Slice sensitivity profiles and pixel noise of multi-slice CT in comparison with single-slice CT.

被引:18
作者
Schorn, C [1 ]
Obenauer, S [1 ]
Funke, M [1 ]
Hermann, KP [1 ]
Kopka, L [1 ]
Grabbe, E [1 ]
机构
[1] Univ Gottingen, Abt Rontgendiagnost 1, D-37075 Gottingen, Germany
来源
ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN | 1999年 / 171卷 / 03期
关键词
Helical computed tomography (CT); CT multi-slice; CT technology; CT image quality;
D O I
10.1055/s-1999-244
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: Presentation and evaluation of slice sensitivity profil and pixel noise of multi-slice CT in comparison to single-slice CT. Methods: Slice sensitivity profiles and pixel noise of a multi-slice CT equiped with a 2D matrix detector array and of a single-slice CT were evaluated in phantom studies. Results: For the single-slice CT the width of the slice sensitivity profiles increased with increasing pitch. In spite of a much higher table speed the slice sensitivity profiles of multi-slice CT were narrower and did not increase with higher pitch. Noise in single-slice CT was independent of pitch. For multi-slice CT noise increased with higher pitch and for the higher pitch decreased slightly with higher detector row collimation. Conclusions: Multi-slice CT provides superior z-resolution and higher volume coverage speed. These qualities fulfill one of the prerequisites for improvement of 3D postprocessing.
引用
收藏
页码:219 / 225
页数:7
相关论文
共 19 条
[1]  
[Anonymous], RADIOLOGY
[2]  
BELAND LL, 1998, RADIOLOGY, V209, P327
[3]   SPIRAL CT - DECREASED SPATIAL-RESOLUTION INVIVO DUE TO BROADENING OF SECTION-SENSITIVITY PROFILE [J].
BRINK, JA ;
HEIKEN, JP ;
BALFE, DM ;
SAGEL, SS ;
DICROCE, J ;
VANNIER, MW .
RADIOLOGY, 1992, 185 (02) :469-474
[4]   SPIRAL CT OF THE THORAX WITH REDUCED VOLUME OF CONTRAST MATERIAL - A COMPARATIVE-STUDY [J].
COSTELLO, P ;
DUPUY, DE ;
ECKER, CP ;
TELLO, R .
RADIOLOGY, 1992, 183 (03) :663-666
[5]   DYNAMIC HEPATIC CT [J].
FOLEY, WD .
RADIOLOGY, 1989, 170 (03) :617-622
[6]   COMPUTERIZED TRANSVERSE AXIAL SCANNING (TOMOGRAPHY) .1. DESCRIPTION OF SYSTEM [J].
HOUNDFIELD, GN .
BRITISH JOURNAL OF RADIOLOGY, 1973, 46 (552) :1016-1022
[7]   Multi-slice helical CT: Scan and reconstruction [J].
Hu, H .
MEDICAL PHYSICS, 1999, 26 (01) :5-18
[8]   SPIRAL VOLUMETRIC CT WITH SINGLE-BREATH-HOLD TECHNIQUE, CONTINUOUS TRANSPORT, AND CONTINUOUS SCANNER ROTATION [J].
KALENDER, WA ;
SEISSLER, W ;
KLOTZ, E ;
VOCK, P .
RADIOLOGY, 1990, 176 (01) :181-183
[9]   PHYSICAL PERFORMANCE-CHARACTERISTICS OF SPIRAL CT SCANNING [J].
KALENDER, WA ;
POLACIN, A .
MEDICAL PHYSICS, 1991, 18 (05) :910-915
[10]  
KALENDER WA, 1998, Z MED PHYS, V8, P7