Evaluation of measurement sites for noninvasive blood glucose sensing with near-infrared transmission spectroscopy

被引:1
|
作者
Burmeister, JJ
Arnold, MA [1 ]
机构
[1] Univ Iowa, Dept Chem, Iowa City, IA 52242 USA
[2] Univ Iowa, Opt Sci & Technol Ctr, Iowa City, IA 52242 USA
关键词
D O I
暂无
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Six putative measurement sites were evaluated for noninvasive sensing of blood glucose by first-overtone near-infrared spectroscopy. The cheek, lower lip, upper lip, nasal septum, tongue, and webbing tissue between the thumb and forefinger were examined. These sites were evaluated on the basis of their chemical and physical properties as they pertain to the noninvasive measurement of glucose. Critical features included the effective optical pathlength of aqueous material within the tissue and the percentage of body fat within the optical path. Aqueous optical paths of 5 mm are required to measure clinically relevant concentrations of glucose in the first-overtone region. All of the tested sites met this requirement. The percentage of body fat affects the signal-to-noise ratio of the measurement and must be minimized for reliable glucose sensing. The webbing tissue contains a considerable amount of fat tissue and is clearly the worse;measurement site. All other sites possess substantially less fat, with the least amount of fat in tongue tissue. For this reason, the tongue provides spectra with the highest signal-to-noise ratio and is, therefore, the site of choice on the basis of spectral quality. (C) 1999 American Association for Clinical Chemistry.
引用
收藏
页码:1621 / 1627
页数:7
相关论文
共 50 条
  • [1] Phantoms for noninvasive blood glucose sensing with near infrared transmission spectroscopy
    Burmeister, JJ
    Chung, HI
    Arnold, MA
    PHOTOCHEMISTRY AND PHOTOBIOLOGY, 1998, 67 (01) : 50 - 55
  • [2] New advancement in noninvasive measurement of blood glucose by near-infrared spectroscopy
    Li, QB
    Wang, Y
    Xu, KX
    OPTICS IN HEALTH CARE AND BIOMEDICAL OPTICS: DIAGNOSTICS AND TREATMENT, 2002, 4916 : 457 - 464
  • [3] Noninvasive blood glucose sensing with near-infrared spectroscopy based on interstitial fluid
    Lu, Q. (luqipeng@126.com), 1600, Chinese Optical Society (33):
  • [4] Near-Infrared Spectroscopy for Noninvasive Measurement of Blood Glucose: Problems, Progress, Tasks
    K. V. Pozhar
    M. O. Mikhailov
    E. L. Litinskaia
    E. A. Polyakova
    Biomedical Engineering, 2022, 56 : 64 - 68
  • [5] Near-Infrared Spectroscopy for Noninvasive Measurement of Blood Glucose: Problems, Progress, Tasks
    Pozhar, K., V
    Mikhailov, M. O.
    Litinskaia, E. L.
    Polyakova, E. A.
    BIOMEDICAL ENGINEERING-MEDITSINSKAYA TEKNIKA, 2022, 56 (01): : 64 - 68
  • [6] Accuracy of Noninvasive Glucose Sensing Based on Near-Infrared Spectroscopy
    Liu, Jin
    Liu, Rong
    Xu, Kexin
    APPLIED SPECTROSCOPY, 2015, 69 (11) : 1313 - 1318
  • [7] On Feasibility of Near-Infrared Spectroscopy for Noninvasive Blood Glucose Measurements
    Vural, N. Mert
    Yoleri, Yigit
    Torun, Hamdi
    OPTICAL DIAGNOSTICS AND SENSING XIX: TOWARD POINT-OF-CARE DIAGNOSTICS, 2019, 10885
  • [8] In vivo noninvasive measurement of blood glucose by near-infrared diffuse-reflectance spectroscopy
    Maruo, K
    Tsurugi, M
    Tamura, M
    Ozaki, Y
    APPLIED SPECTROSCOPY, 2003, 57 (10) : 1236 - 1244
  • [9] Internal floating reference method for noninvasive measurement of blood glucose with near-infrared spectroscopy
    Luo, Yun-Han
    Li, Shiping
    Huang, Furong
    OPTOELECTRONICS AND ADVANCED MATERIALS-RAPID COMMUNICATIONS, 2007, 1 (06): : 307 - 311
  • [10] Internal floating reference method for noninvasive measurement of blood glucose with near-infrared spectroscopy
    Luo, Yun-Han
    Li, Ship-Ing
    Huang, Fu-Rong
    Chen, Zhe
    JOURNAL OF OPTOELECTRONICS AND ADVANCED MATERIALS, 2008, 10 (03): : 592 - 596