Use of measured scatter data for the attenuation correction of single photon emission tomography without transmission scanning

被引:21
作者
Cade, Sarah C. [1 ,2 ]
Arridge, Simon [3 ]
Evans, Martyn J. [2 ]
Hutton, Brian F. [1 ,4 ]
机构
[1] Univ Coll London Hosp, Inst Nucl Med, London NW1 2BU, England
[2] Royal United Hosp NHS Trust, Bath BA1 3NG, Avon, England
[3] UCL, Ctr Med Image Comp, London WC1E 6BT, England
[4] Univ Wollongong, Ctr Med Radiat Phys, Wollongong, NSW 2522, Australia
基金
英国工程与自然科学研究理事会; 美国国家卫生研究院;
关键词
SPECT; attenuation correction; scatter; SPECT; CT; RECONSTRUCTION; ARTIFACTS; SIMULATION; PET/CT;
D O I
10.1118/1.4812686
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: Attenuation correction is essential for reliable interpretation of emission tomography, however, the use of transmission measurements to generate attenuation maps is limited by availability of equipment and potential mismatches between the transmission and emission measurements. The authors present a first step toward a method of estimating an attenuation map from measured scatter data without a transmission scan. Methods: A scatter model has been developed that accurately predicts the distribution of photons which have been scattered once. The scatter model has been used as the basis of a maximum likelihood gradient ascent method to estimate an attenuation map from measured scatter data. In order to estimate both the attenuation map and activity distribution, iterations of the derived scatter based algorithm have been alternated with the maximum likelihood expectation maximization algorithm in a joint estimation process. For each iteration of the attenuation map estimation, the activity distribution is fixed at the values estimated during the previous activity iteration, and in each iteration of the activity distribution estimation the attenuation map is fixed at the values estimated during the previous attenuation iteration. The use of photopeak data to enhance the estimation of the attenuation map compared to the use of scatter data alone has also been considered. The algorithm derived has been used to reconstruct data simulated for an idealized two-dimensional situation and using a physical phantom. Results: The reconstruction of idealized data demonstrated good reconstruction of both the activity distribution and attenuation map. The inclusion of information recorded in the photopeak energy window in the attenuation map estimation step demonstrated an improvement in the accuracy of the reconstruction, enabling an accurate attenuation map to be recovered. Validation of the results with physical phantom data demonstrated that different regions of attenuation could be distinguished in a real situation and produces results that represent a promising first step toward the use of scatter data to estimate the activity distribution and attenuation map from single photon emission tomography (SPECT) data without a transmission scan. Conclusions: The technique presented shows promise as a method of attenuation correction for SPECT data without the need for a separate transmission scan. Further work is required to further improve the method and to compensate for the assumptions used in the scatter model. (C) 2013 American Association of Physicists in Medicine.
引用
收藏
页数:12
相关论文
共 50 条
  • [1] Attenuation map estimation without transmission scanning using measured scatter data
    Cade, Sarah C.
    Arridge, Simon
    Evans, Martyn J.
    Hutton, Brian F.
    2011 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2011, : 2657 - 2663
  • [2] ArtificialIntelligence-Based Data Corrections for Attenuation and Scatter in Position Emission Tomography and Single-Photon Emission Computed Tomography
    McMillan, Alan B.
    Bradshaw, Tyler J.
    PET CLINICS, 2021, 16 (04) : 543 - 552
  • [3] Investigation of Attenuation Correction for Small-Animal Single Photon Emission Computed Tomography
    Lee, Hsin-Hui
    Chen, Jyh-Cheng
    COMPUTATIONAL AND MATHEMATICAL METHODS IN MEDICINE, 2013, 2013
  • [4] Diagnostic performance of myocardial perfusion single-photon emission computed tomography with attenuation correction
    Plachcinska, Anna
    Wlodarczyk, Michal
    Kovacevic-Kusmierek, Katarzyna
    Bienkiewicz, Malgorzata
    Drozdz, Jaroslaw
    Chizynski, Krzysztof
    Kasprzak, Jaroslaw D.
    Peruga, Jan Z.
    Kusmierek, Jacek
    KARDIOLOGIA POLSKA, 2016, 74 (01) : 32 - 39
  • [5] Attenuation correction in emission tomography using the emission data-A review
    Berker, Yannick
    Li, Yusheng
    MEDICAL PHYSICS, 2016, 43 (02) : 807 - 832
  • [6] ATTENUATION CORRECTION IN CARDIAC POSITRON EMISSION TOMOGRAPHY AND SINGLE-PHOTON EMISSION COMPUTED-TOMOGRAPHY
    BACHARACH, SL
    BUVAT, I
    JOURNAL OF NUCLEAR CARDIOLOGY, 1995, 2 (03) : 246 - 255
  • [7] Attenuation correction for single photon emission computed tomography myocardial perfusion imaging
    Corbett J.R.
    Kritzman J.N.
    Ficaro E.P.
    Current Cardiology Reports, 2004, 6 (1) : 32 - 40
  • [8] Quantitative evaluation of single-photon emission computed tomography findings in lower extremity possible without computed tomography-based attenuation correction
    Tsuchitani, Tatsuya
    Kitajima, Kazuhiro
    Takahashi, Yoshiyuki
    Kotoura, Noriko
    NUCLEAR MEDICINE COMMUNICATIONS, 2021, 42 (10) : 1097 - 1103
  • [9] Magnetic Resonace-Based Attenuation Correction for Micro-Single-Photon Emission Computed Tomography
    Keereman, Vincent
    Fierens, Yves
    Vanhove, Christian
    Lahoutte, Tony
    Vandenberghe, Stefaan
    MOLECULAR IMAGING, 2012, 11 (02): : 155 - 165
  • [10] A novel phantom design for emission tomography enabling scatter- and attenuation-"free" single-photon emission tomography imaging
    Larsson, SA
    Jonsson, C
    Pagani, M
    Johansson, L
    Jacobsson, H
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE, 2000, 27 (02) : 131 - 139