Filterscopes: Spectral line monitors for long-pulse plasma devices

被引:15
作者
Brooks, N. H. [1 ]
Colchin, R. J. [2 ]
Fehling, D. T. [2 ]
Hillis, D. L. [2 ]
Mu, Y. [3 ]
Unterberg, E. [4 ]
机构
[1] Gen Atom Co, San Diego, CA 92186 USA
[2] Oak Ridge Natl Lab, Oak Ridge, TN 37831 USA
[3] Univ Toronto, Toronto, ON M5S 3G4, Canada
[4] Oak Ridge Inst Sci & Educ, Oak Ridge, TN 37831 USA
关键词
D O I
10.1063/1.2957777
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
A photomultiplier (PMT)-based diagnostic system for monitoring spectral lines along multiple viewchords, named the "Filterscope" [R. J. Colchin et al., Rev. Sci. Instrum. 74, 2068 (2003)], is currently in use at the DIII-D, NSTX, and CDX-U fusion plasma devices in the US, and has been installed at the KSTAR device in Korea. This diagnostic has recently been upgraded for application to long-pulse devices, such as KSTAR, EAST in China, and the future ITER in France. A new data acquisition system, employing the PXI instrumentation platform with an embedded Windows microprocessor controller, can simultaneously record up to 72 channels at 100 kHz sampling rates for plasma periods lasting up to 20 min. Based on the average signal level during an adjustable time interval (100 ms in the present DIII-D implementation), the controller digitally adjusts PMT dynode voltage throughout the course of a discharge, thereby maintaining the output signals at a level where they are neither saturated nor dominated by digitizer noise. The new system's ability to accommodate large variations in source strength, discharge to discharge and within a single discharge, has proved particularly valuable during DIII-D operations, since changes between top, bottom, and double-null divertor magnetic configurations lead to large temporal variations in signal brightness. (C) 2008 American Institute of Physics. [DOI: 10.1063/1.2957777]
引用
收藏
页数:4
相关论文
共 7 条
[1]   Neutral particle fueling at the midplane of DIII-D [J].
Colchin, RJ ;
Owen, LW .
JOURNAL OF NUCLEAR MATERIALS, 2003, 313 :609-613
[2]   The Filterscope [J].
Colchin, RJ ;
Hillis, DL ;
Maingi, R ;
Klepper, CC ;
Brooks, NH .
REVIEW OF SCIENTIFIC INSTRUMENTS, 2003, 74 (03) :2068-2070
[3]   Physics of slow L-H transitions in the DIII-D tokamak [J].
Colchin, RJ ;
Carreras, BA ;
Maingi, R ;
Baylor, LR ;
Jernigan, TC ;
Schaffer, MJ ;
Carlstrom, TN ;
Brooks, NH ;
Greenfield, CM ;
Gohil, P ;
McKee, GR ;
Rudakov, DL ;
Rhodes, TL ;
Doyle, EJ ;
Austin, ME ;
Watkins, JG .
NUCLEAR FUSION, 2002, 42 (09) :1134-1143
[4]   Detailed comparison of simulated and measured plasma profiles in the scrape-off layer and edge plasma of DIII-D [J].
Porter, GD ;
Isler, R ;
Boedo, J ;
Rognlien, TD .
PHYSICS OF PLASMAS, 2000, 7 (09) :3663-3680
[5]   Edge localized mode particle losses from the DIII-D tokamak [J].
Porter, GD ;
Casper, TA ;
Moller, JM .
PHYSICS OF PLASMAS, 2001, 8 (12) :5140-5150
[6]  
*RES SYST INC BOUL, 1998, BUILD IDL APPL IDL V
[7]   MDSsplus data acquisition system [J].
Stillerman, JA ;
Fredian, TW ;
Klare, KA ;
Manduchi, G .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1997, 68 (01) :939-942