Gastrointestinal Targeted Sampling and Sensing via Embedded Packaging of Integrated Capsule System

被引:15
作者
Banis, George E. [1 ]
Beardslee, Luke A. [2 ]
Stine, Justin M. [3 ]
Sathyam, Rajendra Mayavan [3 ]
Ghodssi, Reza [3 ,4 ]
机构
[1] Univ Maryland, Fischell Dept Bioengn, College Pk, MD 20742 USA
[2] Univ Maryland, Syst Res Inst, College Pk, MD 20742 USA
[3] Univ Maryland, Dept Elect & Comp Engn, College Pk, MD 20742 USA
[4] Univ Maryland, Inst Syst Res, College Pk, MD 20742 USA
基金
美国国家科学基金会;
关键词
Ingestible systems; capsule; capacitive sensing; gastrointestinal tract; system integration; SWALLOWABLE-CAPSULE; ENDOSCOPY; DELIVERY;
D O I
10.1109/JMEMS.2019.2897246
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
Here, we introduce a wireless platform developed to leverage biochemical characteristics of various gastrointestinal (GI) secretions to selectively dissolve pH-sensitive polymers used as coatings over a 3-D printed biocompatible capsule, allowing fluid entry into sensing chambers containing capacitive sensors. Ingestible capsule technology is increasing prevalence in autonomously accessing regions of GI that otherwise require clinical intervention. Our platform is operated using a 3.3-V power source with sensors, possessing a 0.3-220-pF dynamic range with a sensitivity of 3.2 x 10(-3) pF/mV that relay the change in capacitance to a smart system-in-package, which is then transmitted via Bluetooth low energy to an external Android phone. The system is tested with a variety of polymers, which are used in the pharmaceutical industry, to control dissolution at different GI regions characteristic of a specific pH profile. This real-time, cost-effective, and user-friendly platform will ultimately be used to test alternative coating materials that respond to specific biomarkers present in GI secretions, offering significant potential for enhancing gut diagnostics. [2018-0249]
引用
收藏
页码:219 / 225
页数:7
相关论文
共 29 条
[1]  
Adler Samuel N, 2011, Therap Adv Gastroenterol, V4, P265, DOI 10.1177/1756283X11401645
[2]  
Banis G.E., 2018, A Solid-State Sensors, Actuators and Microsystems Workshop, P96, DOI [DOI 10.31438/TRF.HH2018.27, 10.31438/trf.hh2018.27]
[3]   Enteric coated HPMC capsules designed to achieve intestinal targeting [J].
Cole, ET ;
Scott, RA ;
Connor, AL ;
Wilding, IR ;
Petereit, HU ;
Schminke, C ;
Beckert, T ;
Cadé, D .
INTERNATIONAL JOURNAL OF PHARMACEUTICS, 2002, 231 (01) :83-95
[4]   Variability of gastrointestinal transit in healthy women and men [J].
Degen, LP ;
Phillips, SF .
GUT, 1996, 39 (02) :299-305
[5]  
Demosthenous P, 2015, IEEE ENG MED BIO, P2143, DOI 10.1109/EMBC.2015.7318813
[6]   Colon Capsule Endoscopy: Review and Perspectives [J].
Friedel, David ;
Modayil, Rani ;
Stavropoulos, Stavros .
GASTROENTEROLOGY RESEARCH AND PRACTICE, 2016, 2016
[7]   Ingestible capsule for impedance and pH monitoring in the esophagus [J].
Gonzalez-Guillaumin, Jose L. ;
Sadowski, Daniel C. ;
Kaler, Karan V. I. S. ;
Mintchev, Martin P. .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2007, 54 (12) :2231-2236
[8]   Implementation of radiotelemetry in a lab-in-a-pill format [J].
Johannessen, EA ;
Wang, L ;
Reid, SWJ ;
Cumming, DRS ;
Cooper, JM .
LAB ON A CHIP, 2006, 6 (01) :39-45
[9]  
Johnson L.R., 2012, Physiology of the gastrointestinal tract, VFifth
[10]   Ingestible Sensors [J].
Kalantar-zadeh, Kourosh ;
Ha, Nam ;
Ou, Jian Zhen ;
Berean, Kyle J. .
ACS SENSORS, 2017, 2 (04) :468-483