Progressive decrease in amplitude of intracardiac ventricular electrogram and higher left ventricular ejection fraction are associated with conductors externalization in Riata leads

被引:7
作者
Kubala, Maciej [1 ]
Traulle, Sarah [1 ]
Leborgne, Laurent [1 ]
Hermida, Jean-Sylvain [1 ]
机构
[1] Picardie Univ Hosp, Cardiol Unit, Dept Cardiac Arrhythmia, F-80000 Amiens, France
来源
EUROPACE | 2013年 / 15卷 / 08期
关键词
Implantable cardioverter-defibrillator; Riata lead; Lead failure; Fluoroscopy; Insulation; DEFIBRILLATOR LEAD; ICD LEADS; INSULATION; FAILURE; PREVALENCE; DEFECTS;
D O I
10.1093/europace/eut015
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Increased rates of structural abnormalities including externalized conductors have been reported in the Riata family of implantable cardioverter-defibrillator leads (St Jude Medical). Little is known about their reliability and the time lag for emergence of functional or structural abnormalities. Thirty-six patients who received small-caliber leads of the Riata family and who completed face-profile flouroscopies, repeated at every 6 months were included. We assessed the prevalence of conductors externalization and its relation to abnormal electrical parameters or adverse events. Thirty-six patients, mean age 64 10 years, with at least 7-month completed fluoroscopy follow-up were included in the analysis. Externalized conductors were identified in 12 (33) patients after a 53-month (13114) mean delay. A higher left ventricular ejection fraction (LVEF): 47 13 vs. 33 12, P 0.04, and a progressive decrease (30 of the initial value) in amplitude of ventricular electrogram 9/12 (75) vs. 4/24 (17), P 0.03 were independently associated with the fluoroscopic failures. Detection of the conductors externalization was preceded by an electrical lead abnormality in 10 (83) patients. Repeated face-profile fluoroscopies allowed detection of conductors externalization in 33 of patients implanted with Riata leads. Better LVEF and a progressive decrease in amplitude of intracardiac ventricular electrogram were independently associated with externalized conductors. The structural abnormality was preceded by an electrical lead dysfunction 83 of patients.
引用
收藏
页码:1198 / 1204
页数:7
相关论文
共 13 条
[11]  
Noad RL, 2012, HEART RHYTHM, V9, pS60
[12]   High prevalence of insulation failure with externalized cables in St. Jude Medical Riata family ICD leads: Fluoroscopic grading scale and correlation to extracted leads [J].
Parvathaneni, Sunthosh V. ;
Ellis, Christopher R. ;
Rottman, Jeffrey N. .
HEART RHYTHM, 2012, 9 (08) :1218-1224
[13]   Clinical Performance of the St. Jude Medical Riata Defibrillation Lead in a Large Patient Population [J].
Porterfield, James G. ;
Porterfield, Linda M. ;
Kuck, Karl H. ;
Corbisiero, Raffaele ;
Greenberg, Steven M. ;
Hindricks, Gerhard ;
Wazni, Oussama ;
Beau, Scott L. ;
Herre, John M. .
JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2010, 21 (05) :551-556