Low Dead Volume Integrated Separation Columns and Fabry-Perot Sensors for Micro-Gas Chromatography

被引:0
|
作者
Reddy, Karthik [1 ,2 ]
Liu, Jing [1 ]
Oo, Maung Kyaw Khaing [1 ]
Fan, Xudong [1 ]
机构
[1] Univ Michigan, Dept Biomed Engn, 1101 Beal Ave, Ann Arbor, MI 48109 USA
[2] Univ Michigan, Dept Elect Engn & Comp Sci, Ann Arbor, MI 48109 USA
来源
MICROFLUIDICS, BIOMEMS, AND MEDICAL MICROSYSTEMS XI | 2013年 / 8615卷
基金
美国国家科学基金会;
关键词
Micro-gas chromatography; vapor sensing; Fabry-Perot cavity; volatile organic compounds;
D O I
10.1117/12.2003100
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
We developed a monolithic subsystem that integrates a micro-gas chromatography (mu GC) separation column and on-column, non-destructive Fabry-Perot (FP) vapor sensors on a single silicon chip. The device was fabricated using deep reactive ion etching of silicon to create fluidic channels and polymers were deposited on the same silicon chip to act as a stationary phase or an FP sensor, thus avoiding dead volumes caused by the interconnects between the column and sensor traditionally used in mu GC. Two integration designs were studied. In the first design, the mu GC column was coated with a layer of polymer that served as both the stationary phase and the FP sensor, which has the greatest level of integration. In the second design, a FP sensor array spray-coated with different vapor sensing polymers was integrated with the mu GC column, which significantly improves the system flexibility and detection sensitivity. With this design, we show that the FP sensors have a detection limit on the order of tens of picograms with a sub-second response time. Furthermore, FP sensor array are shown to respond to a mixture of analytes separated by the integrated separation channel, allowing for the construction of response patterns, which, along with retention time, can be used as a basis of analyte identification.
引用
收藏
页数:9
相关论文
共 11 条
  • [1] On-chip Fabry-Perot interferometric sensors for micro-gas chromatography detection
    Reddy, Karthik
    Guo, Yunbo
    Liu, Jing
    Lee, Wonsuk
    Oo, Maung Kyaw Khaing
    Fan, Xudong
    SENSORS AND ACTUATORS B-CHEMICAL, 2011, 159 (01): : 60 - 65
  • [2] Integrated Separation Columns and Fabry-Perot Sensors for Microgas Chromatography Systems
    Reddy, Karthik
    Liu, Jing
    Oo, Maung Kyaw Khaing
    Fan, Xudong
    JOURNAL OF MICROELECTROMECHANICAL SYSTEMS, 2013, 22 (05) : 1174 - 1179
  • [3] A portable micro-gas chromatography with integrated photonic crystal slab sensors on chip
    Biswas P.
    Zhang C.
    Chen Y.
    Liu Z.
    Vaziri S.
    Zhou W.
    Sun Y.
    Biosensors, 2021, 11 (09):
  • [4] A Portable Micro-Gas Chromatography with Integrated Photonic Crystal Slab Sensors on Chip
    Biswas, Priyanka
    Zhang, Chen
    Chen, Yudong
    Liu, Zhonghe
    Vaziri, Seyedmohsen
    Zhou, Weidong
    Sun, Yuze
    BIOSENSORS-BASEL, 2021, 11 (09):
  • [5] Rapid, sensitive, and multiplexed on-chip optical sensors for micro-gas chromatography
    Reddy, Karthik
    Guo, Yunbo
    Liu, Jing
    Lee, Wonsuk
    Oo, Maung Kyaw Khaing
    Fan, Xudong
    LAB ON A CHIP, 2012, 12 (05) : 901 - 905
  • [6] High-Q Fabry-Perot Micro-Cavities for High-Sensitivity Volume Refractometry
    Gaber, Noha
    Sabry, Yasser M.
    Erfan, Mazen
    Marty, Frederic
    Bourouina, Tarik
    MICROMACHINES, 2018, 9 (02):
  • [7] High-Gain Low-Profile Substrate-Integrated Fabry-Perot Cavity Antenna
    Liu, Zhiming
    Liu, Shaobin
    Kong, Xiangkun
    Chen, Xinlei
    Liu, Xiaochun
    Sun, Shining
    2018 ASIA-PACIFIC MICROWAVE CONFERENCE PROCEEDINGS (APMC), 2018, : 1387 - 1389
  • [8] Review on micro-gas chromatography system for analysis of multiple low-concentration volatile organic compounds: preconcentration, separation, detection, integration, and challenges
    Lee, Yeongseok
    Son, Hyeonwoo
    Lee, Junwoo
    Lim, Si-Hyung
    MICRO AND NANO SYSTEMS LETTERS, 2024, 12 (01)
  • [9] Compact Powell-Lens-Based Low-Coherence Correlation Interrogation System for Fiber-Optic Fabry-Perot Sensors
    Ma, Zhibo
    Guo, Tongxin
    Zhang, Tianyang
    Wang, Wei
    Chen, Haibin
    Zhang, Xiongxing
    Yuan, Weizheng
    IEEE PHOTONICS JOURNAL, 2019, 11 (04):
  • [10] Graphene Integrated in a Ring Type Fabry-Perot Cavity: Polarization Insensitive, Low Voltage and Tunable Modulation of Light at Near-IR Telecommunications Band
    Jafari, Behnam
    Soofi, Hadi
    IEEE TRANSACTIONS ON NANOTECHNOLOGY, 2021, 20 : 425 - 433