Dose optimization for dual-energy contrast-enhanced digital mammography based on an energy-resolved photon-counting detector: A Monte Carlo simulation study

被引:5
作者
Lee, Youngjin [1 ]
Lee, Seungwan [2 ,3 ]
Kang, Sooncheol [3 ]
Eom, Jisoo [2 ]
机构
[1] Eulji Univ, Coll Hlth Sci, Dept Radiol Sci, 553 Sanseong Daero, Seongnam 13135, Gyeonggi Do, South Korea
[2] Konyang Univ, Coll Med Sci, Dept Radiol Sci, 158 Gwanjeodong Ro, Daejeon 35365, South Korea
[3] Konyang Univ, Dept Med Sci, 158 Gwanjeodong Ro, Daejeon 35365, South Korea
基金
新加坡国家研究基金会;
关键词
Dual-energy imaging; Photon-counting detector; Contrast-enhanced digital mammography; Radiation dose; Monte Carlo simulation; ADDITIONAL SPECTRAL FILTRATION; SOURCE CT; COMPUTED-TOMOGRAPHY; FEASIBILITY; NOISE; DIFFERENTIATION; QUANTIFICATION; RADIOGRAPHY; CALIBRATION; TISSUE;
D O I
10.1016/j.radphyschem.2016.11.016
中图分类号
O64 [物理化学(理论化学)、化学物理学];
学科分类号
070304 ; 081704 ;
摘要
Dual-energy contrast-enhanced digital mammography (CEDM) has been used to decompose breast images and improve diagnostic accuracy for tumor detection. However, this technique causes an increase of radiation dose and an inaccuracy in material decomposition due to the limitations of conventional X-ray detectors. In this study, we simulated the dual-energy CEDM with an energy-resolved photon-counting detector (ERPCD) for reducing radiation dose and improving the quantitative accuracy of material decomposition images. The ERPCD-based dual-energy CEDM was compared to the conventional dual-energy CEDM in terms of radiation dose and quantitative accuracy. The correlation between radiation dose and image quality was also evaluated for optimizing the ERPCD-based dual-energy CEDM technique. The results showed that the material decomposition errors of the ERPCD-based dual-energy CEDM were 0.56-0.67 times lower than those of the conventional dual-energy CEDM. The imaging performance of the proposed technique was optimized at the radiation dose of 1.09 mGy, which is a half of the MGD for a single view mammogram. It can be concluded that the ERPCD-based dual-energy CEDM with an optimal exposure level is able to improve the quality of material decomposition images as well as reduce radiation dose.
引用
收藏
页码:46 / 51
页数:6
相关论文
共 50 条
  • [41] Monte Carlo simulation studies for the determination of microcalcification thickness and glandular ratio through dual-energy mammography
    Del Lama, L. S.
    Godeli, J.
    Poletti, M. E.
    RADIATION PHYSICS AND CHEMISTRY, 2017, 137 : 157 - 162
  • [42] Monte Carlo Simulation of an Amorphous Selenium-Based Multi-layer Photon-Counting Detector for SPECT Applications
    Adnani, Sahar
    Pil-Ali, Abdollah
    Scott, Christopher C.
    Karim, Karim S.
    MEDICAL IMAGING 2021: PHYSICS OF MEDICAL IMAGING, 2021, 11595
  • [43] Evaluating noise reduction techniques while considering anatomical noise in dual-energy contrast-enhanced mammography
    Allec, Nicholas
    Abbaszadeh, Shiva
    Scott, Chris C.
    Karim, Karim S.
    Lewin, John M.
    MEDICAL PHYSICS, 2013, 40 (05)
  • [44] Experimental optimization of single-exposure dual-energy angiography with photon-counting x-ray detectors
    Aubert, Sarah
    Tanguay, Jesse
    MEDICAL PHYSICS, 2023, 50 (02) : 763 - 777
  • [45] Radiation dose efficiency of multi-energy photon-counting-detector CT for dual-contrast imaging
    Ren, Liqiang
    Rajendran, Kishore
    McCollough, Cynthia H.
    Yu, Lifeng
    PHYSICS IN MEDICINE AND BIOLOGY, 2019, 64 (24)
  • [46] Performance improvements of virtual monoenergetic images in photon-counting detector CT compared with dual source dual-energy CT: Fourier-based assessment
    Kawashima, Hiroki
    Ichikawa, Katsuhiro
    Yoshida, Ryoichi
    Katayama, Takuto
    Arimoto, Makoto
    Kataoka, Jun
    Nagata, Hiroji
    Kobayashi, Satoshi
    PHYSICAL AND ENGINEERING SCIENCES IN MEDICINE, 2024, : 143 - 153
  • [47] Projection-based energy weighting on photon-counting X-ray images in digital subtraction mammography: a feasibility study
    Choi, Sung-Hoon
    Lee, Seung-Wan
    Choi, Yu-Na
    Lee, Young-Jin
    Kim, Hee-Joung
    MEDICAL IMAGING 2014: PHYSICS OF MEDICAL IMAGING, 2014, 9033
  • [48] Image quality and dose exposure of contrast-enhanced abdominal CT on a 1st generation clinical dual-source photon-counting detector CT in obese patients vs. a 2nd generation dual-source dual energy integrating detector CT
    Hagen, Florian
    Hofmann, Johannes
    Wrazidlo, Robin
    Gutjahr, Ralf
    Schmidt, Bernhard
    Faby, Sebastian
    Nikolaou, Konstantin
    Horger, Marius
    EUROPEAN JOURNAL OF RADIOLOGY, 2022, 151
  • [49] Contrast and quantum noise in single-exposure dual-energy thoracic imaging with photon-counting x-ray detectors
    Dhari, Jeffrey
    Tanguay, Jesse
    PHYSICS IN MEDICINE AND BIOLOGY, 2024, 69 (19)
  • [50] Experimental characterization of a direct conversion amorphous selenium detector with thicker conversion layer for dual-energy contrast-enhanced breast imaging
    Scaduto, David A.
    Tousignant, Olivier
    Zhao, Wei
    MEDICAL PHYSICS, 2017, 44 (08) : 3965 - 3977