Novel motor design for rotating anode x-ray tubes operating in the fringe field of a magnetic resonance imaging system

被引:3
|
作者
Lillaney, Prasheel [1 ,2 ]
Shin, Mihye [1 ,3 ]
Hinshaw, Waldo [1 ]
Bennett, N. Robert [1 ,4 ]
Pelc, Norbert [1 ,2 ]
Fahrig, Rebecca [1 ]
机构
[1] Stanford Univ, Dept Radiol, Stanford, CA 94305 USA
[2] Stanford Univ, Dept Bioengn, Stanford, CA 94305 USA
[3] Stanford Univ, Dept Mech Engn, Stanford, CA 94305 USA
[4] Qualcomm MEMS Technol, San Jose, CA 95134 USA
关键词
XMR; motor design; x-ray production; magnetic fields; MR compatibility; PROSTATE BRACHYTHERAPY IMPLANTS; COMBINED MR; XMR; FEASIBILITY; DOSIMETRY;
D O I
10.1118/1.4773313
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: Using hybrid x-ray/MR (XMR) systems for image guidance during interventional procedures could enhance the diagnosis and treatment of neurologic, oncologic, cardiovascular, and other disorders. The authors propose a close proximity hybrid system design in which a C-arm fluoroscopy unit is placed immediately adjacent to the solenoid magnet of a MR system with a minimum distance of 1.2 m between the x-ray and MR imaging fields of view. Existing rotating anode x-ray tube designs fail within MR fringe field environments because the magnetic fields alter the electron trajectories in the x-ray tube and act as a brake on the induction motor, reducing the rotation speed of the anode. In this study the authors propose a novel motor design that avoids the anode rotation speed reduction. Methods: The proposed design replaces the permanent magnet stator found in brushed dc motors with the radial component of the MR fringe field. The x-ray tube is oriented such that the radial component of the MR fringe field is orthogonal to the cathode-anode axis. Using a feedback position sensor and the support bearings as electrical slip rings, the authors use electrical commutation to eliminate the need for mechanical brushes and commutators. A vacuum compatible prototype of the proposed motor design was assembled, and its performance was evaluated at various operating conditions. The prototype consisted of a 3.1 in. diameter anode rated at 300 kHU with a ceramic rotor that was 5.6 in. in length and had a 2.9 in. diameter. The material chosen for all ceramic components was MACOR, a machineable glass ceramic developed by Corning Inc. The approximate weight of the entire assembly was 1750 g. The maximum rotation speed, angular acceleration, and acceleration time of the motor design were investigated, as well as the dependence of these parameters on rotor angular offset, magnetic field strength, and field orientation. The resonance properties of the authors' assembly were also evaluated to determine its stability during acceleration, and a pulse width modulation algorithm was implemented to control the rotation speed of the motor. Results: At a magnetic flux density of 41 mT orthogonal to the axis of rotation (on the lower end of the expected flux density in the MR suite) the maximum speed of the motor was found to be 5150 revolutions per minute (rpm). The acceleration time necessary to reach 3000 rpm was found to be approximately 10 s at 59 mT. The resonance frequency of the assembly with the anode attached was 1310 rpm (21.8 Hz) which is far below the desired operating speeds. Pulse width modulation provides an effective method to control the speed of the motor with a resolution of 100 rpm. Conclusions: The proposed design can serve as a direct replacement to the conventional induction motor used in rotating anode x-ray tubes. It does not suffer from a reduced rotation speed when operating in a MR environment. The presence of chromic steel bearings in the prototype prevented testing at the higher field strengths, and future iterations of the design could eliminate this shortcoming. The prototype assembly demonstrates proof of concept of the authors' design and overcomes one of the major obstacles for a MR compatible rotating anode x-ray tube. (c) 2013 American Association of Physicists in Medicine. [http://dx.doi.org/10.1118/1.4773313]
引用
收藏
页数:14
相关论文
共 50 条
  • [1] Resonant Frequency of Rotating Anode X-Ray Tubes
    Shin, M.
    Lillaney, P.
    Hinshaw, W.
    Fahrig, R.
    MEDICAL PHYSICS, 2012, 39 (06) : 3622 - 3622
  • [2] HEAVY-DUTY ROTATING ANODE X-RAY TUBES
    YAMAMURA, T
    SAWA, T
    YONEYAMA, K
    UEMURA, K
    ISHII, Y
    TOSHIBA REVIEW, 1980, (125): : 41 - 46
  • [3] Design and development of a rotating-anode x-ray tube coherent scatter projection imaging system
    Dydula, Christopher
    Xu, Tong
    Johns, Paul C.
    ANOMALY DETECTION AND IMAGING WITH X-RAYS (ADIX) V, 2020, 11404
  • [4] Design optimization of MR-compatible rotating anode x-ray tubes for stable operation
    Shin, Mihye
    Lillaney, Prasheel
    Hinshaw, Waldo
    Fahrig, Rebecca
    MEDICAL PHYSICS, 2013, 40 (11)
  • [5] FACTORS AFFECTING DESIGN AND OPERATION OF ROLLING ELEMENT BEARINGS IN ROTATING ANODE X-RAY TUBES
    KESSLER, GR
    MAHN, GR
    LUBRICATION ENGINEERING, 1967, 23 (09): : 372 - &
  • [7] Design principles of a novel X-ray imaging system
    喻春雨
    常本康
    王世允
    王庆宝
    ChineseOpticsLetters, 2006, (03) : 158 - 159
  • [8] Design of An MR Compatible Rotating Anode X-Ray Tube
    Lillaney, P.
    Shin, M.
    Hinshaw, W.
    Fahrig, R.
    MEDICAL PHYSICS, 2012, 39 (06) : 3949 - 3949
  • [9] NONSTEADY TEMPERATURE-FIELD IN ROTATING ANODE OF AN X-RAY TUBE
    GAMER, U
    ACTA TECHNICA ACADEMIAE SCIENTIARUM HUNGARICAE, 1976, 82 (1-2): : 47 - 52
  • [10] Design of a nanoprobe for high field magnetic resonance imaging, dual energy X-ray computed tomography and luminescent imaging
    Gonzalez-Mancebo, Daniel
    Isabel Becerro, Ana
    Corral, Ariadna
    Garcia-Embid, Sonia
    Balcerzyk, Marcin
    Luisa Garcia-Martin, Maria
    de la Fuente, Jesus M.
    Ocana, Manuel
    JOURNAL OF COLLOID AND INTERFACE SCIENCE, 2020, 573 : 278 - 286