Tip-sample interaction in a "shear-force" near-field scanning optical microscope

被引:22
作者
Hsu, K [1 ]
Gheber, LA [1 ]
机构
[1] Johns Hopkins Univ, Dept Biol, Baltimore, MD 21218 USA
关键词
D O I
10.1063/1.1149967
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The interaction between the tip of a near-field scanning optical microscope (NSOM) and the sample it scans is analyzed and compared to a simple tapping model. The approach curves acquired with the NSOM are in excellent agreement with the model, and additional experiments strongly point against a noncontact interaction (such as shear force). Based on this model we are also able to explain the oscillations pattern of the feedback loop. We conclude that our straight-fiber tip, operating under "shear-force" control, intermittently contacts the surface it is scanning, in a way similar to the tapping mode in atomic force microscope. (C) 1999 American Institute of Physics. [S0034-6748(99)04509-8].
引用
收藏
页码:3609 / 3613
页数:5
相关论文
共 50 条
[1]   Tip-sample interaction in a shear-force near-field scanning optical microscope [J].
Hsu, Kate ;
Gheber, Levi A. .
Review of Scientific Instruments, 70 (09) :3609-3613
[2]   An optical tip-sample distance control for a scanning near-field optical microscope [J].
Kramer, A ;
Hartmann, T ;
Stadler, SM ;
Guckenberger, R .
ULTRAMICROSCOPY, 1995, 61 (1-4) :191-195
[3]   A cryogenic scanning near-field optical microscope with shear-force gapwidth control [J].
Kramer, A ;
Segura, JM ;
Hunkeler, A ;
Renn, A ;
Hecht, B .
REVIEW OF SCIENTIFIC INSTRUMENTS, 2002, 73 (08) :2937-2941
[4]   Dynamics of the tip-sample interaction in near-field scanning optical microscopy and the implications for shear force as an accurate distance measure [J].
Jahncke, CL ;
Huerth, SH ;
Clark, B ;
Hallen, HD .
APPLIED PHYSICS LETTERS, 2002, 81 (21) :4055-4057
[5]   Non-optical shear-force detection for scanning near-field optical microscope [J].
Tamaru, H ;
Miyano, K .
JAPANESE JOURNAL OF APPLIED PHYSICS PART 2-LETTERS & EXPRESS LETTERS, 1997, 36 (6B) :L821-L823
[6]   Non-optical shear-force detection for scanning near-field optical microscope [J].
Univ of Tokyo, Tokyo, Japan .
Japanese Journal of Applied Physics, Part 2: Letters, 1997, 36 (06)
[7]   The ultrasonic/shear-force microscope: Integrating ultrasonic sensing into a near-field scanning optical microscope [J].
La Rosa, A ;
Cui, X ;
McCollum, J ;
Li, N ;
Nordstrom, R .
REVIEW OF SCIENTIFIC INSTRUMENTS, 2005, 76 (09)
[8]   Development of a shear-force scanning near-field cathodoluminescence microscope for characterization of nanostructures' optical properties [J].
Bercu, N. B. ;
Troyon, M. ;
Molinari, M. .
JOURNAL OF MICROSCOPY, 2016, 263 (03) :357-364
[9]   BIOLOGICAL NEAR-FIELD SCANNING OPTICAL MICROSCOPY - INSTRUMENTATION AND SAMPLE ISSUES FOR SHEAR-FORCE FEEDBACK [J].
VALASKOVIC, GA ;
HOLTON, M ;
MORRISON, GH .
ULTRAMICROSCOPY, 1995, 57 (2-3) :212-218
[10]   Reduction of tip-sample interaction forces for scanning near-field optical microscopy in a liquid environment [J].
Lambelet, P ;
Pfeffer, M ;
Sayah, A ;
Marquis-Weible, F .
ULTRAMICROSCOPY, 1998, 71 (1-4) :117-121