Modified carbon paste electrodes for Cu(II) determination

被引:18
|
作者
Ghaedi, Mehrorang [1 ]
Naderi, Sakineh [1 ]
Montazerozohori, Morteza [1 ]
Sahraei, Reza [2 ]
Daneshfar, Ali [2 ]
Taghavimoghadam, Narges [1 ]
机构
[1] Univ Yasuj, Dept Chem, Yasuj 7591874831, Iran
[2] Ilam Univ, Dept Chem, Ilam, Iran
来源
MATERIALS SCIENCE & ENGINEERING C-MATERIALS FOR BIOLOGICAL APPLICATIONS | 2012年 / 32卷 / 08期
关键词
Carbon paste electrode; Copper ion; Palladium nanoparticle; Cadmium sulfide nanoparticles; Potentiometric sensors; 2,2 '-(1E,1 ' E)-1,1 '-(2,2 '-azanediylbis(ethane-2,1-diyl)bis(azan-1-yl-1-ylidene))bis(ethan-1-yl-1-ylidene)diphenol (ADEZEDP); COBALT(III) SCHIFF-BASE; MEMBRANE ELECTRODES; COPPER-ION; SELECTIVE ELECTRODE; SENSORS; PORPHYRINS; CHEMISTRY; CARRIER;
D O I
10.1016/j.msec.2012.06.014
中图分类号
TB3 [工程材料学]; R318.08 [生物材料学];
学科分类号
0805 ; 080501 ; 080502 ;
摘要
A new Cu2+ carbon paste electrode (CPE) using 2,2'-(1E,1'E)-1,1'-(2,2'-azanediylbis (ethane-2,1-diyl) bis(azan-1-yl-1-ylidene))bis(ethan-1-yl-1-ylidene)diphenol (ADEZEDP) has been prepared. The influence of variables including sodium tetraphenylborate (NaTPB). ionophore, and amount of multiwalled carbon nanotubes (MWCNT). CdO nanowires, CdS nanoparticles and palladium nanoparticles loaded on ADEZEDP and Nujol on the electrodes response were studied and optimized. At optimum values of all variables, for each nanomaterial the electrode response was linear in concentration range of 1.0x10(-8) to 1.0x10(-1) mol L-1 for ADEZEDP with Nernstian slope. The good performance of electrode viz. Wide applicable pH range (2.0-5.0), fast response time (approximate to 6 s). and adequate life time (3 months) indicate the utility of the proposed electrodes for evaluation of Cu2+ ion content in various situations. Finally, these electrodes have been successfully applied for the determination of Cu2+ ions content in various real samples. The selectivity of proposed electrode was evaluated by separation solution method and fixed interference method. (C) 2012 Elsevier B.V. All rights reserved.
引用
收藏
页码:2274 / 2279
页数:6
相关论文
共 50 条
  • [1] Sorption and determination of Hg(II) on clay modified carbon paste electrodes
    Kula, P
    Navrátilová, Z
    Kulová, P
    Kotoucek, M
    ANALYTICA CHIMICA ACTA, 1999, 385 (1-3) : 91 - 101
  • [2] ALCOHOL DETERMINATION BY MODIFIED CARBON PASTE ELECTRODES
    BILITEWSKI, U
    SCHMID, RD
    BIOSENSORS : APPLICATIONS IN MEDICINE, ENVIRONMENTAL PROTECTION AND PROCESS CONTROL, 1989, 13 : 99 - 102
  • [3] DETERMINATION OF IRON(II) WITH CHEMICALLY-MODIFIED CARBON-PASTE ELECTRODES
    GAO, ZQ
    LI, PB
    ZHAO, ZF
    TALANTA, 1991, 38 (10) : 1177 - 1184
  • [4] Preconcentration and voltammetric determination of palladium (II) at sodium humate modified carbon paste electrodes
    Sun, QY
    Wang, CM
    Li, LX
    Li, HL
    FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY, 1999, 363 (01): : 114 - 117
  • [5] Preconcentration and voltammetric determination of palladium (II) at sodium humate modified carbon paste electrodes
    Qiaoyu Sun
    C. Wang
    Liangxi Li
    Hulin Li
    Fresenius' Journal of Analytical Chemistry, 1999, 363 : 114 - 117
  • [6] VOLTAMMETRIC DETERMINATION OF CELIPTIUM WITH CARBON PASTE AND LIPID-MODIFIED CARBON PASTE ELECTRODES
    ARCOS, J
    KAUFFMANN, JM
    PATRIARCHE, GJ
    SANCHEZBATANERO, P
    ANALYTICA CHIMICA ACTA, 1990, 236 (02) : 299 - 305
  • [7] Electrochemical Behavior and Determination of Rutin on Modified Carbon Paste Electrodes
    Macikova, Pavla
    Halouzka, Vladimir
    Hrbac, Jan
    Bartak, Petr
    Skopalova, Jana
    SCIENTIFIC WORLD JOURNAL, 2012,
  • [8] Maize tassel-modified carbon paste electrode for voltammetric determination of Cu(II)
    Moyo, Mambo
    Okonkwo, Jonathan O.
    Agyei, Nana M.
    ENVIRONMENTAL MONITORING AND ASSESSMENT, 2014, 186 (08) : 4807 - 4817
  • [9] A Comparison of Variously Modified Carbon Paste Electrodes for Determination of Rutin
    Macikova, Pavla
    Skopalova, Jana
    MODERN ELECTROCHEMICAL METHODS XXX, 2010, : 101 - 105
  • [10] Pyranose oxidase modified carbon paste electrodes for monosaccharide determination
    Liden, H
    Volc, J
    Marko-Varga, G
    Gorton, L
    ELECTROANALYSIS, 1998, 10 (04) : 223 - 230