Influence of the number of elongated fiducial markers on the localization accuracy of the prostate

被引:11
作者
de Boer, Johan [1 ]
de Bois, Josien [1 ]
van Herk, Marcel [1 ]
Sonke, Jan-Jakob [1 ]
机构
[1] Antoni Van Leeuwenhoek Hosp, Netherlands Canc Inst, Dept Radiat Oncol, NL-1066 CX Amsterdam, Netherlands
关键词
EXTERNAL-BEAM RADIOTHERAPY; RIGID-BODY; POSITION VERIFICATION; ERROR; CANCER; CT; FEASIBILITY; PRECISION; STRATEGY; MOTION;
D O I
10.1088/0031-9155/57/19/6211
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Implanting fiducial markers for localization purposes has become an accepted practice in radiotherapy for prostate cancer. While many correction strategies correct for translations only, advanced correction protocols also require knowledge of the rotation of the prostate. For this purpose, typically, three or more markers are implanted. Elongated fiducial markers provide more information about their orientation than traditional round or cylindrical markers. Potentially, fewer markers are required. In this study, we evaluate the effect of the number of elongated markers on the localization accuracy of the prostate. To quantify the localization error, we developed a model that estimates, at arbitrary locations in the prostate, the registration error caused by translational and rotational uncertainties of the marker registration. Every combination of one, two and three markers was analysed for a group of 24 patients. The average registration errors at the prostate surface were 0.3-0.8 mm and 0.4-1 mm for registrations on, respectively, three markers and two markers located on different sides of the prostate. Substantial registration errors (2.0-2.2 mm) occurred at the prostate surface contralateral to the markers when two markers were implanted on the same side of the prostate or only one marker was used. In conclusion, there is no benefit in using three elongated markers: two markers accurately localize the prostate if they are implanted at some distance from each other.
引用
收藏
页码:6211 / 6226
页数:16
相关论文
共 27 条
[1]   Automatic online adaptive radiation therapy techniques for targets with significant shape change: a feasibility study [J].
Court, LE ;
Tishler, RB ;
Petit, J ;
Cormack, R ;
Chin, L .
PHYSICS IN MEDICINE AND BIOLOGY, 2006, 51 (10) :2493-2501
[2]   Clinical feasibility study for the use of implanted gold seeds in the prostate as reliable positioning markers during megavoltage irradiation [J].
Dehnad, H ;
Nederveen, AJ ;
van der Heide, UA ;
van Moorselaar, RJA ;
Hofman, P ;
Lagendijk, JJW .
RADIOTHERAPY AND ONCOLOGY, 2003, 67 (03) :295-302
[3]   Predicting error in rigid-body point-based registration [J].
Fitzpatrick, JM ;
West, JB ;
Maurer, CR .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1998, 17 (05) :694-702
[4]   Strategies to reduce the systematic error due to tumor and rectum motion in radiotherapy of prostate cancer [J].
Hoogeman, MS ;
van Herk, M ;
de Bois, J ;
Lebesque, JV .
RADIOTHERAPY AND ONCOLOGY, 2005, 74 (02) :177-185
[5]   EFFECTIVENESS OF USING FEWER IMPLANTED FIDUCIAL MARKERS FOR PROSTATE TARGET ALIGNMENT [J].
Kudchadker, Rajat J. ;
Lee, Andrew K. ;
Yu, Zhiqian H. ;
Johnson, Jennifer L. ;
Zhang, Lifei ;
Zhang, Yongbin ;
Amos, Richard A. ;
Nakanishi, Hiroyuki ;
Ochiai, Atsushi ;
Dong, Lei .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2009, 74 (04) :1283-1289
[6]  
LANGRON SP, 1985, J ROY STAT SOC B MET, V47, P277
[7]   Assessment of residual error for online cone-beam CT-guided treatment of prostate cancer patients [J].
Létourneau, D ;
Martinez, AA ;
Lockman, D ;
Yan, D ;
Vargas, C ;
Ivaldi, G ;
Wong, J .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2005, 62 (04) :1239-1246
[8]   Daily prostate targeting using implanted radiopaque markers [J].
Litzenberg, D ;
Dawson, LA ;
Sandler, H ;
Sanda, MG ;
McShan, D ;
Ten Haken, RK ;
Lam, KL ;
Brock, KK ;
Balter, JM .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2002, 52 (03) :699-703
[9]   Long-term experience with transrectal and transperineal implantations of fiducial gold markers in the prostate for position verification in external beam radiotherapy; feasibility, toxicity and quality of life [J].
Moman, Maaike R. ;
van der Heide, Uulke A. ;
Kotte, Alexis N. T. J. ;
van Moorselaar, R. Jeroen A. ;
Bol, Gijsbert H. ;
Franken, Stefan P. G. ;
van Vulpen, Marco .
RADIOTHERAPY AND ONCOLOGY, 2010, 96 (01) :38-42
[10]   Fiducial-based targeting accuracy for external-beam radiotherapy [J].
Murphy, MJ .
MEDICAL PHYSICS, 2002, 29 (03) :334-344