The new Roche Vitamin D Total assay: fit for its purpose ?

被引:32
作者
Emmen, Judith M. A. [1 ]
Wielders, Jos P. M. [2 ]
Boer, Arjen-Kars [3 ]
van den Ouweland, Johannes M. W. [4 ]
Vader, Huib L. [1 ]
机构
[1] Maxima Med Ctr, Clin Chem Lab, NL-5500 MB Veldhoven, Netherlands
[2] Meander Med Ctr, Dept Clin Chem, Amersfoort, Netherlands
[3] Catharina Hosp, Clin Lab, Eindhoven, Netherlands
[4] Canisius Wilhelmina Hosp, Dept Clin Chem, Nijmegen, Netherlands
关键词
25-hydroxyvitamin D; HPLC; immunoassay; liquid chromatography-tandem mass spectrometry (LC-MS/MS); method validation; TANDEM MASS-SPECTROMETRY; 25-HYDROXYVITAMIN D; ACCURACY;
D O I
10.1515/cclm-2011-0085
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Background: Measurement of serum 25-hydroxyvitamin D [25(OH)D] is used to assess vitamin D status. We evaluated the analytical performance of a new automated assay, Elecsys Vitamin D Total (Roche Diagnostics, Mannheim, Germany), based on competitive protein binding. Methods: The Elecsys assay was tested for imprecision, linearity and functional sensitivity at three test-sites and compared to a liquid chromatography-tandem mass spectrometry (LC-MS/MS) method, a high-performance liquid chromatography (HPLC) method and the Liaison 25(OH) Vitamin D Total immunoassay (Diasorin). Results: Imprecision testing with human serum specimens showed within-run CVs of <= 6% and between-run CVs of <= 8%. The assay was linear from 33 up to at least 111 nmol/L and showed equivalent 25(OH) D levels for matched serum and heparinized plasma samples. The assay correlated reasonable to well with LC-MS/MS (r=0.93; y=1.07x-5.04 nmol/L), HPLC (r=0.91, y=0.90x+3.03 nmol/L) and the Liaison assay (r=0.86, y=1.19x+2.80 nmol/L). Some of the samples showed large between-method differences. Conclusions: The new Elecsys assay fulfilled present analytical performance requirements and showed close agreement to other well-established methods for 25(OH) D analysis, making it fit for routine assessment of vitamin D status.
引用
收藏
页码:1969 / 1972
页数:4
相关论文
共 12 条
  • [1] Current status of clinical 25-hydroxyvitamin D measurement: An assessment of between-laboratory agreement
    Binkley, Neil
    Krueger, Diane C.
    Morgan, Sarah
    Wiebe, Donald
    [J]. CLINICA CHIMICA ACTA, 2010, 411 (23-24) : 1976 - 1982
  • [2] Carter GD, 2011, CURR DRUG TARGETS, V12, P19
  • [3] Overreporting of vitamin D deficiency with the Roche Elecsys Vitamin D3 (25-OH) method
    Connell, A. B.
    Jenkins, N.
    Black, M.
    Pasco, J. A.
    Kotowicz, M. A.
    Schneider, H-G.
    [J]. PATHOLOGY, 2011, 43 (04) : 368 - 371
  • [4] State-of-the-Art Vitamin D Assays: A Comparison of Automated Immunoassays with Liquid Chromatography-Tandem Mass Spectrometry Methods
    Farrell, Christopher-John L.
    Martin, Steven
    McWhinney, Brett
    Straub, Isabella
    Williams, Paul
    Herrmann, Markus
    [J]. CLINICAL CHEMISTRY, 2012, 58 (03) : 531 - 542
  • [5] Accuracy of 6 Routine 25-Hydroxyvitamin D Assays: Influence of Vitamin D Binding Protein Concentration
    Heijboer, Annemieke C.
    Blankenstein, Marinus A.
    Kema, Ido P.
    Buijs, Madelon M.
    [J]. CLINICAL CHEMISTRY, 2012, 58 (03) : 543 - 548
  • [6] Vitamin D deficiency
    Holick, Michael F.
    [J]. NEW ENGLAND JOURNAL OF MEDICINE, 2007, 357 (03) : 266 - 281
  • [7] Automated Measurement of 25-OH Vitamin D3 on the Roche Modular E170 Analyzer
    Leino, Aila
    Turpeinen, Ursula
    Koskinen, Pertti
    [J]. CLINICAL CHEMISTRY, 2008, 54 (12) : 2059 - 2062
  • [8] The 2011 Report on Dietary Reference Intakes for Calcium and Vitamin D from the Institute of Medicine: What Clinicians Need to Know
    Ross, A. Catharine
    Manson, JoAnn E.
    Abrams, Steven A.
    Aloia, John F.
    Brannon, Patsy M.
    Clinton, Steven K.
    Durazo-Arvizu, Ramon A.
    Gallagher, J. Christopher
    Gallo, Richard L.
    Jones, Glenville
    Kovacs, Christopher S.
    Mayne, Susan T.
    Rosen, Clifford J.
    Shapses, Sue A.
    [J]. JOURNAL OF CLINICAL ENDOCRINOLOGY & METABOLISM, 2011, 96 (01) : 53 - 58
  • [9] Specifications for trueness and precision of a reference measurement system for serum/plasma 25-hydroxyvitamin D analysis
    Stoeckl, Dietmar
    Sluss, Patrick M.
    Thienpont, Linda M.
    [J]. CLINICA CHIMICA ACTA, 2009, 408 (1-2) : 8 - 13
  • [10] Measurement of 25-OH-vitamin D in human serum using liquid chromatography tandem-mass spectrometry with comparison to radioimmunoassay and automated immunoassay
    van den Ouweland, Johannes M. W.
    Beijers, Antonius M.
    Demacker, Pierre N. M.
    van Daal, Henny
    [J]. JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2010, 878 (15-16): : 1163 - 1168