Dosimetric effects of intrafractional isocenter variation during deep inspiration breath-hold for breast cancer patients using surface-guided radiotherapy

被引:43
|
作者
Kugele, Malin [1 ,2 ]
Edvardsson, Anneli [2 ]
Berg, Lovisa [1 ]
Alkner, Sara [1 ]
Ljus, Carina Andersson [1 ]
Ceberg, Sofie [1 ,2 ]
机构
[1] Skane Univ Hosp, Dept Hematol Oncol & Radiat Phys, Lund, Sweden
[2] Lund Univ, Dept Clin Sci, Med Radiat Phys, Lund, Sweden
来源
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS | 2018年 / 19卷 / 01期
关键词
breath hold; intrafractional isocenter variation; optical surface scanning; treatment planning; RADIATION-THERAPY; HEART-DISEASE; RISK; WOMEN; REPRODUCIBILITY; MORTALITY; COHORT;
D O I
10.1002/acm2.12214
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The aim of this study was to investigate potential dose reductions to the heart, left anterior descending coronary artery (LAD), and ipsilateral lung for left-sided breast cancer using visually guided deep inspiration breath-hold (DIBH) with the optical surface scanning system Catalyst (TM), and how these potential dosimetric benefits are affected by intrafractional motion in between breath holds. For both DIBH and free breathing (FB), treatment plans were created for 20 tangential and 20 locoregional left-sided breast cancer patients. During DIBH treatment, beam-on was triggered by a region of interest on the xiphoid process using a 3 mm gating window. Using a novel nonrigid algorithm, the Catalyst (TM) system allows for simultaneous real-time tracking of the isocenter position, which was used to calculate the intrafractional DIBH isocenter reproducibility. The 50% and 90% cumulative probabilities and maximum values of the intrafractional DIBH isocenter reproducibility were calculated and to obtain the dosimetric effect isocenter shifts corresponding to these values were performed in the treatment planning system. For both tangential and locoregional treatment, the dose to the heart, LAD and ipsilateral lung was significantly reduced for DIBH compared to FB. The intrafractional DIBH isocenter reproducibility was very good for the majority of the treatment sessions, with median values of approximately 1 mm in all three translational directions. However, for a few treatment sessions, intrafractional DIBH isocenter reproducibility of up to 5 mm was observed, which resulted in large dosimetric effects on the target volume and organs at risk. Hence, it is of importance to set tolerance levels on the intrafractional isocenter motion and not only perform DIBH based on the xiphoid process.
引用
收藏
页码:25 / 38
页数:14
相关论文
共 50 条
  • [31] Deep Inspiration Breath Hold in Left-Sided Breast Radiotherapy A Balance Between Side Effects and Costs
    Busschaert, Sara-Lise
    Kimpe, Eva
    Gevaert, Thierry
    De Ridder, Mark
    Putman, Koen
    JACC: CARDIOONCOLOGY, 2024, 6 (04): : 514 - 525
  • [32] Voluntary deep inspiration breath-hold reduces the heart dose without compromising the target volume coverage during radiotherapy for left-sided breast cancer
    Al-Hammadi, Noora
    Caparrotti, Palmira
    Naim, Carole
    Hayes, Jillian
    Benson, Katherine Rebecca
    Vasic, Ana
    Al-Abdulla, Hissa
    Hammoud, Rabih
    Divakar, Saju
    Petric, Primoz
    RADIOLOGY AND ONCOLOGY, 2018, 52 (01) : 112 - 120
  • [33] Artificial Intelligence-Based Patient Selection for Deep Inspiration Breath-Hold Breast Radiotherapy from Respiratory Signals
    Vendrame, Alessandra
    Cappelletto, Cristina
    Chiovati, Paola
    Vinante, Lorenzo
    Parvej, Masud
    Caroli, Angela
    Pirrone, Giovanni
    Barresi, Loredana
    Drigo, Annalisa
    Avanzo, Michele
    APPLIED SCIENCES-BASEL, 2023, 13 (08):
  • [34] A uniform and versatile surface-guided radiotherapy procedure and workflow for high-quality breast deep-inspiration breath-hold treatment in a multi-center institution
    Li, Guang
    Lu, Wei
    O'Grady, Kyle
    Yan, Iris
    Yorke, Ellen
    Arriba, Laura I. Cervino
    Powell, Simon
    Hong, Linda
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2022, 23 (03):
  • [35] Dosimetric evaluation of critical organs at risk in mastectomized left-sided breast cancer radiotherapy using breath-hold technique
    Dincoglan, Ferrat
    Beyzadeoglu, Murat
    Sager, Omer
    Oysul, Kaan
    Kahya, Yelda Elcim
    Gamsiz, Hakan
    Uysal, Bora
    Demiral, Selcuk
    Dirican, Bahar
    Surenkok, Serdar
    TUMORI JOURNAL, 2013, 99 (01): : 76 - 82
  • [36] Reproducibility and stability of spirometer-guided deep inspiration breath-hold in left-breast treatments using an optical surface monitoring system
    Cilla, Savino
    Romano, Carmela
    Craus, Maurizio
    Viola, Pietro
    Macchia, Gabriella
    Boccardi, Mariangela
    De Vivo, Livia P.
    Buwenge, Milly
    Morganti, Alessio G.
    Deodato, Francesco
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2023, 24 (06):
  • [37] The necessity of using deep inspiration breath-hold in the radiotherapy of left breast cancer patients who undergo the UK FAST trial
    Hafez, Abdelrahman
    Abdelaziz, Dina M.
    Khalil, Magdy M.
    El-Nagdy, Mohamed S.
    BIOMEDICAL PHYSICS & ENGINEERING EXPRESS, 2021, 7 (01)
  • [38] Dosimetric comparison of deep-inspiration breath-hold and free-breathing treatment delivery techniques for left-sided breast cancer using 3D surface tracking
    Vuong, Winston
    Garg, Rupen
    Bourgeois, Daniel J.
    Yu, Suhong
    Sehgal, Varun
    Daroui, Parima
    MEDICAL DOSIMETRY, 2019, 44 (03) : 193 - 198
  • [39] Review of deep inspiration breath-hold techniques for the treatment of breast cancer
    Latty, Drew
    Stuart, Kirsty E.
    Wang, Wei
    Ahern, Verity
    JOURNAL OF MEDICAL RADIATION SCIENCES, 2015, 62 (01) : 74 - 81
  • [40] Clinical benefits of deep inspiration breath-hold in postoperative radiotherapy for right-sided breast cancer: a meta-analysis
    Li, Zhuocheng
    Jian, Chenxi
    Li, Yuanyuan
    Pan, Zhenyu
    Yang, Guozi
    Sun, Xingru
    BMC CANCER, 2024, 24 (01)