Programmable Shunt Assistant Tested in Cambridge Shunt Evaluation Laboratory

被引:7
作者
Czosnyka, Marek [1 ]
Czosnyka, Zofia [1 ]
Pickard, John D. [1 ]
机构
[1] Univ Cambridge, Dept Clin Neurosci, Cambridge, England
来源
HYDROCEPHALUS | 2012年 / 113卷
关键词
Shunt; Hydrocephalus; Cerebrospinal fluid; HYDROCEPHALUS; FLOW;
D O I
10.1007/978-3-7091-0923-6_15
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Objective: The programmable Shunt Assistant (capped!) (ProSA) shunt system has recently been introduced into clinical use. The system can be in vivo adjusted magnetically, and this adjustability is supposed to affect CS F drainage only in the vertical body position. Materials and methods: We tested a combination of ProSA with fixed differential pressure valve designed to drain cerebrospinal fluid (CSF) from the brain ventricles into the peritoneal cavity. Results: The ProSA showed good mechanical durability and stability of hydrodynamic performance over a 60-day period. The flow-pressure performance curves and operating pressures were stable, fell within the limits specified by the manufacturer, and changed according to the programmed performance levels. The ProSA system has higher than usual hydrodynamic resistance (around 8.8 mmHg/(mL/min)) in the vertical position. Operating pressure in the vertical position reacted repeatedly to changes of settings within the limits 0-40 cmH(2)O and was reduced gradually when the axis of the valve declined from vertical to horizontal. External programming proved to be reliable. Strong magnetic fields (3-T MRI) were not able to change the programming of ProSA. Conclusion: ProSA works in the horizontal position as differential fixed pressure low hydrodynamic resistance and in the vertical position as an adjustable, normal hydrodynamic resistance valve. It is able to compensate for posture-related overdrainage
引用
收藏
页码:71 / 76
页数:6
相关论文
共 4 条
[1]   OVERDRAINAGE AND SHUNT TECHNOLOGY - A CRITICAL COMPARISON OF PROGRAMMABLE, HYDROSTATIC AND VARIABLE RESISTANCE VALVES AND FLOW-REDUCING DEVICES [J].
ASCHOFF, A ;
KREMER, P ;
BENESCH, C ;
FRUH, K ;
KLANK, A ;
KUNZE, S .
CHILDS NERVOUS SYSTEM, 1995, 11 (04) :193-202
[2]   Laboratory testing of hydrocephalus shunts - Conclusion of the UK Shunt Evaluation Programme [J].
Czosnyka, Z ;
Czosnyka, M ;
Richards, HK ;
Pickard, JD .
ACTA NEUROCHIRURGICA, 2002, 144 (06) :525-538
[3]   Pattern of white matter regional cerebral blood flow and autoregulation in normal pressure hydrocephalus [J].
Momjian, S ;
Owler, BK ;
Czosnyka, Z ;
Czosnyka, M ;
Pena, A ;
Pickard, JD .
BRAIN, 2004, 127 :965-972
[4]   In vivo assessment of hydrocephalus shunt [J].
Petrella, G. ;
Czosnyka, M. ;
Smielewski, P. ;
Allin, D. ;
Guazzo, E. P. ;
Pickard, J. D. ;
Czosnyka, Z. H. .
ACTA NEUROLOGICA SCANDINAVICA, 2009, 120 (05) :317-323