Quantification of seven β-lactam antibiotics and two β-lactamase inhibitors in human plasma using a validated UPLC-MS/MS method

被引:85
作者
Carlier, Mieke [1 ]
Stove, Veronique [1 ]
Roberts, Jason A. [2 ]
Van de Velde, Eric [1 ]
De Waele, Jan J. [3 ]
Verstraete, Alain G. [1 ]
机构
[1] Univ Ghent, Dept Clin Chem Microbiol & Immunol, B-9000 Ghent, Belgium
[2] Univ Queensland, Burns Trauma & Crit Care Res Ctr, Brisbane, Qld, Australia
[3] Univ Ghent, Dept Crit Care Med, B-9000 Ghent, Belgium
基金
英国医学研究理事会;
关键词
beta-Lactam antibiotics; Therapeutic drug monitoring; UPLC-MS/MS; PERFORMANCE LIQUID-CHROMATOGRAPHY; TANDEM MASS-SPECTROMETRY; ULTRAVIOLET DETECTION; BIOANALYTICAL METHODS; CLAVULANIC ACID; HPLC; PIPERACILLIN; INFECTION; MEROPENEM; SERUM;
D O I
10.1016/j.ijantimicag.2012.06.022
中图分类号
R51 [传染病];
学科分类号
100401 ;
摘要
There is an increasing interest in monitoring plasma concentrations of beta-lactam antibiotics. The objective of this work was to develop and validate a rapid ultra-performance liquid chromatographic method with tandem mass spectrometric detection (UPLC-MS/MS) for simultaneous quantification of amoxicillin, ampicillin, cefuroxime, cefazolin, ceftazidime, meropenem, piperacillin, clavulanic acid and tazobactam. Sample clean-up included protein precipitation with acetonitrile and back-extraction of acetonitrile with dichloromethane. Six deuterated beta-lactam antibiotics were used as internal standards. Chromatographic separation was performed on a Waters ACQUITY UPLC system using a BEH C-18 column (1.7 mu m, 100 mm x 2.1 mm) applying a binary gradient elution of water and acetonitrile both containing 0.1% formic acid. The total run time was 5.5 min. The developed method was validated in terms of precision, accuracy, linearity, matrix effect and recovery. The assay has now been successfully used to determine concentrations of amoxicillin/clavulanic acid, cefuroxime and meropenem in plasma samples from intensive care patients. (C) 2012 Elsevier B.V. and the International Society of Chemotherapy. All rights reserved.
引用
收藏
页码:416 / 422
页数:7
相关论文
共 24 条
[1]   Linear regression for calibration lines revisited: weighting schemes for bioanalytical methods [J].
Almeida, AM ;
Castel-Branco, MM ;
Falcao, AC .
JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2002, 774 (02) :215-222
[2]  
[Anonymous], APPL LC MS TOXICOLOG
[3]   Daily serum piperacillin monitoring is advisable in critically ill patients [J].
Blondiaux, Nicolas ;
Wallet, Frederic ;
Favory, Raphael ;
Onimus, Thierry ;
Nseir, Saad ;
Courcol, Rene J. ;
Durocher, Alain ;
Roussel-Delvallez, Micheline .
INTERNATIONAL JOURNAL OF ANTIMICROBIAL AGENTS, 2010, 35 (05) :500-503
[4]   A liquid chromatography assay for a quantification of doripenem, ertapenem, imipenem, meropenem concentrations in human plasma: Application to a clinical pharmacokinetic study [J].
Dailly, E. ;
Bouquie, R. ;
Deslandes, G. ;
Jolliet, P. ;
Le Floch, R. .
JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2011, 879 (15-16) :1137-1142
[5]   Simultaneous determination of five β-lactam antibiotics (cefepim, ceftazidim, cefuroxim, meropenem and piperacillin) in human plasma by high-performance liquid chromatography with ultraviolet detection [J].
Denooz, Raphael ;
Charlier, Corinne .
JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2008, 864 (1-2) :161-167
[6]  
EMEA (European Medicines Agency), 2011, EMEA/CHMP/EWP/192217/2009
[7]  
EUCAST, 2012, ANT WILD TYP DISTR M
[8]  
International Conference on Harmonization (ICH), 1996, VAL AN METH METH ICH
[9]  
Leibovici L, 1998, J INTERN MED, V244, P379
[10]   Strategies for the assessment of matrix effect in quantitative bioanalytical methods based on HPLC-MS/MS [J].
Matuszewski, BK ;
Constanzer, ML ;
Chavez-Eng, CM .
ANALYTICAL CHEMISTRY, 2003, 75 (13) :3019-3030