Peripheral nerve conduction recorded by a micro gradiometer system (micro-SQUID) in humans

被引:19
作者
Hoshiyama, M [1 ]
Kakigi, R
Nagata, O
机构
[1] Natl Inst Physiol Sci, Dept Integrat Physiol, Okazaki, Aichi 4448585, Japan
[2] Natl Inst Physiol Sci, Tech Div, Okazaki, Aichi 4448585, Japan
关键词
Micro-SQUID; nerve action field; human; peripheral nerve; gradiometer; quadrupole;
D O I
10.1016/S0304-3940(99)00589-3
中图分类号
Q189 [神经科学];
学科分类号
071006 ;
摘要
We developed a new multi (twelve)-channel gradiometer system with the smallest and highest quality superconducting quantum interference device (micro-SQUID). A very small distance (3.80 mm) between the sensor coils and the skin provides quite high spatial resolution. The actual whole image of the sensory nerve action fields (NAF) of the human median nerve at the wrist were successfully recorded following digital nerve stimulation by using the micro-SQUID. The NAF showed the biphasic waveform at each of the 12 channels, and the isomagnetic field map clearly showed the moving quadrupole pattern. The quadrupole comprised a dipole for depolarization followed by another dipole with the opposite direction for repolarization. The polarized length of the nerve obtained by reconstructing the magnetic field maps was approximately 17 cm, and the magnetic field complex moved along the nerve from the distal to the proximal part of the wrist at 58.7 m/s. (C) 1999 Elsevier Science Ireland Ltd. All rights reserved.
引用
收藏
页码:199 / 202
页数:4
相关论文
共 10 条
[1]  
Buchthal F., 1966, Brain Res, V3, P1
[2]  
ERNE SN, 1988, BIOMAGNETISM 87, P166
[3]  
HAHISMOTO I, 1991, ELECTROEN CLIN NEURO, V81, P332
[4]   MULTICHANNEL DETECTION OF MAGNETIC COMPOUND ACTION FIELDS OF MEDIAN AND ULNAR NERVES [J].
HARI, R ;
HALLSTROM, J ;
TIIHONEN, J ;
JOUTSINIEMI, SL .
ELECTROENCEPHALOGRAPHY AND CLINICAL NEUROPHYSIOLOGY, 1989, 72 (03) :277-280
[5]   VISUALIZATION OF A MOVING QUADRUPOLE WITH MAGNETIC MEASUREMENTS OF PERIPHERAL-NERVE ACTION FIELDS [J].
HASHIMOTO, I ;
MASHIKO, T ;
MIZUTA, T ;
IMADA, T ;
IWASE, K ;
OKAZAKI, H .
ELECTROENCEPHALOGRAPHY AND CLINICAL NEUROPHYSIOLOGY, 1994, 93 (06) :459-467
[6]   USING A MAGNETOMETER TO IMAGE A TWO-DIMENSIONAL CURRENT DISTRIBUTION [J].
ROTH, BJ ;
SEPULVEDA, NG ;
WIKSWO, JP .
JOURNAL OF APPLIED PHYSICS, 1989, 65 (01) :361-372
[7]   BIOMAGNETIC FUNCTIONAL LOCALIZATION OF A PERIPHERAL-NERVE IN MAN [J].
TRAHMS, L ;
ERNE, SN ;
TRONTELJ, Z ;
CURIO, G ;
AUST, P .
BIOPHYSICAL JOURNAL, 1989, 55 (06) :1145-1153
[8]   MAGNETIC-FIELD OF A NERVE IMPULSE - 1ST MEASUREMENTS [J].
WIKSWO, JP ;
BARACH, JP ;
FREEMAN, JA .
SCIENCE, 1980, 208 (4439) :53-55
[9]  
WIKSWO JP, 1990, ADV BIOMAGNETISM, P1
[10]  
WIKSWO JP, 1990, ADV BIOMAGNETISM, P681