Influence of source-detector separation on diffuse correlation spectroscopy measurements of cerebral blood flow with a multilayered analytical model

被引:13
|
作者
Zhao, Hongting [1 ,2 ]
Buckley, Erin M. [1 ,2 ,3 ,4 ]
机构
[1] Georgia Inst Technol, Wallace H Coulter Dept Biomed Engn, Atlanta, GA 30332 USA
[2] Emory Univ, Atlanta, GA 30322 USA
[3] Emory Univ, Sch Med, Dept Pediat, Atlanta, GA 30322 USA
[4] Childrens Res Scholar, Childrens Healthcare Atlanta, Atlanta, GA 30322 USA
基金
美国国家卫生研究院;
关键词
diffuse correlation spectroscopy; cerebral blood flow; THICKNESS;
D O I
10.1117/1.NPh.9.3.035002
中图分类号
Q189 [神经科学];
学科分类号
071006 ;
摘要
Significance: Diffuse correlation spectroscopy (DCS) is an emerging noninvasive optical technology for bedside monitoring of cerebral blood flow. However, extracerebral hemodynamics can significantly influence DCS estimations of cerebral perfusion. Advanced analytical models can be used to remove the contribution of extracerebral hemodynamics; however, these models are highly sensitive to measurement noise. There is a need for an empirical determination of the optimal source-detector separation(s) (SDS) that improves the accuracy and reduces sensitivity to noise in the estimation of cerebral blood flow with these models. Aim: To determine the influence of SDS on solution uniqueness, measurement accuracy, and sensitivity to inaccuracies in model parameters when using the three-layer model to estimate cerebral blood flow with DCS. Approach: We performed a series of in silico simulations on samples spanning a wide range of physiologically-relevant layer optical properties, thicknesses, and flow. Data were simulated at SDS ranging from 0.5 to 3.0 cm using the three-layer solution to the correlation diffusion equation (with and without noise added) and using three-layer slab Monte Carlo simulations. We quantified the influence of SDS on uniqueness, accuracy, and sensitivity to inaccuracies in model parameters using the three-layer inverse model. Results: Two SDS are required to ensure a unique solution of cerebral blood flow index (CBFi). Combinations of 0.5/1.0/1.5 and 2.5 cm provide the optimal choice for balancing the depth penetration with signal-to-noise ratio to minimize the error in CBFi across a wide range of samples with varying optical properties, thicknesses, and dynamics. Conclusions: These results suggest that the choice of SDS is critical for minimizing the estimated error of cerebral blood flow when using the three-layer model to analyze DCS data. (C) The Authors. Published by SPIE under a Creative Commons Attribution 4.0 International License.
引用
收藏
页数:15
相关论文
共 50 条
  • [1] Interferometric diffuse correlation spectroscopy improves measurements at long source-detector separation and low photon count rate
    Robinson, Mitchell B.
    Boas, David A.
    Sakadzic, Sava
    Franceschini, Maria Angela
    Carp, Stefan A.
    JOURNAL OF BIOMEDICAL OPTICS, 2020, 25 (09)
  • [2] Time-domain diffuse correlation spectroscopy at large source detector separation for cerebral blood flow recovery
    Mogharari, Neda
    Wojtkiewicz, Stanislaw
    Borycki, Dawid
    Liebert, Adam
    Kacprzak, Michal
    BIOMEDICAL OPTICS EXPRESS, 2024, 15 (07): : 4330 - 4344
  • [3] Influence of oversimplifying the head anatomy on cerebral blood flow measurements with diffuse correlation spectroscopy
    Zhao, Hongting
    Buckley, Erin M.
    NEUROPHOTONICS, 2023, 10 (01)
  • [4] In vivo time-gated diffuse correlation spectroscopy at quasi-null source-detector separation
    Pagliazzi, M.
    Sekar, S. Konugolu Venkata
    Di Sieno, L.
    Colombo, L.
    Durduran, T.
    Contini, D.
    Torricelli, A.
    Pifferi, A.
    Dalla Mora, A.
    OPTICS LETTERS, 2018, 43 (11) : 2450 - 2453
  • [5] Optimization of source-detector separation for non-invasive regional cerebral blood flow sensing
    Han, Guang
    Feng, Hao
    Chen, Siqi
    Guo, Qianbei
    Wang, Huiquan
    INFRARED PHYSICS & TECHNOLOGY, 2021, 117
  • [6] Accuracy of diffuse correlation spectroscopy measurements of cerebral blood flow when using a three-layer analytical model
    Zhao, Hongting
    Sathialingam, Eashani
    Buckley, Erin M.
    BIOMEDICAL OPTICS EXPRESS, 2021, 12 (11) : 7149 - 7161
  • [7] Influence of skin blood flow and source-detector distance on near-infrared spectroscopy-determined cerebral oxygenation in humans
    Hirasawa, Ai
    Yanagisawa, Shintaro
    Tanaka, Naoki
    Funane, Tsukasa
    Kiguchi, Masashi
    Sorensen, Henrik
    Secher, Niels H.
    Ogoh, Shigehiko
    CLINICAL PHYSIOLOGY AND FUNCTIONAL IMAGING, 2015, 35 (03) : 237 - 244
  • [8] Small separation diffuse correlation spectroscopy for measurement of cerebral blood flow in rodents
    Sathialingam, Eashani
    Lee, Seung Yup
    Sanders, Bharat
    Park, Jaekeun
    McCracken, Courtney E.
    Bryan, Leah
    Buckley, Erin M.
    BIOMEDICAL OPTICS EXPRESS, 2018, 9 (11): : 5719 - 5734
  • [9] Diffuse Correlation Spectroscopy at Short Source-Detector Separations: Simulations, Experiments and Theoretical Modeling
    Vishwanath, Karthik
    Zanfardino, Sara
    APPLIED SCIENCES-BASEL, 2019, 9 (15):
  • [10] Comparison of diffuse correlation spectroscopy analytical models for measuring cerebral blood flow in adults
    Zhao, Hongting
    Sathialingam, Eashani
    Cowdrick, Kyle R.
    Urner, Tara
    Lee, Seung Yup
    Bai, Shasha
    Akbik, Feras
    Samuels, Owen B.
    Kandiah, Prem
    Sadan, Ofer
    Buckley, Erin M.
    JOURNAL OF BIOMEDICAL OPTICS, 2023, 28 (12)