Soft Contact-lens Biosensor for Real-time Tear Sugar Monitoring at the Eye

被引:0
作者
Kudo, Hiroyuki [1 ]
Chu, MingXing [1 ]
Miyajima, Kumiko [1 ]
Arakawa, Takahiro [1 ]
Mitsubayashi, Kohji [1 ]
Hiranuma, Yoshitaka [2 ]
Hiramatsu, Hideo [2 ]
机构
[1] Tokyo Med & Dent Univ, Inst Biomat & Bioengn, Tokyo, Japan
[2] Nihon Univ, Grad Sch Ind Technol, Chiba, Japan
来源
PROCEEDINGS 2012 IEEE INTERNATIONAL CONFERENCE ON SYSTEMS, MAN, AND CYBERNETICS (SMC) | 2012年
基金
日本学术振兴会;
关键词
glucose; biosensor; monitoring; continuous; tear; GLUCOSE;
D O I
暂无
中图分类号
TP3 [计算技术、计算机技术];
学科分类号
0812 ;
摘要
A soft contact-lens (SCL) biosensor for in-situ tear sugar monitoring was fabricated and tested. In contrast to the previously reported continuous glucose monitoring (CGM) devices, we focused on the relationship between blood glucose and tear glucose. The SCL biosensor was constructed by applying microelectromechanical systems (MEMS) techniques to functional polymers. The sensor has film electrodes on the surface of a rounded surface of polydimethyl siloxane (PDMS) contact-lens and glucose oxidase (GOD) was immobilized on the sensing region of the electrodes. Sufficiently adhesive and flexible electrodes were formed on the polymer device (PDMS and 2-methacryloyloxyethyl phosphorylcholine (MPC) polymer). In the in-vitro characterization, the SCL-biosensor showed excellent relationship between the output current and glucose concentration from 0.03 to 10.0 mmol/L, which included the reported tear glucose concentrations (0.14 mM) of humans. Based on the result, in-situ tear glucose monitoring with the SCL-biosensor was carried out. The output current of the SCL-biosensor was sufficiently stable and reflected the tear glucose levels. The static glucose level in the basal tear of a rabbit was measured and estimated to 0.13 mmol/L. Since the biosensor was also possible to be sotored for more than 50 days, daily disposable application was expected with the SCL biosensor.
引用
收藏
页码:2048 / 2051
页数:4
相关论文
共 9 条
  • [1] Badugu R., J FLUORESCENCE, V13, P371
  • [2] Chen R, 1996, J CAPILLARY ELECTROP, V3, P243
  • [3] DAUM KM, 1982, INVEST OPHTH VIS SCI, V22, P509
  • [4] Determination of glucose in submicroliter samples by CE-LIF using precolumn or on-column enzymatic reactions
    Jin, Z
    Chen, R
    Colon, LA
    [J]. ANALYTICAL CHEMISTRY, 1997, 69 (07) : 1326 - 1331
  • [5] EFFECTS OF TEMPERATURE, RELATIVE HUMIDITY, AND MODE OF BREATHING ON CANINE AIRWAY SECRETIONS
    MAN, SFP
    ADAMS, GK
    PROCTOR, DF
    [J]. JOURNAL OF APPLIED PHYSIOLOGY, 1979, 46 (02) : 205 - 210
  • [6] ANALYSIS OF METABOLITES IN SWEAT AS A MEASURE OF PHYSICAL CONDITION
    MITSUBAYASHI, K
    SUZUKI, M
    TAMIYA, E
    KARUBE, I
    [J]. ANALYTICA CHIMICA ACTA, 1994, 289 (01) : 27 - 34
  • [7] In vivo glucose monitoring: the clinical reality and the promise
    Pickup, JC
    Hussain, F
    Evans, ND
    Sachedina, N
    [J]. BIOSENSORS & BIOELECTRONICS, 2005, 20 (10) : 1897 - 1902
  • [8] Romano A, 1988, Metab Pediatr Syst Ophthalmol (1985), V11, P78
  • [9] TEAR GLUCOSE-LEVELS IN NORMAL PEOPLE AND IN DIABETIC-PATIENTS
    SEN, DK
    SARIN, GS
    [J]. BRITISH JOURNAL OF OPHTHALMOLOGY, 1980, 64 (09) : 693 - 695