Rise in chronic defibrillation energy requirements necessitating implantable defibrillator lead system revision

被引:7
|
作者
Daoud, EG [1 ]
Man, KC [1 ]
Morady, F [1 ]
Strickberger, SA [1 ]
机构
[1] UNIV MICHIGAN HOSP,DEPT INTERNAL MED,DIV CARDIOL,ANN ARBOR,MI 48109
来源
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY | 1997年 / 20卷 / 03期
关键词
defibrillation energy; implantable defibrillator;
D O I
10.1111/j.1540-8159.1997.tb03890.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
The chronic defibrillation energy requirement (DER) is believed to remain clinically stable in patients with defibrillators. Six patients (two with an epicardial and four with a nonthoracotomy system) were identified with a rise in their chronic DER, which eliminated a 10-J safety margin, thus necessitating a defibrillator lead system revision. The mean increase in DER was 14.7 +/- 4J and was discovered at a mean of 16.0 +/- 18 months (range 2-41) following implantation. Management included placement of a defibrillator with a biphasic waveform, placement of an additional defibrillation electrode, or both. At 2 months following revision of the defibrillation system, a 10-J DER safety margin was present in each patient. In some patients, there is a progressive increase in the chronic DER with elimination of a 10-J safety margin necessitating revision of the defibrillation system. Routine reevaluation of the chronic DER, therefore, is necessary to identify these patients.
引用
收藏
页码:714 / 719
页数:6
相关论文
共 50 条
  • [31] IDENTIFICATION OF AN IMPLANTABLE DEFIBRILLATOR LEAD FRACTURE WITH A NEW HOLTER SYSTEM
    FELDMAN, CL
    OLSON, WH
    HUBBELBANK, M
    MITCHELL, R
    KELSO, D
    BARDY, GH
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1993, 16 (06): : 1342 - 1344
  • [32] Performance of the lead system for the Metrix automatic implantable atrial defibrillator
    Jung, W
    Lüderitz, B
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1998, 31 (02) : 513A - 513A
  • [33] The use of a dual shock lead in an implantable atrial defibrillator system
    Kriplen, M
    Ayers, G
    Reuter, D
    PACEMAKER LEADS 1997, 1997, : 573 - 580
  • [34] Implantable Intravascular Defibrillator: Evaluation of Defibrillation Waveforms with Inferior Vena Cava Electrode System
    Sanders, William E., Jr.
    Malkin, Robert A.
    Richey, Mark W.
    Masson, Stephen C.
    Ransbury, Terrance J.
    Urtz, Mark W.
    Ideker, Raymond E.
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2011, 34 (05): : 577 - 583
  • [35] EFFECTS OF ANTIARRHYTHMIC DRUGS ON EPICARDIAL DEFIBRILLATION ENERGY-REQUIREMENTS AND THE RATE OF DEFIBRILLATOR DISCHARGES
    JUNG, W
    MANZ, M
    PFEIFFER, D
    TEBBENJOHANNS, J
    PIZZULLI, L
    LUDERITZ, B
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1993, 16 (01): : 198 - 201
  • [36] DEFIBRILLATION ENERGY-REQUIREMENTS WITH SINGLE ENDOCARDIAL (ENDOTAK(TM)) LEAD
    WINTER, J
    VESTER, EG
    KUHLS, S
    KANTARTZIS, M
    PERINGS, C
    PAUSCHINGER, M
    STRAUER, BE
    BIRCKS, W
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1993, 16 (03): : 540 - 546
  • [37] CHRONIC ORAL CLOFILIUM DECREASES DEFIBRILLATION ENERGY-REQUIREMENTS
    DORIAN, P
    WANG, MJ
    DAVID, I
    FEINDEL, CM
    CIRCULATION, 1990, 82 (04) : 528 - 528
  • [38] Effect of chronic amiodarone therapy on defibrillation energy requirements in humans
    Pelosi, F
    Oral, H
    Kim, MH
    Sticherling, C
    Horwood, L
    Knight, BP
    Michaud, GF
    Morady, F
    Strickberger, SA
    JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2000, 11 (07) : 736 - 740
  • [39] Implantable cardioverter-defibrillator lead revision following left ventricular assist device implantation
    Black-Maier, Eric
    Lewis, Robert K.
    Rehorn, Michael
    Loungani, Rahul
    Friedman, Daniel J.
    Frazier-Mills, Camille
    Jackson, Kevin P.
    Atwater, Brett D.
    Milano, Carmelo A.
    Schroder, Jacob N.
    Pokorney, Sean D.
    Piccini, Jonathan P.
    JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2020, 31 (06) : 1509 - 1518