Eliminating the interference of ascorbic acid and uric acid to the amperometric glucose biosensor by cation exchangers membrane and size exclusion membrane

被引:59
作者
Yuan, CJ
Hsu, CL
Wang, SC
Chang, KS [1 ]
机构
[1] Yuanpei Univ Sci & Technol, Dept Food Sci, Hsinchu 300, Taiwan
[2] Natl Chiao Tung Univ, Dept Biol Sci & Technol, Hsinchu 300, Taiwan
[3] Yuanpei Univ Sci & Technol, Dept Biotechnol, Hsinchu 300, Taiwan
关键词
PVA-SbQ; nafion; planar electrode; amperometric biosensor; polyion;
D O I
10.1002/elan.200503359
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The characteristics of a multiuse planar amperometric biosensor modified with Nafion and/or polyion membrane were investigated. A new enzyme immobilization process was proposed, in which the polyvinyl alcohol bearing a styrylpyridinium (SbQ)/glucose oxidase composite was treated with glutaraldehyde vapor prior to the photo-crosslinking reaction. The resulting planar enzyme electrode remains active for at least 150 days. Compared with poly-L-lysine/poly (4-stryenesulfonate) polyion complex membrane the Nafion membrane showed marked effect to reduce the electrochemical response of the modified planar enzyme electrode to the biological interferents, such as ascorbic acid and uric acid. Furthermore, Nafion membrane can effective restricting the oxidized anionic interferent to adhear on its surface, thereby the fouling of the electrode was avoided.
引用
收藏
页码:2239 / 2245
页数:7
相关论文
共 28 条
[1]   Configurations used in the design of screen-printed enzymatic biosensors.: A review [J].
Albareda-Sirvent, M ;
Merkoçi, A ;
Alegret, S .
SENSORS AND ACTUATORS B-CHEMICAL, 2000, 69 (1-2) :153-163
[2]   Hydrogen peroxide detection with improved selectivity and sensitivity using constant current potentiometry [J].
Anh, DTV ;
Olthuis, W ;
Bergveld, P .
SENSORS AND ACTUATORS B-CHEMICAL, 2003, 91 (1-3) :1-4
[3]  
[Anonymous], 1994, BIOSENS BIOELECTRON, V9, P243
[4]   Determination of glutamate pyruvate transaminase activity in clinical specimens using a biosensor composed of immobilized L-glutamate oxidase in a photo-crosslinkable polymer membrane on a palladium-deposited screen-printed carbon electrode [J].
Chang, KS ;
Hsu, WL ;
Chen, HY ;
Chang, CK ;
Chen, CY .
ANALYTICA CHIMICA ACTA, 2003, 481 (02) :199-208
[5]   Amperometric nitrogen dioxide gas sensor based on PAn/Au/Nafion® prepared by constant current and cyclic voltammetry methods [J].
Do, JS ;
Chang, WB .
SENSORS AND ACTUATORS B-CHEMICAL, 2004, 101 (1-2) :97-106
[6]  
FANG X, 2002, TALANTA, V57, P365
[7]   Amperometric biosensors based on nafion coated screen-printed electrodes for the determination of cholinesterase inhibitors [J].
Gogol, EV ;
Evtugyn, GA ;
Marty, JL ;
Budnikov, HC ;
Winter, VG .
TALANTA, 2000, 53 (02) :379-389
[8]  
Grennan K, 2001, ELECTROANAL, V13, P745, DOI 10.1002/1521-4109(200105)13:8/9<745::AID-ELAN745>3.0.CO
[9]  
2-B
[10]   IMMOBILIZATION OF ENZYMES WITH USE OF PHOTOSENSITIVE POLYMERS HAVING THE STILBAZOLIUM GROUP [J].
ICHIMURA, K ;
WATANABE, S .
JOURNAL OF POLYMER SCIENCE PART A-POLYMER CHEMISTRY, 1980, 18 (03) :891-902