Evaluation of Acute Cardiac and Chest Wall Damage after Shocks with a Subcutaneous Implantable Cardioverter Defibrillator in Swine

被引:27
|
作者
Killingsworth, Cheryl R. [1 ]
Melnick, Sharon B. [1 ]
Litovsky, Silvio H. [2 ]
Ideker, Raymond E. [1 ,3 ,4 ]
Walcott, Gregory P. [1 ,3 ]
机构
[1] Univ Alabama Birmingham, Dept Med, Div Cardiovasc Dis, Cardiac Rhythm Management Lab, Birmingham, AL 35294 USA
[2] Univ Alabama Birmingham, Dept Pathol, Birmingham, AL 35294 USA
[3] Univ Alabama Birmingham, Dept Biomed Engn, Birmingham, AL 35294 USA
[4] Univ Alabama Birmingham, Dept Physiol & Biophys, Birmingham, AL 35294 USA
来源
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY | 2013年 / 36卷 / 10期
关键词
implantable cardioverter defibrillator; troponin I; creatine kinase; inflammation; swine; AUTOMATIC DEFIBRILLATOR; EXTERNAL DEFIBRILLATION; MYOCARDIAL INJURY; TROPONIN-I; ARRHYTHMIAS; PREVENTION;
D O I
10.1111/pace.12173
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
BackgroundA subcutaneous implantable cardioverter defibrillator (S-ICD) could ease placement and reduce complications of transvenous ICDs, but requires more energy than transvenous ICDs. Therefore we assessed cardiac and chest wall damage caused by the maximum energy shocks delivered by both types of clinical devices. MethodsDuring sinus rhythm, anesthetized pigs (38 6 kg) received an S-ICD (n = 4) and five 80-Joule (J) shocks, or a transvenous ICD (control, n = 4) and five 35-J shocks. An inactive S-ICD electrode was implanted into the same control pigs to study implant trauma. All animals survived 24hours. Troponin I and creatine kinase muscle isoenzyme (CK-MM) were measured as indicators of myocardial and skeletal muscle injury. Histopathological injury of heart, lungs, and chest wall was assessed using semiquantitative scoring. ResultsTroponin I was significantly elevated at 4hours and 24hours (22.6 +/- 16.3 ng/mL and 3.1 +/- 1.3 ng/mL; baseline 0.07 +/- 0.09 ng/mL) in control pigs but not in S-ICD pigs (0.12 +/- 0.11 ng/mL and 0.13 +/- 0.13 ng/mL; baseline 0.06 +/- 0.03 ng/mL). CK-MM was significantly elevated in S-ICD pigs after shocks (6,544 +/- 1,496 U/L and 9,705 +/- 6,240 U/L; baseline 704 +/- 398 U/L) but not in controls. Electrocardiogram changes occurred postshock in controls but not in S-ICD pigs. The myocardium and lungs were histologically normal in both groups. Subcutaneous injury was greater in S-ICD compared to controls. ConclusionAlthough CK-MM suggested more skeletal muscle injury in S-ICD pigs, significant cardiac, lung, and chest wall histopathological changes were not detected in either group. Troponin I data indicate significantly less cardiac injury from 80-J S-ICD shocks than 35-J transvenous shocks.
引用
收藏
页码:1265 / 1272
页数:8
相关论文
共 50 条
  • [1] Analysis of troponin I levels after spontaneous implantable cardioverter defibrillator shocks
    Hasdemir, C
    Shah, N
    Rao, AP
    Acosta, H
    Matsudaira, K
    Neas, BR
    Reynolds, DW
    Po, S
    Lazzara, R
    Beckman, KJ
    JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2002, 13 (02) : 144 - 150
  • [2] Survival After Implantable Cardioverter-Defibrillator Shocks
    Aktas, Mehmet K.
    Younis, Arwa
    Zareba, Wojciech
    Kutyifa, Valentina
    Klein, Helmut
    Daubert, James P.
    Estes, Mark
    McNitt, Scott
    Polonsky, Bronislava
    Goldenberg, Ilan
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2021, 77 (20) : 2453 - 2462
  • [3] Uncovering Phantom Shocks in Cardiac Patients with an Implantable Cardioverter Defibrillator
    Bilanovic, Ana
    Irvine, Jane
    Kovacs, Adrienne H.
    Hill, Ann
    Cameron, Doug
    Katz, Joel
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2013, 36 (06): : 673 - 683
  • [4] Development of the subcutaneous implantable cardioverter-defibrillator for reducing sudden cardiac death
    Sanghera, Rick
    Sanders, Richard
    Husby, Michael
    Bentsen, James G.
    PHARMACEUTICAL SCIENCE TO IMPROVE THE HUMAN CONDITION: WINNERS AND FINALISTS FOR THE PRIX GALIEN USA AWARDS 2013, 2014, 1329 : 1 - 17
  • [5] Multiple shocks after upgrade of an implantable cardioverter-defibrillator to a cardiac resynchronization therapy-defibrillator device
    Francisco, Gregory M.
    Sharma, Saumya
    Dougherty, Anne H.
    Kantharia, Bharat K.
    CARDIOLOGY JOURNAL, 2009, 16 (05) : 473 - 476
  • [6] Inappropriate Subcutaneous Implantable Cardioverter Defibrillator Shocks Secondary to Cardiac Remodeling: A Unique Case of T Wave Oversensing
    Saleem, Maleeha
    Pahuja, Karan
    Fatima, Tehreem
    Hamilton, Steven
    Wjasow, Christina
    Fox, Justin
    CUREUS JOURNAL OF MEDICAL SCIENCE, 2022, 14 (06)
  • [7] Inappropriate Shocks With Subcutaneous Implantable Cardioverter-Defibrillator in a Young Patient: A Case Report
    Khan, Zahid
    Sethumadhavan, Dinesh
    Rayner, Tom
    Kyaw, Sithu
    CUREUS JOURNAL OF MEDICAL SCIENCE, 2023, 15 (02)
  • [8] Prognostic evaluation of the troponin I elevation after multiple spontaneous shocks of the implantable cardioverter/defibrillator
    Miranda, Carlos Henrique
    Schmidt, Andre
    Pazin-Filho, Antonio
    AMERICAN JOURNAL OF EMERGENCY MEDICINE, 2014, 32 (09): : 1085 - 1088
  • [9] Subcutaneous implantable cardioverter defibrillator after transvenous lead extraction: safety, efficacy and outcome
    Giacomin, Enrico
    Falzone, Pasquale Valerio
    Dall'Aglio, Pietro Bernardo
    Pittorru, Raimondo
    De Lazzari, Manuel
    Vianello, Riccardo
    Bertaglia, Emanuele
    Tarzia, Vincenzo
    Iliceto, Sabino
    Gerosa, Gino
    Migliore, Federico
    JOURNAL OF INTERVENTIONAL CARDIAC ELECTROPHYSIOLOGY, 2022,
  • [10] Comparison of complications and shocks in paediatric and young transvenous and subcutaneous implantable cardioverter-defibrillator patients
    A. B. E. Quast
    T. F. Brouwer
    K. M. Kooiman
    P. F. H. M van Dessel
    N. A. Blom
    A. A. M. Wilde
    R. E. Knops
    Netherlands Heart Journal, 2018, 26 : 612 - 619