Microfluidic Analyzer Enabling Quantitative Measurements of Specific Intracellular Proteins at the Single-Cell Level

被引:2
|
作者
Liu, Lixing [1 ,2 ]
Fan, Beiyuan [1 ,2 ]
Wang, Diancan [3 ]
Li, Xiufeng [1 ,2 ]
Song, Yeqing [3 ]
Zhang, Ting [1 ,2 ]
Chen, Deyong [1 ,2 ]
Wang, Yixiang [3 ]
Wang, Junbo [1 ,2 ]
Chen, Jian [1 ,2 ]
机构
[1] Chinese Acad Sci, Inst Elect, State Key Lab Transducer Technol, Beijing 100190, Peoples R China
[2] Univ Chinese Acad Sci, Beijing 100190, Peoples R China
[3] Peking Univ, Sch Stomatol, Beijing 10081, Peoples R China
来源
MICROMACHINES | 2018年 / 9卷 / 11期
基金
中国国家自然科学基金;
关键词
instrumentation; microfluidic flow cytometry; intracellular proteins; absolute quantification; FLOW-CYTOMETRY; HETEROGENEITY;
D O I
10.3390/mi9110588
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
This paper presents a microfluidic instrument capable of quantifying single-cell specific intracellular proteins, which are composed of three functioning modules and two software platforms. Under the control of a LabVIEW platform, a pressure module flushed cells stained with fluorescent antibodies through a microfluidic module with fluorescent intensities quantified by a fluorescent module and translated into the numbers of specific intracellular proteins at the single-cell level using a MATLAB platform. Detection ranges and resolutions of the analyzer were characterized as 896.78-6.78 x 10(5) and 334.60 nM for Alexa 488, 314.60-2.11 x 10(5) and 153.98 nM for FITC, and 77.03-5.24 x 10(4) and 37.17 nM for FITC-labelled anti-beta-actin antibodies. As a demonstration, the numbers of single-cell beta-actins of two paired oral tumor cell types and two oral patient samples were quantified as: 1.12 +/- 0.77 x 10(6)/cell (salivary adenoid cystic carcinoma parental cell line (SACC-83), n(cell) = 13,689) vs. 0.90 +/- 0.58 x 10(5)/cell (salivary adenoid cystic carcinoma lung metastasis cell line (SACC-LM), n(cell) = 15,341); 0.89 +/- 0.69 x 10(6)/cell (oral carcinoma cell line (CAL 27), n(cell) = 7357) vs. 0.93 +/- 0.69 x 10(6)/cell (oral carcinoma lymphatic metastasis cell line (CAL 27-LN2), n(cell) = 6276); and 0.86 +/- 0.52 x 10(6)/cell (patient I) vs. 0.85 +/- 0.58 x 10(6)/cell (patient II). These results (1) validated the developed analyzer with a throughput of 10 cells/s and a processing capability of similar to 10,000 cells for each cell type, and (2) revealed that as an internal control in cell analysis, the expressions of beta-actins remained stable in oral tumors with different malignant levels.
引用
收藏
页数:9
相关论文
共 50 条
  • [1] A microfluidic flow cytometer enabling absolute quantification of single-cell intracellular proteins
    Li, Xiufeng
    Fan, Beiyuan
    Cao, Shanshan
    Chen, Deyong
    Zhao, Xiaoting
    Men, Dong
    Yue, Wentao
    Wang, Junbo
    Chen, Jian
    LAB ON A CHIP, 2017, 17 (18) : 3129 - 3137
  • [2] DEVELOPMENT OF DROPLET MICROFLUIDICS ENABLING QUANTITATIVE MEASUREMENTS OF MULTIPLE PROTEIONS AT SINGLE-CELL LEVEL
    Yang, Hongyu
    Yang, Guang
    Zhang, Ting
    Chen, Deyong
    Wang, Junbo
    Chen, Jian
    2022 IEEE 35TH INTERNATIONAL CONFERENCE ON MICRO ELECTRO MECHANICAL SYSTEMS CONFERENCE (MEMS), 2022, : 872 - 875
  • [3] Microfluidic electroporation for selective release of intracellular molecules at the single-cell level
    Bao, Ning
    Wang, Jun
    Lu, Chang
    ELECTROPHORESIS, 2008, 29 (14) : 2939 - 2944
  • [4] Microfluidic single-cell analysis of intracellular compounds
    Chao, Tzu-Chiao
    Ros, Alexandra
    JOURNAL OF THE ROYAL SOCIETY INTERFACE, 2008, 5 (S139-S150) : S139 - S150
  • [5] Microelectrophoretic single-cell measurements with microfluidic devices
    Sibbitts, Jay
    Sadeghi, Jalal
    Culbertson, Christopher T.
    ENZYME ACTIVITY IN SINGLE CELLS, 2019, 628 : 223 - 241
  • [6] A Microfluidic System Enabling Continuous Characterization of Single-Cell Specific Membrane Capacitance and Cytoplasm Conductivity
    Zhao, Yang
    Chen, Deyong
    Li, Hao
    Luo, Yana
    Deng, Bin
    Huang, Song-Bin
    Chiu, Tzu-Keng
    Wu, Min-Hsien
    Long, Rong
    Wang, Junbo
    Chen, Jian
    2013 8TH ANNUAL IEEE INTERNATIONAL CONFERENCE ON NANO/MICRO ENGINEERED AND MOLECULAR SYSTEMS (IEEE NEMS 2013), 2013, : 344 - 347
  • [7] Microfluidic iterative mechanical characteristics (iMECH) analyzer for single-cell metastatic identification
    Babahosseini, Hesam
    Strobl, Jeannine S.
    Agah, Masoud
    ANALYTICAL METHODS, 2017, 9 (05) : 847 - 855
  • [8] Profiling lymphocyte interactions at the single-cell level by microfluidic cell pairing
    Burak Dura
    Stephanie K. Dougan
    Marta Barisa
    Melanie M. Hoehl
    Catherine T. Lo
    Hidde L. Ploegh
    Joel Voldman
    Nature Communications, 6
  • [9] Profiling lymphocyte interactions at the single-cell level by microfluidic cell pairing
    Dura, Burak
    Dougan, Stephanie K.
    Barisa, Marta
    Hoehl, Melanie M.
    Lo, Catherine T.
    Ploegh, Hidde L.
    Voldman, Joel
    NATURE COMMUNICATIONS, 2015, 6
  • [10] A microfluidic platform enabling single-cell RNA-seq of multigenerational lineages
    Kimmerling, Robert J.
    Szeto, Gregory Lee
    Li, Jennifer W.
    Genshaft, Alex S.
    Kazer, Samuel W.
    Payer, Kristofor R.
    Borrajo, Jacob de Riba
    Blainey, Paul C.
    Irvine, Darrell J.
    Shalek, Alex K.
    Manalis, Scott R.
    NATURE COMMUNICATIONS, 2016, 7