittel, FK ;
Fornage, BD ;
Thomsen, SL ;
Stelling, C ;
Oraevsky, AA .
OPTICAL TOMOGRAPHY AND SPECTROSCOPY OF TISSUE: THEORY, INSTRUMENTATION, MODEL, AND HUMAN STUDIES II, PROCEEDINGS OF, 1997, 2979 :71-82
[4]   Three-dimensional photoacoustic imaging of blood vessels in tissue [J].
Hoelen, CGA ;
de Mul, FFM ;
Pongers, R ;
Dekker, A .
OPTICS LETTERS, 1998, 23 (08) :648-650
[5]   Backward mode detection of laser-induced wide-band ultrasonic transients with optoacoustic transducer [J].
Karabutov, AA ;
Savateeva, EV ;
Podymova, NB ;
Oraevsky, AA .
JOURNAL OF APPLIED PHYSICS, 2000, 87 (04) :2003-2014
[6]  
KARABUTOV AA, 1995, P SOC PHOTO-OPT INS, V2389, P209
[7]  
Köstli KP, 2000, J APPL PHYS, V88, P1632, DOI 10.1063/1.373864
[8]   Thermoacoustic computed tomography-technical considerations [J].
Kruger, RA ;
Reinecke, DR ;
Kruger, GA .
MEDICAL PHYSICS, 1999, 26 (09) :1832-1837
[9]   Multiple-bandwidth photoacoustic tomography [J].
Ku, G ;
Wang, XD ;
Stoica, G ;
Wang, LHV .
PHYSICS IN MEDICINE AND BIOLOGY, 2004, 49 (07) :1329-1338
[10]   Real-time three-dimensional optoacoustic imaging using an acoustic lens system [J].
Niederhauser, JJ ;
Jaeger, M ;
Frenz, M .
APPLIED PHYSICS LETTERS, 2004, 85 (05) :846-848