High dose-per-pulse electron beam dosimetry: Commissioning of the Oriatron eRT6 prototype linear accelerator for preclinical use

被引:158
|
作者
Jaccard, Maud [1 ]
Duran, Maria Teresa [1 ]
Petersson, Kristoffer [1 ]
Germond, Jean-Francois [1 ]
Liger, Philippe [2 ]
Vozenin, Marie-Catherine [3 ,4 ]
Bourhis, Jean [3 ,4 ]
Bochud, Francois [1 ]
Bailat, Claude [1 ]
机构
[1] Lausanne Univ Hosp, Inst Radiat Phys, Lausanne, Switzerland
[2] PMB Alcen, Peynier, France
[3] Lausanne Univ Hosp, Dept Radiat Oncol, Lausanne, Switzerland
[4] Lausanne Univ Hosp, Radiooncol Lab, DO CHUV, Lausanne, Switzerland
基金
瑞士国家科学基金会;
关键词
beam monitoring; commissioning; high dose-per-pulse; high dose-rate; linac; EBT3 GAFCHROMIC FILMS; RADIATION-THERAPY; CONFORMAL RADIOTHERAPY; IMAGE GUIDANCE; IRRADIATION; ONCOLOGY; TRACKING;
D O I
10.1002/mp.12713
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: The Oriatron eRT6 is an experimental high dose-per-pulse linear accelerator (linac) which was designed to deliver an electron beam with variable dose-rates, ranging from a few Gy/min up to hundreds of Gy/s. It was built to study the radiobiological effects of high dose-per-pulse/dose-rate electron beam irradiation, in the context of preclinical and cognitive studies. In this work, we report on the commissioning and beam monitoring of the Oriatron eRT6 prototype linac. Materials and Methods: The beam was characterized in different steps. The output stability was studied by performing repeated measurements over a period of 20 months. The relative output variations caused by changing beam parameters, such as the temporal electron pulse width, the pulse repetition frequency and the pulse amplitude were also analyzed. Finally, depth dose curves and field sizes were measured for two different beam settings, resulting in one beam with a conventional radiotherapy dose-rate and one with a much higher dose-rate. Measurements were performed with Gaf-chromic EBT3 films and with a PTW Advanced Markus ionization chamber. In addition, we developed a beam current monitoring system based on the signals from an induction torus positioned at the beam exit of the waveguide and from a graphite beam collimator. Results: The stability of the output over repeated measurements was found to be good, with a standard deviation smaller than 1%. However, non-negligible day-to-day variations of the beam output were observed. Those output variations showed different trends depending on the dose-rate. The analysis of the relative output variation as a function of various beam parameters showed that in a given configuration, the dose-rate could be reliably varied over three orders of magnitude. Interdependence effects on the output variation between the parameters were also observed. The beam energy and field size were found to be slightly dose-rate-dependent and suitable mainly for small animal irradiation. The beam monitoring system was able to measure in a reproducible way the total charge of electrons that exit the machine, as long as the electron pulse amplitude remains above a given threshold. Furthermore, we were able to relate the charge measured with the monitoring system to the absorbed dose in a solid water phantom. Conclusion: The Oriatron eRT6 was successfully commissioned for preclinical use and is currently in full operation, with studies being performed on the radiobiological effects of high dose-per-pulse irradiation. (C) 2017 American Association of Physicists in Medicine
引用
收藏
页码:863 / 874
页数:12
相关论文
共 4 条
  • [1] High dose-per-pulse electron beam dosimetry: Usability and dose-rate independence of EBT3 Gafchromic films
    Jaccard, Maud
    Petersson, Kristoffer
    Buchillier, Thierry
    Germond, Jean-Francois
    Duran, Maria Teresa
    Vozenin, Marie-Catherine
    Bourhis, Jean
    Bochud, Francois O.
    Bailat, Claude
    MEDICAL PHYSICS, 2017, 44 (02) : 725 - 735
  • [2] High dose-per-pulse electron beam dosimetry - A model to correct for the ion recombination in the Advanced Markus ionization chamber
    Petersson, Kristoffer
    Jaccard, Maud
    Germond, Jean-Francois
    Buchillier, Thierry
    Bochud, Francois
    Bourhis, Jean
    Vozenin, Marie-Catherine
    Bailat, Claude
    MEDICAL PHYSICS, 2017, 44 (03) : 1157 - 1167
  • [3] Response of diamond detectors in ultra-high dose-per-pulse electron beams for dosimetry at FLASH radiotherapy
    Kranzer, R.
    Schueller, A.
    Bourgouin, A.
    Hackel, T.
    Poppinga, D.
    Lapp, M.
    Looe, H. K.
    Poppe, B.
    PHYSICS IN MEDICINE AND BIOLOGY, 2022, 67 (07)
  • [4] Investigating ultra-high dose rate water radiolysis using the Geant4-DNA toolkit and a Geant4 model of the Oriatron eRT6 electron linac
    Chappuis, Flore
    Tran, Hoang Ngoc
    Jorge, Patrik Goncalves
    Zein, Sara A.
    Kyriakou, Ioanna
    Emfietzoglou, Dimitris
    Bailat, Claude
    Bochud, Francois
    Incerti, Sebastien
    Desorgher, Laurent
    SCIENTIFIC REPORTS, 2024, 14 (01):