Use of a novel platelet function analyzer (PFA-100™) with high sensitivity to disturbances in von willebrand factor to screen for von Willebrand's disease and other disorders

被引:0
|
作者
Favaloro, EJ [1 ]
Facey, D [1 ]
Henniker, A [1 ]
机构
[1] Westmead Hosp, Western Sydney Area Hlth Serv, Inst Clin Pathol & Med Res, Dept Haematol,Diagnost Hemostasis Lab, Westmead, NSW 2145, Australia
关键词
von Willebrand factor; von Willebrand's disease; laboratory assessment; screening; diagnosis;
D O I
10.1002/(SICI)1096-8652(199911)62:3<165::AID-AJH6>3.0.CO;2-C
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
The PFA-100(TM) is a platelet function analyzer which uses whole blood and high shear stress blood flow to simulate primary hemostasis and assess platelet function. A small volume of blood is introduced into a disposable cartridge, and forced through a capillary tube, Platelet adhesion and aggregation is then initiated following exposure to either collagen/ADP [C/ADP] or collagen/epinephrine [C/Epi] coated membranes. Movement of blood through the capillary, and its subsequent occlusion is monitored and yields the measured endpoint (closure time [CT] in seconds). Using two approaches, we assessed the sensitivity of this system to disturbances in the function of von Willebrand Factor (VWF), Firstly, we assessed the ability of the PFA-100(TM) to detect the presence of von Willebrands Disease (VWD), Using normal individuals (N=18), CTs (in seconds; mean [range = mean +/- 2SD]) were (i) C/ADP, 95 [66-124], (ii) C/Epi, 128 [98-158]. A panel of 47 patients undergoing evaluation for clinical hemostatic defects inclusive of VWD were also evaluated. All samples from patients confirmed to have VWD following specific VWF studies [N = 9; 3 x Type 1, 1 x Type 3, 1 x Type 2A, 4 x Type 2B] gave prolonged CTs (greater than or equal to 200 s) for both C/ADP and C/Epi membranes; in contrast, all patients yielding normal CT values were found to yield normal VWF results (i.e., were found not to suffer from VWD), Patients with hemophilia (1 x hemophilia A, 1 x hemophilia C) gave normal PFA-100(TM) CT, while those with clinical thrombocytopaenia (N = 3) gave prolonged PFA-100(TM) CT. A number of other patient samples also gave abnormal CT values which in some cases could be linked to recent aspirin consumption, In the second evaluation process, and using normal blood, we have assessed the ability of various antibodies to influence the CT. Of the monoclonal antibody panel tested [N = 20], only a proportion of those against VWF [6/10] or gp1b/lX [CD42; 2/5] were found to be inhibitory (i.e., prolonged the CT), Data using polyclonal antibodies (against platelets, VWF, fibrinogen and fibronectin) is more complex but largely confirms the sensitivity of the system to VWF. On the basis of these results, we conclude that the PFA-100(TM) is highly sensitive to disturbances in VWF and to the presence of VWD and may thus provide a valuable screening test for VWD in certain specific circumstances (i.e., acute need conditions or remote testing sites; normal CT result generally effective as negative predictor, i.e. not severe VWD), However, since abnormal CT values were obtained in clinical situations other than evident VWD, the PFA-100(TM) cannot be used as a specific diagnostic tool to establish the presence of VWD. Thus, any abnormal PFA-100TM CT result should be thoroughly evaluated by follow-up specific testing to establish the true clinical disorder affecting the individual under investigation, inclusive of appropriate VWF assays if VWD is clinically suspected. (C) 1999 Wiley-Liss, Inc.
引用
收藏
页码:165 / 174
页数:10
相关论文
共 33 条
  • [1] The role of the platelet function analyser (PFA-100™) in the characterization of patients with von Willebrand's disease and its relationships with von Willebrand factor and the ABO blood group
    Nitu-Whalley, IC
    Lee, CA
    Brown, SA
    Riddell, A
    Hermans, C
    HAEMOPHILIA, 2003, 9 (03) : 298 - 302
  • [2] The utility of the PFA-100 in the identification of von Willebrand disease: A concise review
    Favaloro, Emmanuel J.
