Improving SNR and B1 transmit field for an endorectal coil in 7 T MRI and MRS of prostate cancer

被引:33
作者
de Castro, C. S. Arteaga [1 ,2 ]
van den Bergen, B. [2 ]
Luijten, P. R. [1 ]
van der Heide, U. A. [2 ]
van Vulpen, M. [2 ]
Klomp, D. W. J. [1 ]
机构
[1] Univ Med Ctr Utrecht, Dept Radiol, NL-3508 GA Utrecht, Netherlands
[2] Univ Med Ctr Utrecht, Dept Radiotherapy, NL-3508 GA Utrecht, Netherlands
关键词
endorectal coil; 7; T; spectroscopy; prostate; APPARENT DIFFUSION-COEFFICIENT; CONTRAST-ENHANCED MRI; ADIABATIC PULSES; HUMAN BRAIN; TISSUE; SPECTROSCOPY;
D O I
10.1002/mrm.23200
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Higher magnetic field strengths like 7 T and above are desirable for MR spectroscopy given the increased spectral resolution and signal to noise ratio. At these field strengths, substantial nonuniformities in B1+/- and radiofrequency power deposition become apparent. In this investigation, we propose an improvement on a conventionally used endorectal coil, through the addition of a second element (stripline). Both elements are used as transceivers. In the center of the prostate, approximately 40% signal to noise ratio increase is achieved. In fact, the signal to noise ratio gain obtained with the quadrature configuration locally can be even greater than 40% when compared to the single loop configuration. This is due to the natural asymmetry of the B1+/- fields at high frequencies, which causes destructive and constructive interference patterns. Global specific absorption rate is reduced by almost a factor of 2 as expected. Furthermore, approximately a 4-fold decrease in local specific absorption rate is observed when normalized to the B1 values in the center of the prostate. Because of the 4-fold local specific absorption rate decrease obtained with the dual channel setup for the same reference B1 value (20 mu T at 3.5 cm depth into the prostate) as compared to the single loop, the transmission power B1 duty cycle can be increased by a factor 4. Consequently, when using the two-element endorectal coil, the radiofrequency power deposition is significantly reduced and radiofrequency intense sequences with adiabatic pulses can be safely applied at 7 T for 1H magnetic resonance spectroscopy and MRI in the prostate. Altogether, in vivo 1H magnetic resonance spectroscopic imaging of prostate cancer with a fully adiabatic sequence operated at a minimum B1+ of 20 mu T shows insensitivity to the nonuniform transmit field, while remaining within local specific absorption rate guidelines of 10 W/kg. Magn Reson Med, 2012. (C) 2011 Wiley Periodicals, Inc.
引用
收藏
页码:311 / 318
页数:8
相关论文
共 31 条
[1]   Tissue-equivalent phantoms for high frequencies [J].
Beck, BL ;
Jenkins, KA ;
Rocca, JR ;
Fitzsimmons, JR .
CONCEPTS IN MAGNETIC RESONANCE PART B-MAGNETIC RESONANCE ENGINEERING, 2004, 20B (01) :30-33
[2]   B1+ Interferometry for the Calibration of RF Transmitter Arrays [J].
Brunner, David O. ;
Pruessmann, Klaas P. .
MAGNETIC RESONANCE IN MEDICINE, 2009, 61 (06) :1480-1488
[3]  
DeGraaf RA, 1997, CONCEPT MAGNETIC RES, V9, P247, DOI 10.1002/(SICI)1099-0534(1997)9:4<247::AID-CMR4>3.0.CO
[4]  
2-Z
[5]   Design of internal MRI coils using ultimate intrinsic SNR [J].
Eryaman, Yigitcan ;
Oner, Yusuf ;
Atalar, Ergin .
MAGNETIC RESONANCE MATERIALS IN PHYSICS BIOLOGY AND MEDICINE, 2009, 22 (04) :221-228
[6]   The return of the frequency sweep: Designing adiabatic pulses for contemporary NMR [J].
Garwood, M ;
DelaBarre, L .
JOURNAL OF MAGNETIC RESONANCE, 2001, 153 (02) :155-177
[7]   Endorectal diffusion-weighted imaging in prostate cancer to differentiate malignant and benign peripheral zone tissue [J].
Hosseinzadeh, K ;
Schwarz, SD .
JOURNAL OF MAGNETIC RESONANCE IMAGING, 2004, 20 (04) :654-661
[8]  
Hoult DI, 2000, J MAGN RESON IMAGING, V12, P46, DOI 10.1002/1522-2586(200007)12:1<46::AID-JMRI6>3.0.CO
[9]  
2-D
[10]   In vivo measurement of the apparent diffusion coefficient in normal and malignant prostatic tissues using echo-planar imaging [J].
Issa, B .
JOURNAL OF MAGNETIC RESONANCE IMAGING, 2002, 16 (02) :196-200