Demonstration of Saw Blade Accuracy and Excursion A Cadaveric Comparison Study of Blade Types Used in Total Knee Arthroplasty

被引:10
作者
Wetzel, Robert J. [1 ]
Shah, Ritesh R. [2 ]
Puri, Lalit [1 ]
机构
[1] Northwestern Univ, Feinberg Sch Med, Dept Orthopaed Surg, Chicago, IL 60611 USA
[2] LLC, Illinois Bone & Joint Inst, Morton Grove, IL USA
关键词
total knee arthroplasty; saw blade; accuracy; excursion; tip oscillating blade; NAVIGATION; ALIGNMENT; AXIS;
D O I
10.1016/j.arth.2013.02.006
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
In total knee arthroplasty, outcomes partly depend on accurate osteotomies and integrity of stabilizing structures. We compared accuracy and excursion between a conventional and an oscillating tip saw blade. Two sets of osteotomies were made on cadaveric knees. Bi-planar accuracy was compared using computer navigation, and excursion was compared using methylene blue. Wilcoxon-Mann-Whitney testing demonstrated no significant difference in blade accuracy (p=0.35). Blades were within 0.5 degrees of neutral coronally and 2.0 degrees sagittally. The oscillating tip blade demonstrated less dye markings on the surrounding tissues. Accurate osteotomies and soft tissue protection are critical to successful arthroplasties. Although comparative accuracy was equal, the oscillating tip blade exhibited less excursion displaying potential for less iatrogenic soft tissue injuries leading to catastrophic failure. (C) 2013 Elsevier Inc. All rights reserved.
引用
收藏
页码:985 / 987
页数:3
相关论文
共 11 条
  • [1] Computer Assisted Navigation in Knee Arthroplasty
    Bae, Dae Kyung
    Song, Sang Jun
    [J]. CLINICS IN ORTHOPEDIC SURGERY, 2011, 3 (04) : 259 - 267
  • [2] Randomized control trial comparing radiographic total knee arthroplasty implant placement using computer navigation versus conventional technique
    Chin, PL
    Yang, KY
    Yeo, SJ
    Lo, NN
    [J]. JOURNAL OF ARTHROPLASTY, 2005, 20 (05) : 618 - 626
  • [3] Leg axis after computer-navigated total knee arthroplasty - A prospective randomized trial comparing computer-navigated and manual implantation
    Decking, R
    Markmann, Y
    Fuchs, J
    Puhl, W
    Scharf, HP
    [J]. JOURNAL OF ARTHROPLASTY, 2005, 20 (03) : 282 - 288
  • [4] Excursion of Oscillating Saw Blades in Total Knee Arthroplasty
    Dimitris, Kirk
    Taylor, Benjamin C.
    Steensen, Robert N.
    [J]. JOURNAL OF ARTHROPLASTY, 2010, 25 (01) : 158 - 160
  • [5] Coronal Alignment in Total Knee Arthroplasty Just How Important is it?
    Fang, David M.
    Ritter, Merrill A.
    Davis, Kenneth E.
    [J]. JOURNAL OF ARTHROPLASTY, 2009, 24 (06) : 39 - 43
  • [6] Primary repair of intraoperative disruption of the medial collateral ligament during total knee arthroplasty
    Leopold, SS
    McStay, C
    Klafeta, K
    Jacobs, JJ
    Berger, RA
    Rosenberg, AG
    [J]. JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2001, 83A (01) : 86 - 91
  • [7] MORELAND JR, 1988, CLIN ORTHOP RELAT R, P49
  • [8] The Value of Patient-Matched Instrumentation in Total Knee Arthroplasty
    Noble, John W., Jr.
    Moore, Chris A.
    Liu, Ning
    [J]. JOURNAL OF ARTHROPLASTY, 2012, 27 (01) : 153 - 155
  • [9] Effect of Postoperative Mechanical Axis Alignment on the Fifteen-Year Survival of Modern, Cemented Total Knee Replacements
    Parratte, Sebastien
    Pagnano, Mark W.
    Trousdale, Robert T.
    Berry, Daniel J.
    [J]. JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2010, 92A (12) : 2143 - 2149
  • [10] Extensor Mechanism Allograft Reconstruction After Total Knee Arthroplasty
    Springer, Bryan D.