    SEMINARS IN THROMBOSIS AND HEMOSTASIS, 2006, 32 (05) : 537 - 545
  • [3] Potential supplementary utility of combined PFA-100 and functional von Willebrand factor testing for the laboratory assessment of desmopressin and factor concentrate therapy in von Willebrand disease
    Favaloro, Emmanuel J.
    Thom, Jim
    Patterson, David
    Just, Sarah
    Baccala, Maria
    Dixon, Tracy
    Meiring, Muriel
    Koutts, Jerry
    Rowell, John
    Baker, Ross
    BLOOD COAGULATION & FIBRINOLYSIS, 2009, 20 (06) : 475 - 483
  • [4] The Prevalence of von Willebrand Disease and Significance of in Vitro Bleeding Time (PFA-100) in von Willebrand Disease Screening in the Izmir Region
    Sap, Fatih
    Kavakli, Tulay
    Kavakli, Kaan
    Dizdarer, Ceyhun
    TURKISH JOURNAL OF HEMATOLOGY, 2013, 30 (01) : 40 - 47
  • [5] Utility of the platelet function analyser (PFA-100/200) for exclusion or detection of von Willebrand disease: A study 22 years in the making
    Favaloro, Emmanuel J.
    THROMBOSIS RESEARCH, 2020, 188 : 17 - 24
  • [6] Platelet function analyser (PFA-100) results and von Willebrand factor deficiency: a 16-year 'real-world' experience
    Ardillon, L.
    Ternisien, C.
    Fouassier, M.
    Sigaud, M.
    Lefrancois, A.
    Pacault, M.
    Ribeyrol, O.
    Fressinaud, E.
    Boisseau, P.
    Trossaert, M.
    HAEMOPHILIA, 2015, 21 (05) : 646 - 652
  • [7] High Resolution Multimer Analysis and the PFA-100 Platelet Function Analyser Can Detect von Willebrand Disease Type 2A without a Pathological Ratio of Ristocetin Cofactor Activity and von Willebrand Antigen Level
    Weiss, Dominik R.
    Strasser, Erwin F.
    Ringwald, Juergen
    Zimmermann, Robert
    Eckstein, Reinhold
    CLINICAL LABORATORY, 2012, 58 (11-12) : 1203 - 1209
  • [8] The impact of bleeding history, von Willebrand factor and PFA-100® on the diagnosis of type 1 von Willebrand disease: results from the European study MCMDM-1VWD
    Castaman, Giancarlo
    Tosetto, Alberto
    Goodeve, Anne
    Federici, Augusto B.
    Lethagen, Stefan
    Budde, Ulrich
    Batlle, Javier
    Meyer, Dominique
    Mazurier, Claudine
    Goudemand, Jenny
    Eikenboom, Jeroen
    Schneppenheim, Reinhard
    Ingerslev, Jorgen
    Habart, David
    Hill, Frank
    Peake, Ian
    Rodeghiero, Francesco
    BRITISH JOURNAL OF HAEMATOLOGY, 2010, 151 (03) : 245 - 251
  • [9] Evaluation of the PFA-100 system for monitoring desmopressin therapy in patients with type 1 von Willebrand's disease
    Franchini, M
    Gandini, G
    Manzato, F
    Lippi, G
    HAEMATOLOGICA, 2002, 87 (06) : 670 - 670
  • [10] Sensitive and specific assessment of recombinant von Willebrand factor in platelet function analyzer
    Pekrul, Isabell
    Kragh, Thorsten
    Turecek, Peter L.
    Novack, Aaron R.
    Ott, Helmut W.
    Spannagl, Michael
    PLATELETS, 2019, 30 (02) : 264 - 270