Quantitative Analysis Tools and Digital Phantoms for Deformable Image Registration Quality Assurance

被引:9
作者
Kim, Haksoo [1 ]
Park, Samuel B. [2 ]
Monroe, James I. [1 ,3 ]
Traughber, Bryan J. [1 ,4 ]
Zheng, Yiran [1 ,4 ]
Lo, Simon S. [1 ,4 ]
Yao, Min [1 ,4 ]
Mansur, David [1 ,4 ]
Ellis, Rodney [1 ,4 ]
Machtay, Mitchell [1 ,4 ]
Sohn, Jason W. [1 ,4 ]
机构
[1] Case Western Reserve Univ, Sch Med, Dept Radiat Oncol, Cleveland, OH 44106 USA
[2] Natl Canc Ctr, Goyang Si, Gyeonggi Do, South Korea
[3] St Anthonys Med Ctr, St Louis, MO USA
[4] Univ Hosp Cleveland, Cleveland, OH 44106 USA
关键词
deformable image registration; quality assurance; FREE-FORM DEFORMATIONS; RADIATION-THERAPY; NONRIGID REGISTRATION; MUTUAL-INFORMATION; DEMONS ALGORITHM; MEDICAL IMAGES; B-SPLINES; MR-IMAGES; ACCURACY; OPTIMIZATION;
D O I
10.1177/1533034614553891
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
This article proposes quantitative analysis tools and digital phantoms to quantify intrinsic errors of deformable image registration (DIR) systems and establish quality assurance (QA) procedures for clinical use of DIR systems utilizing local and global error analysis methods with clinically realistic digital image phantoms. Landmark-based image registration verifications are suitable only for images with significant feature points. To address this shortfall, we adapted a deformation vector field (DVF) comparison approach with new analysis techniques to quantify the results. Digital image phantoms are derived from data sets of actual patient images (a reference image set, R, a test image set, T). Image sets from the same patient taken at different times are registered with deformable methods producing a reference DVFref. Applying DVFref to the original reference image deforms T into a new image R. The data set, R, T, and DVFref, is from a realistic truth set and therefore can be used to analyze any DIR system and expose intrinsic errors by comparing DVFref and DVFtest. For quantitative error analysis, calculating and delineating differences between DVFs, 2 methods were used, (1) a local error analysis tool that displays deformation error magnitudes with color mapping on each image slice and (2) a global error analysis tool that calculates a deformation error histogram, which describes a cumulative probability function of errors for each anatomical structure. Three digital image phantoms were generated from three patients with a head and neck, a lung and a liver cancer. The DIR QA was evaluated using the case with head and neck.
引用
收藏
页码:428 / 439
页数:12
相关论文
共 56 条
  • [1] Non-rigid registration using free-form deformation for prostate images
    Abolhassani, N
    Samani, A
    [J]. NAFIPS 2005 - 2005 ANNUAL MEETING OF THE NORTH AMERICAN FUZZY INFORMATION PROCESSING SOCIETY, 2005, : 51 - 54
  • [2] 4D-CT lung motion estimation with deformable registration: Quantification of motion nonlinearity and hysteresis
    Boldea, Vlad
    Sharp, Gregory C.
    Jiang, Steve B.
    Sarrut, David
    [J]. MEDICAL PHYSICS, 2008, 35 (03) : 1008 - 1018
  • [3] Registration of real and CT-derived virtual bronchoscopic images to assist transbronchial biopsy
    Bricault, I
    Ferretti, G
    Cinquin, P
    [J]. IEEE TRANSACTIONS ON MEDICAL IMAGING, 1998, 17 (05) : 703 - 714
  • [4] RESULTS OF A MULTI-INSTITUTION DEFORMABLE REGISTRATION ACCURACY STUDY (MIDRAS)
    Brock, Kristy K.
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2010, 76 (02): : 583 - 596
  • [5] A framework for evaluation of deformable image registration spatial accuracy using large landmark point sets
    Castillo, Richard
    Castillo, Edward
    Guerra, Rudy
    Johnson, Valen E.
    McPhail, Travis
    Garg, Amit K.
    Guerrero, Thomas
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2009, 54 (07) : 1849 - 1870
  • [6] Chiang MC, 2006, I S BIOMED IMAGING, P193
  • [7] Tracking lung tissue motion and expansion/compression with inverse consistent image registration and spirometry
    Christensen, Gary E.
    Song, Joo Hyun
    Lu, Wei
    El Naqa, Issarn
    Low, Daniel A.
    [J]. MEDICAL PHYSICS, 2007, 34 (06) : 2155 - 2163
  • [8] Evaluating commercial image registration packages for radiotherapy treatment planning
    Dalah, Entesar Z.
    Nisbet, A.
    Reise, S.
    Bradley, D.
    [J]. APPLIED RADIATION AND ISOTOPES, 2008, 66 (12) : 1948 - 1953
  • [9] DOSE-VOLUME HISTOGRAMS
    DRZYMALA, RE
    MOHAN, R
    BREWSTER, L
    CHU, J
    GOITEIN, M
    HARMS, W
    URIE, M
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1991, 21 (01): : 71 - 78
  • [10] Registration of 3-D intraoperative MR images of the brain using a finite-element biomechanical model
    Ferrant, M
    Nabavi, A
    Macq, B
    Jolesz, FA
    Kikinis, R
    Warfield, SK
    [J]. IEEE TRANSACTIONS ON MEDICAL IMAGING, 2001, 20 (12) : 1384 - 1397