Downloadable Algorithm to Reduce Inappropriate Shocks Caused by Fractures of Implantable Cardioverter-Defibrillator Leads

被引:102
作者
Swerdlow, Charles D. [1 ]
Gunderson, Bruce D. [2 ]
Ousdigian, Kevin T. [2 ]
Abeyratne, Athula [2 ]
Stadler, Robert W. [2 ]
Gillberg, Jeffrey M. [2 ]
Patel, Amisha S. [2 ]
Ellenbogen, Kenneth A. [3 ]
机构
[1] Cedars Sinai Med Ctr, Dept Cardiol, Los Angeles, CA 90048 USA
[2] Medtronic Inc, Mounds View, MN USA
[3] Virginia Commonwealth Univ, Div Cardiol, Med Coll Virginia, Richmond, VA 23284 USA
关键词
defibrillators; lead; lead failure; recall;
D O I
10.1161/CIRCULATIONAHA.108.796136
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background-The primary method for monitoring implantable cardioverter-defibrillator lead integrity is periodic measurement of impedance. Sprint Fidelis leads are prone to pace-sense lead fractures, which commonly present as inappropriate shocks caused by oversensing. Methods and Results-We developed and tested an algorithm to enhance early identification of lead fractures and to reduce inappropriate shocks. This lead-integrity algorithm, which can be downloaded into presently implanted implantable cardioverter-defibrillators, alerts the patient and/or physician when triggered by either oversensing or excessive increases in impedance. To reduce inappropriate shocks, the lead-integrity algorithm increases the number of intervals to detect (NID) ventricular fibrillation when triggered. The lead-integrity algorithm was tested on data from 15 970 patients with Fidelis leads (including 121 with clinically diagnosed fractures) and 95 other fractured leads confirmed by analysis of returned product. The effect of the NID on inappropriate shocks was tested in 92 patients with 927 shocks caused by lead fracture. Increasing the NID reduced inappropriate shocks (P < 0.0001). The lead-integrity algorithm provided at least a 3-day warning of inappropriate shocks in 76% (95% CI, 66 to 84) of patients versus 55% (95% CI, 43 to 64) for optimal impedance monitoring (P = 0.007). Its positive predictive value was 72% for lead fractures and 81% for lead fractures or header-connector problems requiring surgical intervention. The false-positive rate was 1 per 372 patient-years of monitoring. Conclusions-A lead-integrity algorithm developed for download into existing implantable cardioverter-defibrillators increases short-term warning of inappropriate shocks in patients with lead fractures and reduces the likelihood of inappropriate shocks. It is the first downloadable RAMware to enhance the performance of nominally functioning implantable cardioverter-defibrillators and the first implantable cardioverter-defibrillator monitoring feature that triggers real-time changes in ventricular fibrillation detection parameters to reduce inappropriate shocks. (Circulation. 2008; 118: 2122-2129.)
引用
收藏
页码:2122 / 2129
页数:8
相关论文
共 20 条
[1]   Patient alert in implantable cardioverter defibrillators: Toy or tool? [J].
Becker, R ;
Ruf-Richter, J ;
Senges-Becker, JC ;
Bauer, A ;
Weretka, S ;
Voss, F ;
Katus, HA ;
Schoels, W .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2004, 44 (01) :95-98
[2]   When life-saving devices terminate life [J].
Catanchin, Andrei ;
Anderson, Lisa ;
Jones, Sue ;
Ward, David .
JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2008, 19 (03) :316-318
[3]  
GROVES R, 2007, URGENT MED DEVICE IN
[4]  
GUNDERSON B, 2008, HEART RHYTHM, V5, pS340
[5]   An algorithm to predict implantable cardioverter-defibrillator lead failure [J].
Gunderson, BD ;
Patel, AS ;
Bounds, CA ;
Shepard, RK ;
Wood, MA ;
Ellenbogen, KA .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2004, 44 (09) :1898-1902
[6]   Automatic identification of implantable cardioverter-defibrillator lead problems using intracardiac electrograms [J].
Gunderson, BD ;
Patel, AS ;
Bounds, CA .
COMPUTERS IN CARDIOLOGY 2002, VOL 29, 2002, 29 :121-124
[7]   Effect of programmed number of intervals to detect ventricular fibrillation on implantable cardioverter-defibrillator aborted and unnecessary shocks [J].
Gunderson, Bruce D. ;
Abeyratne, Athula I. ;
Olson, Walter H. ;
Swerdlow, Charles D. .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2007, 30 (02) :157-165
[8]   Early failure of a small-diameter high-voltage implantable cardioverter-defibrillator lead [J].
Hauser, Robert G. ;
Kallinen, Linda M. ;
Almquist, Adrian K. ;
Gornick, Charles C. ;
Katsiyiannis, William T. .
HEART RHYTHM, 2007, 4 (07) :892-896
[9]   Failure of impedance monitoring to prevent adverse clinical events caused by fracture of a recalled high-voltage implantable cardioverter-defibrillator lead [J].
Kallinen, Linda M. ;
Hauser, Robert G. ;
Lee, Ken W. ;
Almquist, Adrian K. ;
Katsiyiannis, William T. ;
Tang, Chuen Y. ;
Melby, Daniel P. ;
Gornick, Charles C. .
HEART RHYTHM, 2008, 5 (06) :775-779
[10]   Annual rate of transvenous defibrillation lead defects in implantable cardioverter-defibrillators over a period of &gt; 10 years [J].
Kleemann, Thomas ;
Becker, Torsten ;
Doenges, Klaus ;
Vater, Margit ;
Senges, Jochen ;
Schneider, Steffen ;
Saggau, Werner ;
Weisse, Udo ;
Seidl, Karlheinz .
CIRCULATION, 2007, 115 (19) :2474-2480