A Sutureless Aortic Stent-Graft Based on a Nitinol Scaffold Bonded to a Compliant Nanocomposite Polymer Is Durable for 10 Years in a Simulated In Vitro Model

被引:16
作者
Desai, Mital [1 ]
Bakhshi, Raheleh [2 ]
Zhou, Xiang [3 ]
Odlyha, Marianne [4 ]
You, Zhong [3 ]
Seifalian, Alexander M. [1 ,2 ]
Hamilton, George [1 ,2 ]
机构
[1] Royal Free London NHS Fdn Trust, Dept Vasc Surg, London NW3 2QG, England
[2] UCL, Ctr Nanotechnol & Regenerat Med, UCL Div Surg & Intervent Sci, London WC1E 6BT, England
[3] Univ Oxford, Dept Engn Sci, Oxford OX1 2JD, England
[4] Univ London Birkbeck Coll, Dept Chem Mat, London WC1E 7HX, England
基金
英国工程与自然科学研究理事会;
关键词
endovascular aneurysm repair; stent-graft; nanocomposite polymer; nitinol; fatigue testing; flow circuit; polyhedral oligomeric silsesquioxane; poly(carbonate-urea) urethane; thermal resistance; compliance; SILSESQUIOXANE NANOCOMPOSITES; III ENDOLEAK; FAILURE; ANEURYSM; REPAIR; ENDOTHELIALIZATION; POLYURETHANES; IMPLANTS; DEVICES; IMPACT;
D O I
10.1583/11-3740MR.1
中图分类号
R61 [外科手术学];
学科分类号
摘要
Purpose: To physiologically test the durability of a sutureless aortic stent-graft based on nitinol bonded to polyhedral oligomeric silsesquioxane (POSS) and poly(carbonate-urea) urethane (PCU) for 10 years according to Food and Drug Administration guidelines. Methods: Aortic stent-grafts (n=4) were tested in 37 degrees C distilled water using simulated in vivo hydrodynamic pulse loading. After 400 million cycles, surface topography was assessed by scanning electron microscopy (SEM) and Fourier transform infrared (FTIR) spectroscopy. Dynamic compliance was measured using a pulsatile flow phantom. Mechanical and elastic properties were determined by stress-strain studies and elastic deformation tests. Dynamic scanning calorimetry (DSC) and thermomechanical analysis (TMA) were used to assess thermal resistance. Comparison was made with a zero-cycled control. Results: All stent-grafts successfully completed accelerated pulsatile fatigue at 94 +/- 14-mmHg pulse pressure. SEM images confirmed uniform surface topography without any fractures. FTIR showed increased intensity of -NHCO- bonds, but there was no significant sign of biodegradation. Tensile stress of fatigue-tested polymer compared favorably with the zero-cycled control at 50% to 500% strain (p=0.69). At a mean pressure range of 60 to 120 mmHg, overall compliance of the fatigue-tested grafts was 3.48 +/- 1.27%mmHg(-1)x10(-2) with no significant difference compared to control (3.26 +/- 0.65%mmHg(-1)x10(-2); p=0.47). DSC and TMA showed comparable thermotropic transition. Conclusion: Simulated physiological in vivo hydrodynamic loading has no significant degradative effect on an innovative sutureless stent-graft made from POSS-PCU nanocomposite polymer. Sutureless technology incorporating nitinol stents proved to be robust, with no separation over an accelerated 10-year cycle, which may allow development of durable stent-grafts with better compliance. J Endovasc Ther. 2012;19:415-427
引用
收藏
页码:415 / 427
页数:13
相关论文
共 30 条
  • [1] Polymeric coating of surface modified nitinol stent with POSS-nanocomposite polymer
    Bakhshi, Raheleh
    Darbyshire, Arnold
    Evans, James Eaton
    You, Zhong
    Lu, Jian
    Seifalian, Alexander M.
    [J]. COLLOIDS AND SURFACES B-BIOINTERFACES, 2011, 86 (01) : 93 - 105
  • [2] Beebe HG, 2001, J VASC SURG, V33, pS55
  • [3] Late type III endoleak and graft failure of an Ancure stent-graft
    Brown, Katherine E.
    Heyer, Kamaldeep S.
    Matsumura, Jon S.
    Eskandari, Mark K.
    [J]. JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2008, 19 (10) : 1506 - 1508
  • [4] Durability of Endovascular Infrarenal Aneurysm Repair: When Does Late Failure Occur and Why?
    Chuter, Timothy A. M.
    [J]. SEMINARS IN VASCULAR SURGERY, 2009, 22 (02) : 102 - 110
  • [5] In situ endothelialisation potential of a biofunctionalised nanocomposite biomaterial-based small diameter bypass graft
    de Mel, Achala
    Punshon, Geoffrey
    Ramesh, Bala
    Sarkar, Sandip
    Darbyshire, Arnold
    Hamilton, George
    Seifalian, Alexander M.
    [J]. BIO-MEDICAL MATERIALS AND ENGINEERING, 2009, 19 (4-5) : 317 - 331
  • [6] Endovascular versus Open Repair of Abdominal Aortic Aneurysm
    Greenhalgh, Roger M.
    Brown, Louise C.
    Powell, Janet T.
    Thompson, Simon G.
    Epstein, David
    Sculpher, Mark J.
    [J]. NEW ENGLAND JOURNAL OF MEDICINE, 2010, 362 (20) : 1863 - 1871
  • [7] Guidoin R, 2000, J ENDOVASC THER, V7, P105, DOI 10.1583/1545-1550(2000)007<0105:FGAEAO>2.3.CO
  • [8] 2
  • [9] Hancock S., 2005, Pat., Patent No. [WO 2005070998, 2005070998]
  • [10] Mechanical failure of prosthetic human implants: A 10-year experience with aortic stent graft devices
    Jacobs, TS
    Won, J
    Gravereaux, EC
    Faries, PL
    Morrissey, N
    Teodorescu, VJ
    Hollier, LH
    Marin, ML
    [J]. JOURNAL OF VASCULAR SURGERY, 2003, 37 (01) : 16 - 26