Rise in chronic defibrillation energy requirements necessitating implantable defibrillator lead system revision

被引:7
|
作者
Daoud, EG [1 ]
Man, KC [1 ]
Morady, F [1 ]
Strickberger, SA [1 ]
机构
[1] UNIV MICHIGAN HOSP,DEPT INTERNAL MED,DIV CARDIOL,ANN ARBOR,MI 48109
来源
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY | 1997年 / 20卷 / 03期
关键词
defibrillation energy; implantable defibrillator;
D O I
10.1111/j.1540-8159.1997.tb03890.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
The chronic defibrillation energy requirement (DER) is believed to remain clinically stable in patients with defibrillators. Six patients (two with an epicardial and four with a nonthoracotomy system) were identified with a rise in their chronic DER, which eliminated a 10-J safety margin, thus necessitating a defibrillator lead system revision. The mean increase in DER was 14.7 +/- 4J and was discovered at a mean of 16.0 +/- 18 months (range 2-41) following implantation. Management included placement of a defibrillator with a biphasic waveform, placement of an additional defibrillation electrode, or both. At 2 months following revision of the defibrillation system, a 10-J DER safety margin was present in each patient. In some patients, there is a progressive increase in the chronic DER with elimination of a 10-J safety margin necessitating revision of the defibrillation system. Routine reevaluation of the chronic DER, therefore, is necessary to identify these patients.
引用
收藏
页码:714 / 719
页数:6
相关论文
共 50 条
  • [41] His bundle pacing implantable cardiac defibrillator with defibrillation lead placement in the middle cardiac vein: a case report
    Bisson, Arnaud
    Bodin, Alexandre
    Clementy, Nicolas
    EUROPACE, 2021, 23 (04):
  • [42] High defibrillation threshold with a subcutaneous implantable cardiac defibrillator due to the lead having been positioned in the fat layer
    Hirao, Tatsuhiko
    Nitta, Junichi
    Sato, Akira
    Takahashi, Yoshihide
    Goya, Masahiko
    Hirao, Kenzo
    JOURNAL OF ARRHYTHMIA, 2018, 34 (02) : 198 - 200
  • [43] TRANSVENOUS-SUBCUTANEOUS LEAD SYSTEM FOR IMPLANTABLE CARDIOVERTER-DEFIBRILLATOR
    HIEF, C
    PODCZECK, A
    VEIT, F
    FROHNER, K
    KALTENBRUNNER, W
    NURNBERG, M
    STEINBACH, K
    WIENER KLINISCHE WOCHENSCHRIFT, 1993, 105 (01) : 12 - 16
  • [44] Compatibility of a nonthoracotomy lead system with a biphasic implantable cardioverter-defibrillator
    Porterfield, JG
    Porterfield, LM
    Levine, JH
    Luceri, RM
    Hsia, HH
    AbiSamra, F
    Alvarez, L
    Buxton, AE
    Camm, J
    Cutler, JE
    Fogoros, RN
    Hsia, HH
    Janosik, DL
    Kelly, P
    Levine, JH
    Luceri, RM
    Mann, D
    Meesmann, M
    McCowan, R
    Nathan, A
    Porterfield, JG
    Rothbart, ST
    Steinberg, JS
    Torres, VI
    Trappe, HJ
    Winters, SL
    AMERICAN JOURNAL OF CARDIOLOGY, 1996, 77 (08): : 586 - 590
  • [45] TRANSVENOUS IMPLANTABLE CARDIOVASCULAR DEFIBRILLATOR LEAD SYSTEM - EXPERIENCE IN 28 PATIENTS
    MATSA, M
    COOPERMAN, I
    VISKIN, S
    GLICK, A
    GUREVITCH, Y
    YAKIREVICH, V
    BELHASSEN, B
    JOURNAL OF CARDIOVASCULAR SURGERY, 1994, 35 (06): : 207 - 209
  • [46] CHRONIC CHANGES IN DEFIBRILLATION PARAMETERS WITH A TRANSVENOUS LEAD SYSTEM
    TUMMALA, RV
    RIGGIO, DW
    WEISS, D
    FELICIANO, Z
    SHOROFSKY, SR
    GOLD, MR
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1994, : A112 - A112
  • [47] A PROSPECTIVE RANDOMIZED EVALUATION OF IMPLANTABLE CARDIOVERTER-DEFIBRILLATOR SIZE ON UNIPOLAR DEFIBRILLATION SYSTEM EFFICACY
    JONES, GK
    POOLE, JE
    KUDENCHUK, PJ
    DOLACK, L
    JOHNSON, G
    DEGROOT, P
    GLEVA, MJ
    RAITT, M
    BARDY, GH
    CIRCULATION, 1995, 92 (10) : 2940 - 2943
  • [48] HIGH FAILURE RATE FOR A CHRONIC EPICARDIAL IMPLANTABLE CARDIOVERTER-DEFIBRILLATOR LEAD
    BRADY, P
    FRIEDMAN, PA
    TRUSTY, JM
    GRICE, SE
    HAMMILL, SC
    STANTON, MS
    CIRCULATION, 1995, 92 (08) : 716 - 716
  • [49] A prospective, randomized evaluation of a nonthoracotomy implantable cardioverter defibrillator lead system
    Baker, JH
    Epstein, AE
    VoshageStahl, L
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1997, 20 (01): : 72 - 78
  • [50] Leakage of energy to the body surface during defibrillation shock by an implantable cardioverter-defibrillator (TCD) system - Experimental evaluation during defibrillation shocks through the right ventricular lead and the subcutaneous active-can in canines
    Niwano, S
    Kitano, Y
    Moriguchi, M
    Yoshizawa, N
    Kashiwa, T
    Suyama, M
    Toyoshima, T
    Izumi, T
    JAPANESE CIRCULATION JOURNAL-ENGLISH EDITION, 2001, 65 (03): : 219 - 225