Free-Breathing Contrast-Enhanced Multiphase MRI of the Liver Using a Combination of Compressed Sensing, Parallel Imaging, and Golden-Angle Radial Sampling

被引:212
作者
Chandarana, Hersh [1 ]
Feng, Li [1 ]
Block, Tobias K. [1 ]
Rosenkrantz, Andrew B. [1 ]
Lim, Ruth P. [1 ]
Babb, James S. [1 ]
Sodickson, Daniel K. [1 ]
Otazo, Ricardo [1 ]
机构
[1] NYU, Langone Med Ctr, Dept Radiol, Ctr Biomed Imaging, New York, NY 10016 USA
基金
美国国家卫生研究院;
关键词
contrast-enhanced MRI; liver MRI; free-breathing MRI; compressed sensing reconstruction; multicoil compressed sensing reconstruction; golden-angle radial acquisition; GRADIENT-ECHO SEQUENCE; PROJECTION RECONSTRUCTION; LESIONS;
D O I
10.1097/RLI.0b013e318271869c
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objective: The objectives of this study were to develop a new method for free-breathing contrast-enhanced multiphase liver magnetic resonance imaging (MRI) using a combination of compressed sensing, parallel imaging, and radial k-space sampling and to demonstrate the feasibility of this method by performing image quality comparison with breath-hold cartesian T1-weighted (conventional) postcontrast acquisitions in healthy participants. Materials and Methods: This Health Insurance Portability and Accountability Act-compliant prospective study received approval from the institutional review board. Eight participants underwent 3 separate contrast-enhanced fat-saturated T1-weighted gradient-echo MRI examinations with matching imaging parameters: conventional breath-hold examination with cartesian k-space sampling volumetric interpolate breath hold examination (BH-VIBE) and free-breathing acquisitions with interleaved angle-bisection and continuous golden-angle radial sampling schemes. Interleaved angle-bisection and golden-angle data from each 100 consecutive spokes were reconstructed using a combination of compressed sensing and parallel imaging (interleaved-angle radial sparse parallel [IARASP] and golden-angle radial sparse parallel [GRASP]) to generate multiple postcontrast phases. Arterial-and venous-phase BH-VIBE, IARASP, and GRASP reconstructions were evaluated by 2 radiologists in a blinded fashion. The readers independently assessed quality of enhancement (QE), overall image quality (IQ), and other parameters of image quality on a 5-point scale, with the highest score indicating the most desirable examination. Mixed model analysis of variance was used to compare each measure of image quality. Results: Images of BH-VIBE and GRASP had significantly higher QE and IQ values compared with IARASP for both phases (P < 0.05). The differences in QE between BH-VIBE and GRASP for the arterial and venous phases were not significant (P 9 0.05). Although GRASP had lower IQ score compared with BH-VIBE for the arterial (3.9 vs 4.8; P < 0.0001) and venous (4.2 vs 4.8; P = 0.005) phases, GRASP received IQ scores of 3 or more in all participants, which was consistent with acceptable or better diagnostic image quality. Conclusion: Contrast-enhanced multiphase liver MRI of diagnostic quality can be performed during free breathing using a combination of compressed sensing, parallel imaging, and golden-angle radial sampling.
引用
收藏
页码:10 / 16
页数:7
相关论文
共 19 条
  • [1] Free-Breathing 3D T1-Weighted Gradient-Echo Sequence With Radial Data Sampling in Abdominal MRI: Preliminary Observations
    Azevedo, Rafael M.
    de Campos, Rafael O. P.
    Ramalho, Miguel
    Heredia, Vasco
    Dale, Brian M.
    Semelka, Richard C.
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 2011, 197 (03) : 650 - 657
  • [2] Undersampled radial MRI with multiple coils. Iterative image reconstruction using a total variation constraint
    Block, Kai Tobias
    Uecker, Martin
    Frahm, Jens
    [J]. MAGNETIC RESONANCE IN MEDICINE, 2007, 57 (06) : 1086 - 1098
  • [3] Chandarana H, 2012, P 20 ANN M ISMRM MEL
  • [4] Free-Breathing Radial 3D Fat-Suppressed T1-Weighted Gradient Echo Sequence A Viable Alternative for Contrast-Enhanced Liver Imaging in Patients Unable to Suspend Respiration
    Chandarana, Hersh
    Block, Tobias K.
    Rosenkrantz, Andrew B.
    Lim, Ruth P.
    Kim, Danny
    Mossa, David J.
    Babb, James S.
    Kiefer, Berthold
    Lee, Vivian S.
    [J]. INVESTIGATIVE RADIOLOGY, 2011, 46 (10) : 648 - 653
  • [5] Focal hepatic lesions: Diagnostic value of enhancement pattern approach with contrast-enhanced 3D gradient-echo MR imaging
    Elsayes, KM
    Narra, VR
    Yin, YM
    Mukundan, G
    Lammle, M
    Brown, JJ
    [J]. RADIOGRAPHICS, 2005, 25 (05) : 1299 - 1320
  • [6] Generalized Autocalibrating Partially Parallel Acquisitions (GRAPPA)
    Griswold, MA
    Jakob, PM
    Heidemann, RM
    Nittka, M
    Jellus, V
    Wang, JM
    Kiefer, B
    Haase, A
    [J]. MAGNETIC RESONANCE IN MEDICINE, 2002, 47 (06) : 1202 - 1210
  • [7] Respiratory Motion-Compensated Radial Dynamic Contrast-Enhanced (DCE)-MRI of Chest and Abdominal Lesions
    Lin, Wei
    Guo, Junyu
    Rosen, Mark A.
    Song, Hee Kwon
    [J]. MAGNETIC RESONANCE IN MEDICINE, 2008, 60 (05) : 1135 - 1146
  • [8] Sparse MRI: The application of compressed sensing for rapid MR imaging
    Lustig, Michael
    Donoho, David
    Pauly, John M.
    [J]. MAGNETIC RESONANCE IN MEDICINE, 2007, 58 (06) : 1182 - 1195
  • [9] The effects of incomplete breath-holding on 3D MR image quality
    Maki, JH
    Chenevert, TL
    Prince, MR
    [J]. JMRI-JOURNAL OF MAGNETIC RESONANCE IMAGING, 1997, 7 (06): : 1132 - 1139
  • [10] Combination of Compressed Sensing and Parallel Imaging for Highly Accelerated First-Pass Cardiac Perfusion MRI
    Otazo, Ricardo
    Kim, Daniel
    Axel, Leon
    Sodickson, Daniel K.
    [J]. MAGNETIC RESONANCE IN MEDICINE, 2010, 64 (03) : 767 - 776