Non-invasive method for rapid bedside estimation of inotropy: theory and preliminary clinical validation

被引:31
作者
Smith, B. E. [1 ,2 ]
Madigan, V. M. [1 ]
机构
[1] Charles Sturt Univ, Sch Biomed Sci, Bathurst, NSW 2795, Australia
[2] Bathurst Base Hosp, Dept Anaesthet & Intens Care, Bathurst, NSW 2795, Australia
关键词
Doppler ultrasonography; inotropism cardiac; left ventricular function; myocardial contractility; systemic vascular resistance; CONGESTIVE-HEART-FAILURE; VENTRICULAR EJECTION FRACTION; CARDIAC-OUTPUT MONITOR; CONTRACTILITY; LOAD; ECHOCARDIOGRAPHY; RELIABILITY; PERFORMANCE; PREVALENCE; PARAMETERS;
D O I
10.1093/bja/aet118
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Background. There are numerous techniques which attempt to quantify inotropy (or myocardial contractility). None has yet found general acceptance in anaesthesia and critical care as a practical method. We report a novel approach to the determination of inotropy as a bedside procedure which could identify low inotropy states in patients with clinical heart failure. Methods. We estimated the potential and kinetic energy delivered by the left ventricle using continuous-wave Doppler ultrasound (ultrasonic cardiac output monitor, Uscom, Sydney, Australia) and data available at the point of care. A formula to calculate effective inotropy [Smith-Madigan inotropy index (SMII)] was tested against historical haemodynamic data for 250 control subjects (ASA I patients from preoperative clinic) and 83 patients with acute left ventricular failure (LVF) of New York Heart Association Grade 4 (LVF group). The ratio of potential to kinetic energy (PKR) was investigated as a measure of arterial impedance. Results. Significant differences were found between the control and LVF groups for cardiac index, mean (range)=3.37 (2.84-5.32) vs 1.84 (1.43-2.26) litre min (-1) m(-2); stroke volume index (SVI), 49.2 (39-55) vs 34.3 (23-37) ml m(-2); systemic vascular resistance, 893 (644-1242) vs 1960 (1744-4048) dyn s cm(-5); SMII, 1.78 (1.35-2.24) vs 0.73 (0.43-0.97) W m(-2); and PKR, 29:1 (24-35:1) vs 124:1 (96-174:1), P<0.001 in each case. Normal ranges were calculated for SMII and PKR as mean (+/-1.96) standard deviations, yielding 1.6-2.2 W M-2 for SMII, and 25-34:1 for PKR. Conclusion. The method clearly identified the two clinical groups with no overlap of data points. The discriminant power of SMII and PKR may offer valuable diagnostic methods and monitoring tools in anaesthesia and critical care. This is the first report of normal ranges for SMII and PKR.
引用
收藏
页码:580 / 588
页数:9
相关论文
共 33 条
  • [1] Myocardial contractility by strain echocardiography: comparison with physiological measurements in an in vitro model
    Abraham, TP
    Laskowski, C
    Zhan, WZ
    Belohlavek, M
    Martin, EA
    Greenleaf, JF
    Sieck, GC
    [J]. AMERICAN JOURNAL OF PHYSIOLOGY-HEART AND CIRCULATORY PHYSIOLOGY, 2003, 285 (06): : H2599 - H2604
  • [2] Comparison of left ventricular ejection fraction and volumes in heart failure by echocardiography, radionuclide ventriculography and cardiovascular magnetic resonance - Are they interchangeable?
    Bellenger, NG
    Burgess, MI
    Ray, SG
    Lahiri, A
    Coats, AJS
    Cleland, JGF
    Pennell, DJ
    [J]. EUROPEAN HEART JOURNAL, 2000, 21 (16) : 1387 - 1396
  • [3] Outcome of heart failure with preserved ejection fraction in a population-based study
    Bhatia, R. Sacha
    Tu, Jack V.
    Lee, Douglas S.
    Austin, Peter C.
    Fang, Jiming
    Haouzi, Annick
    Gong, Yanyan
    Liu, Peter P.
    [J]. NEW ENGLAND JOURNAL OF MEDICINE, 2006, 355 (03) : 260 - 269
  • [4] Increased ventricular contractility is not sufficient for effective positive inotropic intervention
    Binkley, PF
    VanFossen, DB
    Haas, GJ
    Leier, CV
    [J]. AMERICAN JOURNAL OF PHYSIOLOGY-HEART AND CIRCULATORY PHYSIOLOGY, 1996, 271 (04): : H1635 - H1642
  • [5] INFLUENCE OF POSITIVE INOTROPIC THERAPY ON PULSATILE HYDRAULIC LOAD AND VENTRICULAR-VASCULAR COUPLING IN CONGESTIVE HEART-FAILURE
    BINKLEY, PF
    VANFOSSEN, DB
    NUNZIATA, E
    UNVERFERTH, DV
    LEIER, CV
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1990, 15 (05) : 1127 - 1135
  • [6] Cannon WB, 1924, BIOGRAPHICAL MEMOIR
  • [7] The normal ranges of cardiovascular parameters in children measured using the Ultrasonic Cardiac Output Monitor
    Cattermole, Giles N.
    Leung, P. Y. Mia
    Mak, Paulina S. K.
    Chan, Stewart S. W.
    Graham, Colin A.
    Rainer, Timothy H.
    [J]. CRITICAL CARE MEDICINE, 2010, 38 (09) : 1875 - 1881
  • [8] CATTERMOLE GN., 2009, HONG KONG J EMERG ME, V16, P288
  • [9] The effects of changes in loading conditions and modulation of inotropic state on the myocardial performance index: comparison with conductance catheter measurements
    Cheung, MMH
    Smallhorn, JF
    Redington, AN
    Vogel, M
    [J]. EUROPEAN HEART JOURNAL, 2004, 25 (24) : 2238 - 2242
  • [10] LEFT-VENTRICULAR END-SYSTOLIC WALL STRESS-VELOCITY OF FIBER SHORTENING RELATION - A LOAD-INDEPENDENT INDEX OF MYOCARDIAL-CONTRACTILITY