Rapid and Reliable Assessment of the Contrast Sensitivity Function on an iPad

被引:83
作者
Dorr, Michael [1 ,2 ]
Lesmes, Luis A. [1 ,2 ]
Lu, Zhong-Lin [3 ]
Bex, Peter J. [1 ]
机构
[1] Harvard Univ, Sch Med, Dept Ophthalmol, Schepens Eye Res Inst,Massachusetts Eye & Ear Inf, Boston, MA 02114 USA
[2] Adapt Sensory Technol, Boston, MA USA
[3] Ohio State Univ, Dept Psychol, Ctr Cognit & Behav Brain Imaging, Columbus, OH 43210 USA
基金
美国国家卫生研究院;
关键词
contrast sensitivity; adaptive testing; telemonitoring; BAYESIAN ADAPTIVE ESTIMATION; VISUAL-ACUITY; MACULAR DEGENERATION; SPATIAL-CONTRAST; MOBILE PHONE; RISK-FACTORS; GLAUCOMA; RELIABILITY; PERFORMANCE; RETINOPATHY;
D O I
10.1167/iovs.13-11743
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
PURPOSE. Letter acuity, the predominant clinical assessment of vision, is relatively insensitive to slow vision loss caused by eye disease. While the contrast sensitivity function (CSF) has demonstrated the potential to monitor the slow progress of blinding eye diseases, current tests of CSF lack the reliability or ease-of-use to capture changes in vision timely. To improve the current state of home testing for vision, we have developed and validated a computerized adaptive test on a commercial tablet device (iPad) that provides an efficient and easy-to-use assessment of the CSF. METHODS. We evaluated the reliability, accuracy, and flexibility of tablet-based CSF assessment. Repeated tablet-based assessments of the spatial CSF, obtained from four normally-sighted observers, which each took 3 to 5 minutes, were compared to measures obtained on CRT-based laboratory equipment; additional tablet-based measures were obtained from six subjects under three different luminance conditions. RESULTS. A Bland-Altman analysis demonstrated that tablet-based assessment was reliable for estimating sensitivities at specific spatial frequencies (coefficient of repeatability 0.14-0.40 log units). The CRT- and tablet-based results demonstrated excellent agreement with absolute mean sensitivity differences <0.05 log units. The tablet-based test also reliably identified changes in contrast sensitivity due to different luminance conditions. CONCLUSIONS. We demonstrate that CSF assessment on a mobile device is indistinguishable from that obtained with specialized laboratory equipment. We also demonstrate better reliability than tests used currently for clinical trials of ophthalmic therapies, drugs, and devices.
引用
收藏
页码:7266 / 7273
页数:8
相关论文
共 62 条
  • [1] ABRAHAMSSON M, 1986, INVEST OPHTH VIS SCI, V27, P1131
  • [2] SPENDING ON NEW DRUG DEVELOPMENT
    Adams, Christopher Paul
    Brantner, Van Vu
    [J]. HEALTH ECONOMICS, 2010, 19 (02) : 130 - 141
  • [3] The noisy-bit method for digital displays: Converting a 256 luminance resolution into a continuous resolution
    Allard, Remy
    Faubert, Jocelyn
    [J]. BEHAVIOR RESEARCH METHODS, 2008, 40 (03) : 735 - 743
  • [4] American National Standards Institute, 2007, AM NAT STAND OPHTH M
  • [5] Applegate RA., 1997, VISION SCI ITS APPL, P98
  • [6] Bland JM, 1999, STAT METHODS MED RES, V8, P135, DOI 10.1177/096228029900800204
  • [7] VISUAL-ACUITY AND CONTRAST SENSITIVITY IN PATIENTS WITH CEREBRAL LESIONS
    BODISWOL.I
    [J]. SCIENCE, 1972, 178 (4062) : 769 - &
  • [8] Mobile Phone Based Clinical Microscopy for Global Health Applications
    Breslauer, David N.
    Maamari, Robi N.
    Switz, Neil A.
    Lam, Wilbur A.
    Fletcher, Daniel A.
    [J]. PLOS ONE, 2009, 4 (07):
  • [9] Mobile communication using a mobile phone with a glucometer for glucose control in Type 2 patients with diabetes: as effective as an Internet-based glucose monitoring system
    Cho, Jae-Hyoung
    Lee, Hye-Chung
    Lim, Dong-Jun
    Kwon, Hyuk-Sang
    Yoon, Kun-Ho
    [J]. JOURNAL OF TELEMEDICINE AND TELECARE, 2009, 15 (02) : 77 - 82
  • [10] Age related macular degeneration
    Chopdar, A
    Chakravarthy, U
    Verma, D
    [J]. BRITISH MEDICAL JOURNAL, 2003, 326 (7387): : 485 - 488