Microcontroller-Based Clip Force Reading System for Brain Aneurysms

被引:1
作者
Kutbay, Ugurhan [1 ]
机构
[1] Gazi Univ, Dept Elect & Elect Engn, Ankara, Turkey
关键词
Brain aneurysm; Clipping; Clip force sensing; Embedded system; Load cell; Strain gauge; FLEXIBLE TACTILE SENSOR; INTRACRANIAL ANEURYSMS;
D O I
10.1007/s40846-020-00543-6
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Purpose An aneurysm is a balloon-like bulge on an artery wall. Aneurysms can rupture easily, potentially causing a cerebral hemorrhage. To prevent this, a small clip is surgically installed. Importantly, if the clip is opened repeatedly during the operation, the clip closing force may change. Sometimes, the closing force of the clip cannot be accurately predicted during the operation. The main objectives of the present study were to prevent re-bleeding and to accurately measure the force of the clip for reducing the duration of the procedure. Methods A well-designed aneurysm clip microforce sensing system was developed in this study. The Brain Aneurysm Clip Force Reading System allows the clip force in the surgical operation to be measured by using an S-type load cell. The system has a simple structure that includes a clip force sensor and a liquid-crystal display screen. The main contribution of the study is a small, low-cost printed circuit board design that can be used in surgery. Results The proposed system was verified with 12 differently shaped aneurysm clips, and the results were analyzed. For the obtained and measured values, the mean squared error and maximum variance were 1.53 and 6.76 g(2), respectively. Conclusion The results indicated that the proposed system can be used in surgical operations. Because the developed system informs doctors of the correct clip force, it can save lives in operations.
引用
收藏
页码:748 / 756
页数:9
相关论文
共 26 条
[1]   Compensatory force measurement and multimodal force feedback for remote-controlled vascular interventional robot [J].
Bao, Xianqiang ;
Guo, Shuxiang ;
Xiao, Nan ;
Li, Youxiang ;
Shi, Liwei .
BIOMEDICAL MICRODEVICES, 2018, 20 (03)
[2]   Polymer Micromachined Flexible Tactile Sensor for Three-Axial Loads Detection [J].
Choi, Woo-Chang .
TRANSACTIONS ON ELECTRICAL AND ELECTRONIC MATERIALS, 2010, 11 (03) :130-133
[3]   Development of polyimide flexible tactile sensor skin [J].
Engel, J ;
Chen, J ;
Liu, C .
JOURNAL OF MICROMECHANICS AND MICROENGINEERING, 2003, 13 (03) :359-366
[4]   Influence of Spinal Manipulative Therapy Force Magnitude and Application Site on Spinal Tissue Loading: A Biomechanical Robotic Serial Dissection Study in Porcine Motion Segments [J].
Funabashi, Martha ;
Nougarou, Francois ;
Descarreaux, Martin ;
Prasad, Narasimha ;
Kawchuk, Greg .
JOURNAL OF MANIPULATIVE AND PHYSIOLOGICAL THERAPEUTICS, 2017, 40 (06) :387-396
[5]   Paradoxical Trends in the Management of Unruptured Cerebral Aneurysms in the United States Analysis of Nationwide Database Over a 10-Year Period [J].
Huang, Michael C. ;
Baaj, Ali A. ;
Downes, Katheryne ;
Youssef, A. Samy ;
Sauvageau, Eric ;
van Loveren, Harry R. ;
Agazzi, Siviero .
STROKE, 2011, 42 (06) :1730-1735
[6]   A Computer-Aided Diagnosis System for Measuring Carotid Artery Intima-Media Thickness (IMT) Using Quaternion Vectors [J].
Kutbay, Ugurhan ;
Hardalac, Firat ;
Akbulut, Mehmet ;
Akaslan, Unsal ;
Serhathoglu, Selami .
JOURNAL OF MEDICAL SYSTEMS, 2016, 40 (06)
[7]   Titanium aneurysm clips .1. Mechanical, radiological, and biocompatibility testing [J].
Lawton, MT ;
Ho, JC ;
Bichard, WD ;
Coons, SW ;
Zabramski, LM ;
Spetzler, RF .
NEUROSURGERY, 1996, 38 (06) :1158-1163
[8]   International subarachnoid aneurysm trial (ISAT) of neurosurgical clipping versus endovascular coiling in 2143 patients with ruptured intracranial aneurysms: a randomised comparison of effects on survival, dependency, seizures, rebleeding, subgroups, and aneurysm occlusion [J].
Molyneux, AJ ;
Kerr, RSC ;
Yu, LM ;
Clarke, M ;
Sneade, M ;
Yarnold, JA ;
Sandercock, P .
LANCET, 2005, 366 (9488) :809-817
[9]   Simple Method of Using Strain Gages to Estimate Stress Fields Near a Notch in Structures Subject to Large Deflections [J].
Morita, N. ;
Harada, T. ;
Noguchi, H. .
JOURNAL OF TESTING AND EVALUATION, 2017, 45 (05)
[10]   Predictors and impact of aneurysm rebleeding after subarachnoid hemorrhage [J].
Naidech, AM ;
Janjua, N ;
Kreiter, KT ;
Ostapkovich, ND ;
Fitzsimmons, BF ;
Parra, A ;
Commichau, C ;
Connolly, ES ;
Mayer, SA .
ARCHIVES OF NEUROLOGY, 2005, 62 (03) :410-416