Dose optimization for dual-energy contrast-enhanced digital mammography based on an energy-resolved photon-counting detector: A Monte Carlo simulation study

被引:5
作者
Lee, Youngjin [1 ]
Lee, Seungwan [2 ,3 ]
Kang, Sooncheol [3 ]
Eom, Jisoo [2 ]
机构
[1] Eulji Univ, Coll Hlth Sci, Dept Radiol Sci, 553 Sanseong Daero, Seongnam 13135, Gyeonggi Do, South Korea
[2] Konyang Univ, Coll Med Sci, Dept Radiol Sci, 158 Gwanjeodong Ro, Daejeon 35365, South Korea
[3] Konyang Univ, Dept Med Sci, 158 Gwanjeodong Ro, Daejeon 35365, South Korea
基金
新加坡国家研究基金会;
关键词
Dual-energy imaging; Photon-counting detector; Contrast-enhanced digital mammography; Radiation dose; Monte Carlo simulation; ADDITIONAL SPECTRAL FILTRATION; SOURCE CT; COMPUTED-TOMOGRAPHY; FEASIBILITY; NOISE; DIFFERENTIATION; QUANTIFICATION; RADIOGRAPHY; CALIBRATION; TISSUE;
D O I
10.1016/j.radphyschem.2016.11.016
中图分类号
O64 [物理化学(理论化学)、化学物理学];
学科分类号
070304 ; 081704 ;
摘要
Dual-energy contrast-enhanced digital mammography (CEDM) has been used to decompose breast images and improve diagnostic accuracy for tumor detection. However, this technique causes an increase of radiation dose and an inaccuracy in material decomposition due to the limitations of conventional X-ray detectors. In this study, we simulated the dual-energy CEDM with an energy-resolved photon-counting detector (ERPCD) for reducing radiation dose and improving the quantitative accuracy of material decomposition images. The ERPCD-based dual-energy CEDM was compared to the conventional dual-energy CEDM in terms of radiation dose and quantitative accuracy. The correlation between radiation dose and image quality was also evaluated for optimizing the ERPCD-based dual-energy CEDM technique. The results showed that the material decomposition errors of the ERPCD-based dual-energy CEDM were 0.56-0.67 times lower than those of the conventional dual-energy CEDM. The imaging performance of the proposed technique was optimized at the radiation dose of 1.09 mGy, which is a half of the MGD for a single view mammogram. It can be concluded that the ERPCD-based dual-energy CEDM with an optimal exposure level is able to improve the quality of material decomposition images as well as reduce radiation dose.
引用
收藏
页码:46 / 51
页数:6
相关论文
共 50 条
  • [31] Evaluation of gold K-edge imaging using spectral computed tomography with a photon-counting detector: A Monte Carlo simulation study
    Kang, Sooncheol
    Eom, Jisoo
    Kim, Burnyoung
    Lee, Seungwan
    OPTIK, 2017, 140 : 253 - 260
  • [32] Monte Carlo Simulation of X-Ray Spectra in Mammography and Contrast-Enhanced Digital Mammography Using the Code PENELOPE
    Cunha, Diego M.
    Tomal, Alessandra
    Poletti, Martin E.
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2013, 60 (02) : 495 - 502
  • [33] Investigation of test methods for QC in dual-energy based contrast-enhanced digital mammography systems: II. Artefacts/uniformity, exposure time and phantom-based dosimetry
    Marshall, N. W.
    Cockmartin, L.
    Bosmans, H.
    PHYSICS IN MEDICINE AND BIOLOGY, 2023, 68 (21)
  • [34] Contrast-Enhanced Abdominal CT with Clinical Photon-Counting Detector CT: Assessment of Image Quality and Comparison with Energy-Integrating Detector CT
    Higashigaito, Kai
    Euler, Andre
    Eberhard, Matthias
    Flohr, Thomas G.
    Schmidt, Bernhard
    Alkadhi, Hatem
    ACADEMIC RADIOLOGY, 2022, 29 (05) : 689 - 697
  • [35] Application of BI-RADS Descriptors in Contrast-Enhanced Dual-Energy Mammography: Comparison with MRI
    Knogler, Thomas
    Homolka, Peter
    Hoernig, Mathias
    Leithner, Robert
    Langs, Georg
    Waitzbauer, Martin
    Pinker, Katja
    Leitner, Sabine
    Helbich, Thomas H.
    BREAST CARE, 2017, 12 (04) : 212 - 216
  • [36] Organ dose in cardiac dual-energy computed tomography: a Monte Carlo study
    Delaram Pakravan
    Farshid Babapour Mofrad
    Mohammad Reza Deevband
    Mahdi Ghorbani
    Hamidreza Pouraliakbar
    Physical and Engineering Sciences in Medicine, 2022, 45 : 157 - 166
  • [37] Organ dose in cardiac dual-energy computed tomography: a Monte Carlo study
    Pakravan, Delaram
    Mofrad, Farshid Babapour
    Deevband, Mohammad Reza
    Ghorbani, Mahdi
    Pouraliakbar, Hamidreza
    PHYSICAL AND ENGINEERING SCIENCES IN MEDICINE, 2022, 45 (01) : 157 - 166
  • [38] Dual-energy approach to contrast-enhanced mammography using the balanced filter method: Spectral optimization and preliminary phantom measurement
    Saito, Masatoshi
    MEDICAL PHYSICS, 2007, 34 (11) : 4236 - 4246
  • [39] A single detector energy-resolved proton radiography system: a proof of principle study by Monte Carlo simulations
    Huo, Wanli
    Zwart, Townsend
    Cooley, James
    Huang, Kai
    Finley, Caitlin
    Jee, Kyung-Wook
    Sharp, Gregory C.
    Rosenthal, Stanley
    Xu, X. George
    Lu, Hsiao-Ming
    PHYSICS IN MEDICINE AND BIOLOGY, 2019, 64 (02)
  • [40] Dual-energy contrast enhanced digital mammography using a new approach for breast tissue canceling
    Puong, Sylvie
    Bouchevreau, Xavier
    Patoureaux, Fanny
    Iordache, Razvan
    Muller, Serge
    MEDICAL IMAGING 2007: PHYSICS OF MEDICAL IMAGING, PTS 1-3, 2007, 6510