共 50 条
Comprehensive approach to the validation of the standard method for total reflection X-ray fluorescence analysis of water
被引:31
|作者:
Borgese, Laura
[1
,2
]
Dalipi, Rogerta
[1
,2
,3
]
Riboldi, Alessandro
[1
,2
]
Bilo, Fabjola
[1
,2
]
Zacco, Annalisa
[1
,2
]
Federici, Stefania
[1
,2
]
Bettinelli, Maurizio
[4
]
Bontempi, Elza
[1
,2
]
Depero, Laura Eleonora
[1
,2
]
机构:
[1] Univ Brescia, INSTM, Via Branze 38, I-25123 Brescia, Italy
[2] Univ Brescia, Chem Technol Lab, Dept Mech & Ind Engn, Via Branze 38, I-25123 Brescia, Italy
[3] Ist Zooprofilattico Sperimentale Lombardia & Emil, Via Antonio Bianchi 7-9, I-25124 Brescia, Italy
[4] UNICHIM Assoc Unificaz Settore Ind Chimica, Milan, Italy
来源:
关键词:
Water;
TXRF;
Validation;
Calibration;
Elements;
MULTIELEMENT ANALYSIS;
SAMPLES;
SPECTROMETRY;
UNCERTAINTY;
STRATEGIES;
D O I:
10.1016/j.talanta.2017.12.087
中图分类号:
O65 [分析化学];
学科分类号:
070302 ;
081704 ;
摘要:
In this work, we present the validation of the chemical method for total reflection X-ray fluorescence (TXRF) analysis of water, proposed as a standard to the International Standard Organization. The complete experimental procedure to define the linear calibration range, elements sensitivities, limits of detection and quantification, precision and accuracy is presented for a commercial TXRF spectrometer equipped with Mo X-ray tube. Least squares linear regression, including all statistical tests is performed separately for each element of interest to extract sensitivities. Relative sensitivities with respect to Ga, as internal standard, are calculated. Accuracy and precision of the quantification procedure using Ga as internal standard is evaluated with reference water samples. A detailed discussion on the calibration procedure and the limitation of the use of this method for quantitative analysis of water is presented.
引用
收藏
页码:165 / 171
页数:7