New adaptive statistical iterative reconstruction ASiR-V: Assessment of noise performance in comparison to ASiR

被引:63
作者
De Marco, Paolo [1 ]
Origgi, Daniela [1 ]
机构
[1] European Inst Oncol, Med Phys Unit, Milan, Italy
关键词
computed tomography; iterative reconstruction; noise power spectrum; COMPUTED-TOMOGRAPHY; RADIATION RISKS; POWER SPECTRUM; IMAGE QUALITY; REDUCTION; PHANTOM; MDCT;
D O I
10.1002/acm2.12253
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
PurposeTo assess the noise characteristics of the new adaptive statistical iterative reconstruction (ASiR-V) in comparison to ASiR. MethodsA water phantom was acquired with common clinical scanning parameters, at five different levels of CTDIvol. Images were reconstructed with different kernels (STD, SOFT, and BONE), different IR levels (40%, 60%, and 100%) and different slice thickness (ST) (0.625 and 2.5mm), both for ASiR-V and ASiR. Noise properties were investigated and noise power spectrum (NPS) was evaluated. ResultsASiR-V significantly reduced noise relative to FBP: noise reduction was in the range 23%-60% for a 0.625mm ST and 12%-64% for the 2.5mm ST. Above 2mGy, noise reduction for ASiR-V had no dependence on dose. Noise reduction for ASIR-V has dependence on ST, being greater for STD and SOFT kernels at 2.5mm. For the STD kernel ASiR-V has greater noise reduction for both ST, if compared to ASiR. For the SOFT kernel, results varies according to dose and ST, while for BONE kernel ASIR-V shows less noise reduction. NPS for CT Revolution has dose dependent behavior at lower doses. NPS for ASIR-V and ASiR is similar, showing a shift toward lower frequencies as the IR level increases for STD and SOFT kernels. The NPS is different between ASiR-V and ASIR with BONE kernel. NPS for ASiR-V appears to be ST dependent, having a shift toward lower frequencies for 2.5mm ST. ConclusionsASiR-V showed greater noise reduction than ASiR for STD and SOFT kernels, while keeping the same NPS. For the BONE kernel, ASiR-V presents a completely different behavior, with less noise reduction and modified NPS. Noise properties of the ASiR-V are dependent on reconstruction slice thickness. The noise properties of ASiR-V suggest the need for further measurements and efforts to establish new CT protocols to optimize clinical imaging.
引用
收藏
页码:275 / 286
页数:12
相关论文
共 27 条
[1]  
BEIR VII, 2006, Health Risks from Exposure to Low Levels of Ionizing Radiation: BEIR VII Phase 2, DOI [DOI 10.17226/11340, 10.17226/11340]
[2]   Application of the noise power spectrum in modern diagnostic MDCT: Part I. Measurement of noise power spectra and noise equivalent quanta [J].
Boedeker, K. L. ;
Cooper, V. N. ;
McNitt-Gray, M. F. .
PHYSICS IN MEDICINE AND BIOLOGY, 2007, 52 (14) :4027-4046
[3]   Current concepts - Computed tomography - An increasing source of radiation exposure [J].
Brenner, David J. ;
Hall, Eric J. .
NEW ENGLAND JOURNAL OF MEDICINE, 2007, 357 (22) :2277-2284
[4]   Radiation risks potentially associated with low-dose CT screening of adult smokers for lung cancer [J].
Brenner, DJ .
RADIOLOGY, 2004, 231 (02) :440-445
[5]   Estimating Risk of Low Radiation Doses - A Critical Review of the BEIR VII Report and its Use of the Linear No-Threshold (LNT) Hypothesis [J].
Calabrese, Edward J. ;
O'Connor, Michael K. .
RADIATION RESEARCH, 2014, 182 (05) :463-474
[6]  
Fan J, 2014, BENEFITS ASIR V RECO
[7]   Computed tomography-old ideas and new technology [J].
Fleischmann, Dominik ;
Boas, F. Edward .
EUROPEAN RADIOLOGY, 2011, 21 (03) :510-517
[8]   Radiation Risks of Medical Imaging: Separating Fact from Fantasy [J].
Hendee, William R. ;
O'Connor, Michael K. .
RADIOLOGY, 2012, 264 (02) :312-321
[9]  
Hsieh J., 2008, ADAPTIVE STAT ITERAT
[10]   Reducing Radiation Dose in Adult Head CT using Iterative Reconstruction - A Clinical Study in 177 Patients [J].
Kaul, D. ;
Kahn, J. ;
Huizing, L. ;
Wiener, E. ;
Grupp, U. ;
Boening, G. ;
Ghadjar, P. ;
Renz, D. M. ;
Streitparth, F. .
ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 2016, 188 (02) :155-162