Cardiac pacemaker: In vitro assessment at 1.5 T

被引:36
作者
Shellock, FG
Fieno, DS
Thomson, LJ
Talavage, TM
Berman, DS
机构
[1] Univ So Calif, Keck Sch Med, Dept Radiol, Los Angeles, CA 90045 USA
[2] Univ So Calif, Keck Sch Med, Dept Med, Los Angeles, CA 90045 USA
[3] Inst Magnet Resonance Safety Educ & Res, Los Angeles, CA USA
[4] Cedars Sinai Med Ctr, Dept Imaging, Los Angeles, CA 90048 USA
[5] Purdue Univ, Sch Elect & Comp Engn, W Lafayette, IN 47907 USA
关键词
D O I
10.1016/j.ahj.2005.04.015
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background In vitro testing is used to determine safe parameters before performing magnetic resonance imaging (MRI) on a patient with an implant. Therefore, the objective of this study was to evaluate a cardiac pacemaker using a 1.5-T magnetic resonance (MR) system. Methods A modern cardiac pacemaker (INSIGNIA I PLUS, Model 1298, and FINELINE II, Model 4471, pacing leads; Guidant Corporation, St Paul, MN) was evaluated for magnetic field interactions at 1.5 T. Magnetic resonance imaging-related heating was assessed using 3 different 1.5-T scanners operating at various levels of radio-frequency power and imaging conditions. Functional aspects of the pacemaker were evaluated immediately before and after MRI (9 different pulse sequences). Artifacts were also characterized. Results Magnetic field interactions for the pacemaker were minor. Temperature changes measured in vitro were at levels that are not expected to pose a risk for specific MR conditions (< 4.0 degrees C). The function of the pacemaker was unaffected by MRI. Artifacts were minor for the leads and relatively large for the implantable pulse generator. Conclusion The findings indicated that this pacemaker exhibited acceptable safety features relative to the use of a 1.5-T MR system. If induced currents do not occur for this device, it may be safe for a patient to undergo MRI by following specific conditions. The results are specific to the pacemaker tested, the MR systems, and conditions used in this evaluation.
引用
收藏
页码:436 / 443
页数:8
相关论文
共 30 条
[1]   NUCLEAR MAGNETIC-RESONANCE IMAGING IN A PATIENT WITH A DDD PACEMAKER [J].
ALAGONA, P ;
TOOLE, JC ;
MANISCALCO, BS ;
GLOVER, MU ;
ABERNATHY, GT ;
PRIDA, XE .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1989, 12 (04) :619-619
[2]  
American Society for Testing and Materials (ASTM), 2003, INT F2119 TEST METH
[3]  
*ASTM, 2003, INT F2213 02 STAND T
[4]  
*ASTM, 2002, STANT TEST METH MEAS, P1576
[5]  
*ASTM, 2003, INT F2U82 TEST METH
[6]   Evaluation of specific absorption rate as a dosimeter of MRI-related implant heating [J].
Baker, KB ;
Tkach, JA ;
Nyenhuis, JA ;
Phillips, M ;
Shellock, FG ;
Gonzalez-Martinez, J ;
Rezai, AR .
JOURNAL OF MAGNETIC RESONANCE IMAGING, 2004, 20 (02) :315-320
[7]  
*CEN, 455021 CEN CENELEC
[8]   Pacing in magnetic resonance imaging environment:: Clinical and technical considerations on compatibility [J].
Duru, F ;
Luechinger, R ;
Scheidegger, MB ;
Lüscher, TF ;
Boesiger, P ;
Candinas, R .
EUROPEAN HEART JOURNAL, 2001, 22 (02) :113-124
[9]  
Garcia-Bolao I, 1998, ACTA CARDIOL, V53, P33
[10]  
INBOR S, 1993, AM J MED SCI, V3, P174