High intensity focused ultrasound ablation of kidney guided by MRI

被引:47
作者
Damianou, C
Pavlou, M
Velev, O
Kyriakou, K
Trimikliniotis, M
机构
[1] FIT, Limmasol, Cyprus
[2] Poliklin Igia, Limmasol, Cyprus
[3] Intercoll, Limmasol, Cyprus
关键词
ultrasound; kidney; MRI; lesion; cortex; cavitation;
D O I
10.1016/j.ultrasmedbio.2003.10.018
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
The effectiveness of magnetic resonance imaging (MRI) to monitor therapeutic protocols of high-intensity focused ultrasound (HIFU), in freshly excised pig kidney cortex is investigated. For high quality imaging, the pulse sequence fast spin echo (FSE) T1- and T2-weighted, and proton density were evaluated. For fast imaging, the pulse sequence T1-weighted fast spoiled gradient (FSPGR) was used. The main goal was to evaluate the MRI detection of large lesions (bigger than 1 cm x 1 cm x 1 cm) that is achieved by moving the transducer in a predetermined pattern. The contrast between lesion and kidney tissue is excellent with either T1-weighted or T2-weighted FSE. With T1-weighted FSE, the best contrast is observed for recovery time (TR) between 200 ms and 400 ms. With T2-weighted FSE best contrast can be achieved for echo time (TE) between 16 and 32 ms. T2-weighted FSE was proven as the best pulse sequence to detect cavitational activity. This advantage is attributed to the significant difference in signal intensity between air spaces and necrotic tissue. Air spaces appear brighter than thermal lesions. Therefore, for therapeutic protocols created using cavitational mode, T2-weighted FSE may be the optimum pulse sequence to use. The proton density pulse sequence does not provide any advantage over the T1- and T2-weighted pulse sequences. Using T1-weighted FSPGR, acquisition time as low as 5 s could be achieved. Good contrast and signal-to-noise ratio (SNR) are achieved with TR = 100 ms and flip angle between 75 to 90degrees. The above techniques were very successful in detecting large lesion volumes. (C) 2004 World Federation for Ultrasound in Medicine Biology.
引用
收藏
页码:397 / 404
页数:8
相关论文
共 52 条
[31]   Temperature mapping using the water proton chemical shift: A chemical shift selective phase mapping method [J].
Kuroda, K ;
Oshio, K ;
Chung, AH ;
Hynynen, K ;
Jolesz, FA .
MAGNETIC RESONANCE IN MEDICINE, 1997, 38 (05) :845-851
[32]   Calibration of water proton chemical shift with temperature for noninvasive temperature imaging during focused ultrasound surgery [J].
Kuroda, K ;
Chung, AH ;
Hynynen, K ;
Jolesz, FA .
JMRI-JOURNAL OF MAGNETIC RESONANCE IMAGING, 1998, 8 (01) :175-181
[33]  
Law P, 2000, J MAGN RESON IMAGING, V12, P565, DOI 10.1002/1522-2586(200010)12:4<565::AID-JMRI8>3.0.CO
[34]  
2-4
[35]   TEMPERATURE MAPPING WITH MR IMAGING OF MOLECULAR-DIFFUSION - APPLICATION TO HYPERTHERMIA [J].
LEBIHAN, D ;
DELANNOY, J ;
LEVIN, RL .
RADIOLOGY, 1989, 171 (03) :853-857
[36]  
MATSUMOTO R, 1992, J MAGN RESON IMAGING, V2, P255
[37]   MRI monitoring of the thermal ablation of tissue: Effects of long exposure times [J].
McDannold, N ;
Hynynen, K ;
Jolesz, F .
JOURNAL OF MAGNETIC RESONANCE IMAGING, 2001, 13 (03) :421-427
[38]   Determination of the optimal delay between sonications during focused ultrasound surgery in rabbits by using MR imaging to monitor thermal buildup in vivo [J].
McDannold, NJ ;
Jolesz, FA ;
Hynynen, KH .
RADIOLOGY, 1999, 211 (02) :419-426
[39]   Drain edema development after MRI-guided focused ultrasound treatment [J].
Mórocz, IA ;
Hynynen, K ;
Gudbjartsson, H ;
Peled, S ;
Colucci, V ;
Jólesz, FA .
JOURNAL OF MAGNETIC RESONANCE IMAGING, 1998, 8 (01) :136-142
[40]   Laser therapy for breast cancer: MR imaging and histopathologic correlation [J].
Mumtaz, H ;
HallCraggs, MA ;
Wotherspoon, A ;
Paley, M ;
Buonaccorsi, G ;
Amin, Z ;
Wilkinson, I ;
Kissin, MW ;
Davidson, I ;
Taylor, I ;
Bown, SG .
RADIOLOGY, 1996, 200 (03) :651-658