Head-to-Head Comparison of Global Longitudinal Strain Measurements among Nine Different Vendors The EACVI/ASE Inter-Vendor Comparison Study

被引:493
作者
Farsalinos, Konstantinos E. [1 ]
Daraban, Ana M. [1 ]
Unlu, Serkan [1 ]
Thomas, James D. [2 ]
Badano, Luigi P. [3 ]
Voigt, Jens-Uwe [1 ]
机构
[1] Katholieke Univ Leuven, Univ Hosp Gasthuisberg, B-3000 Leuven, Belgium
[2] Northwestern Univ, Bluhm Cardiovasc Inst, Chicago, IL 60611 USA
[3] Univ Padua, Dept Cardiac Thorac & Vasc Sci, Padua, Italy
关键词
Deformation; Strain; Left ventricle; Systolic function; Speckle-tracking echocardiography; SPECKLE TRACKING ECHOCARDIOGRAPHY; EARLY MYOCARDIAL DYSFUNCTION; 2-DIMENSIONAL STRAIN; DEFORMATION; VARIABILITY; REPRODUCIBILITY; LOAD;
D O I
10.1016/j.echo.2015.06.011
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: This study was planned by the EACVI/ASE/Industry Task Force to Standardize Deformation Imaging to (1) test the variability of speckle-tracking global longitudinal strain(GLS) measurements among different vendors and (2) compare GLS measurement variability with conventional echocardiographic parameters. Methods: Sixty-two volunteers were studied using ultrasound systems from seven manufacturers. Each volunteer was examined by the same sonographer on all machines. Inter- and intraobserver variability was determined in a true test-retest setting. Conventional echocardiographic parameters were acquired for comparison. Using the software packages of the respective manufacturer and of two software-only vendors, endocardial GLS was measured because it was the only GLS parameter that could be provided by all manufactures. We compared GLS(AV) (the average from the three apical views) and GLS4(CH) (measured in the four-chamber view) measurements among vendors and with the conventional echocardiographic parameters. Results: Absolute values of GLS(AV) ranged from 18.0% to 21.5%, while GLS4(CH) ranged from 17.9% to 21.4%. The absolute difference between vendors for GLS(AV) was up to 3.7% strain units (P < .001). The interobserver relative mean errors were 5.4% to 8.6% for GLS(AV) and 6.2% to 11.0% for GLS4(CH), while the intraobserver relative mean errors were 4.9% to 7.3% and 7.2% to 11.3%, respectively. These errors were lower than for left ventricular ejection fraction and most other conventional echocardiographic parameters. Conclusion: Reproducibility of GLS measurements was good and in many cases superior to conventional echocardiographic measurements. The small but statistically significant variation among vendors should be considered in performing serial studies and reflects a reference point for ongoing standardization efforts.
引用
收藏
页码:1171 / +
页数:13
相关论文
共 38 条
  • [1] Noninvasive myocardial strain measurement by speckle tracking echocardiography - Validation against sonomicrometry and tagged magnetic resonance imaging
    Amundsen, BH
    Helle-Valle, T
    Edvardsen, T
    Torp, H
    Crosby, J
    Lyseggen, E
    Stoylen, A
    Ihlen, H
    Lima, JAC
    Smiseth, OA
    Slordahl, SA
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2006, 47 (04) : 789 - 793
  • [2] [Anonymous], HEART
  • [3] Use of myocardial strain to assess global left ventricular function: A comparison with cardiac magnetic resonance and 3-dimensional echocardiography
    Brown, Joseph
    Jerkins, Carly
    Marwick, Thomas H.
    [J]. AMERICAN HEART JOURNAL, 2009, 157 (01) : 102 - U15
  • [4] Left ventricular strain and strain rate: characterization of the effect of load in human subjects
    Burns, Andrew T.
    La Gerche, Andre
    D'hooge, Jan
    MacIsaac, Andrew I.
    Prior, David L.
    [J]. EUROPEAN JOURNAL OF ECHOCARDIOGRAPHY, 2010, 11 (03): : 283 - 289
  • [5] Prospective comparison of speckle tracking longitudinal bidimensional strain between two vendors
    Castel, Anne-Laure
    Szymanski, Catherine
    Delelis, Francois
    Levy, Franck
    Menet, Aymeric
    Mailliet, Amandine
    Marotte, Nathalie
    Graux, Pierre
    Tribouilloy, Christophe
    Marechaux, Sylvestre
    [J]. ARCHIVES OF CARDIOVASCULAR DISEASES, 2014, 107 (02) : 96 - 104
  • [6] Global 2-Dimensional Strain as a New Prognosticator in Patients With Heart Failure
    Cho, Goo-Yeong
    Marwick, Thomas H.
    Kim, Hyun-Sook
    Kim, Min-Kyu
    Hong, Kyung-Soon
    Oh, Dong-Jin
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2009, 54 (07) : 618 - 624
  • [7] Global and regional longitudinal strain assessed by two-dimensional speckle tracking echocardiography identifies early myocardial dysfunction and transmural extent of myocardial scar in patients with acute ST elevation myocardial infarction and relatively preserved LV function
    Cimino, S.
    Canali, E.
    Petronilli, V.
    Cicogna, F.
    De Luca, L.
    Francone, M.
    Sardella, G.
    Iacoboni, C.
    Agati, L.
    [J]. EUROPEAN HEART JOURNAL-CARDIOVASCULAR IMAGING, 2013, 14 (08) : 805 - 811
  • [8] Quantification of the Variability Associated with Repeat Measurements of Left Ventricular Two-Dimensional Global Longitudinal Strain in a Real-World Setting
    Costa, Salvatore P.
    Beaver, Timothy A.
    Rollor, Joyce L.
    Vanichakarn, Pantila
    Magnus, Patrick C.
    Palac, Robert T.
    [J]. JOURNAL OF THE AMERICAN SOCIETY OF ECHOCARDIOGRAPHY, 2014, 27 (01) : 50 - 54
  • [9] D'hooge J, 2015, EUROPEAN HEART J CAR
  • [10] Detection of myocardial dysfunction in the presence of normal ejection fraction
    El-Menyar, Ayman A.
    Galzerano, Domenico
    Asaad, Nidal
    Al-Mulla, Abdulwahid
    Arafa, Salah E. O.
    Al Suwaidi, Jassim
    [J]. JOURNAL OF CARDIOVASCULAR MEDICINE, 2007, 8 (11) : 923 - 933