Acute thrombus formation on phosphorilcholine surface modified flow diverters

被引:53
作者
Marosfoi, Miklos [1 ]
Clarencon, Frederic [2 ]
Langan, Erin T. [1 ]
King, Robert M. [1 ]
Brooks, Olivia W. [1 ]
Tamura, Takamisu [1 ]
Wainwright, John M. [3 ]
Gounis, Matthew J. [1 ]
Vedantham, Srinivasan [4 ]
Puri, Ajit S. [1 ]
机构
[1] Univ Massachusetts, Sch Med, Dept Radiol, New England Ctr Stroke Res, Worcester, MA USA
[2] Hop La Pitie Salpetriere, Dept Intervent Neuroradiol, Paris, France
[3] Medtron Neurovasc, Res & Dev, Irvine, CA USA
[4] Univ Arizona, Banner Univ Med Ctr, Dept Med Imaging, Tucson, AZ USA
关键词
optical coherence tomography; flow diverters; intracranial aneurysms; dual antiplatelet therapy; PIPELINE EMBOLIZATION DEVICE; OPTICAL COHERENCE TOMOGRAPHY; INTRACRANIAL ANEURYSMS; RABBIT MODEL; CLOPIDOGREL; COMPLICATIONS; ARTERY; APPOSITION; CONJUGATE; THERAPY;
D O I
10.1136/neurintsurg-2017-013175
中图分类号
R445 [影像诊断学];
学科分类号
100207 ;
摘要
Purpose Thromboembolic complications remain a limitation of flow diverting stents. We hypothesize that phosphorilcholine surface modified flow diverters (Pipeline Flex with Shield Technology, sPED) would have less acute thrombus formation on the device surface compared with the classic Pipeline Embolization device (cPED). Methods Elastase-induced aneurysms were created in 40 rabbits and randomly assigned to receive cPED or sPED devices with and without dual antiplatelet therapy (DAPT) (four groups, n=10/group). Angioplasty was performed to enhance apposition and create intimal injury for a pro-thrombotic environment. Both before and after angioplasty, the flow diverter was imaged with intravascular optical coherence tomography. The outcome measure was the number of predefined segments along the implant relative to the location of the aneurysm with a minimum of 0 (no clot formation) and maximum of 3 (all segments with thrombus). Clot formation over the device at ostia of branch arteries was assessed as either present or absent. Results Following angioplasty, the number of flow diverter segments with clots was significantly associated with the flow diverter (p<0.0001), but not with DAPT (p=0.3872) or aneurysm neck size (p=0.8555). The incidence rate for clots with cPED was 1.72 times more than with sPED. The clots on the flow diverter at the location corresponding to side branch ostia was significantly lower with sPED than with cPED (OR 0.180; 95%CI 0.044 to 0.734; p=0.0168), but was not associated with DAPT (p=0.3198). Conclusion In the rabbit model, phosphorilcholine surface modified flow diverters are associated with less thrombus formation on the surface of the device.
引用
收藏
页码:406 / 410
页数:5
相关论文
共 28 条
[11]   Flow Diverters Can Occlude Aneurysms and Preserve Arterial Branches: A New Experimental Model [J].
Darsaut, T. E. ;
Bing, F. ;
Salazkin, I. ;
Gevry, G. ;
Raymond, J. .
AMERICAN JOURNAL OF NEURORADIOLOGY, 2012, 33 (10) :2004-2009
[12]   Thrombogenicity of flow diverters in an ex vivo shunt model: effect of phosphorylcholine surface modification [J].
Hagen, Matthew W. ;
Girdhar, Gaurav ;
Wainwright, John ;
Hinds, Monica T. .
JOURNAL OF NEUROINTERVENTIONAL SURGERY, 2017, 9 (10) :1006-1011
[13]  
Herbert JM, 1998, THROMB HAEMOSTASIS, V80, P512
[14]   Safety and efficacy of the Pipeline embolization device for treatment of intracranial aneurysms: a pooled analysis of 3 large studies [J].
Kallmes, David F. ;
Brinjikji, Waleed ;
Cekirge, Saruhan ;
Fiorella, David ;
Hanel, Ricardo A. ;
Jabbour, Pascal ;
Lopes, Demetrius ;
Lylyk, Pedro ;
McDougall, Cameron G. ;
Siddiqui, Adnan .
JOURNAL OF NEUROSURGERY, 2017, 127 (04) :775-780
[15]   Use of self-expanding stents for better intracranial flow diverter wall apposition [J].
Kuhn, Anna Luisa ;
Wakhloo, Ajay K. ;
Gounis, Matthew J. ;
Kan, Peter ;
Rodrigues, Katyucia de Macedo ;
Lozano, J. Diego ;
Marosfoi, Miklos G. ;
Perras, Mary ;
Brooks, Christopher ;
Howk, Mary C. ;
Rex, David E. ;
Massari, Francesco ;
Puri, Ajit S. .
INTERVENTIONAL NEURORADIOLOGY, 2017, 23 (02) :129-136
[16]   Assessment of coronary arterial thrombus by optical coherence tomography [J].
Kume, Teruyoshi ;
Akasaka, Takashi ;
Kawamoto, Takahiro ;
Ogasawara, Yasuo ;
Watanabe, Nozorni ;
Toyota, Eiji ;
Neishi, Yoji ;
Sukmawan, Renan ;
Sadahira, Yoshito ;
Yoshida, Kiyoshi .
AMERICAN JOURNAL OF CARDIOLOGY, 2006, 97 (12) :1713-1717
[17]   Intra-DIC (distal intracranial catheter) deployment of the Pipeline embolization device: a novel rescue strategy for failed device expansion [J].
Lin, Li-Mei ;
Colby, Geoffrey P. ;
Jiang, Bowen ;
Nundkumar, Neelesh ;
Huang, Judy ;
Tamargo, Rafael J. ;
Coon, Alexander L. .
JOURNAL OF NEUROINTERVENTIONAL SURGERY, 2016, 8 (08) :840-846
[18]   Evaluation of cerebral artery perforators and the pipeline embolization device using optical coherence tomography [J].
Lopes, Demetrius Klee ;
Johnson, Andrew Kelly .
JOURNAL OF NEUROINTERVENTIONAL SURGERY, 2012, 4 (04) :291-294
[19]   Antiplatelet Loading Improves Behavioral Outcome in a Rabbit Model of Stroke [J].
Meyer, Dawn M. ;
Compton, Peggy ;
Eastwood, Jo-Ann ;
Gylys, Karen ;
Zivin, Justin A. .
STROKE, 2013, 44 (11) :3246-3248
[20]  
Miskolczi L, 2005, AM J NEURORADIOL, V26, P1964