Dosimetric effects of intrafractional isocenter variation during deep inspiration breath-hold for breast cancer patients using surface-guided radiotherapy

被引:45
作者
Kugele, Malin [1 ,2 ]
Edvardsson, Anneli [2 ]
Berg, Lovisa [1 ]
Alkner, Sara [1 ]
Ljus, Carina Andersson [1 ]
Ceberg, Sofie [1 ,2 ]
机构
[1] Skane Univ Hosp, Dept Hematol Oncol & Radiat Phys, Lund, Sweden
[2] Lund Univ, Dept Clin Sci, Med Radiat Phys, Lund, Sweden
关键词
breath hold; intrafractional isocenter variation; optical surface scanning; treatment planning; RADIATION-THERAPY; HEART-DISEASE; RISK; WOMEN; REPRODUCIBILITY; MORTALITY; COHORT;
D O I
10.1002/acm2.12214
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The aim of this study was to investigate potential dose reductions to the heart, left anterior descending coronary artery (LAD), and ipsilateral lung for left-sided breast cancer using visually guided deep inspiration breath-hold (DIBH) with the optical surface scanning system Catalyst (TM), and how these potential dosimetric benefits are affected by intrafractional motion in between breath holds. For both DIBH and free breathing (FB), treatment plans were created for 20 tangential and 20 locoregional left-sided breast cancer patients. During DIBH treatment, beam-on was triggered by a region of interest on the xiphoid process using a 3 mm gating window. Using a novel nonrigid algorithm, the Catalyst (TM) system allows for simultaneous real-time tracking of the isocenter position, which was used to calculate the intrafractional DIBH isocenter reproducibility. The 50% and 90% cumulative probabilities and maximum values of the intrafractional DIBH isocenter reproducibility were calculated and to obtain the dosimetric effect isocenter shifts corresponding to these values were performed in the treatment planning system. For both tangential and locoregional treatment, the dose to the heart, LAD and ipsilateral lung was significantly reduced for DIBH compared to FB. The intrafractional DIBH isocenter reproducibility was very good for the majority of the treatment sessions, with median values of approximately 1 mm in all three translational directions. However, for a few treatment sessions, intrafractional DIBH isocenter reproducibility of up to 5 mm was observed, which resulted in large dosimetric effects on the target volume and organs at risk. Hence, it is of importance to set tolerance levels on the intrafractional isocenter motion and not only perform DIBH based on the xiphoid process.
引用
收藏
页码:25 / 38
页数:14
相关论文
共 30 条
[1]  
Abe O, 2005, LANCET, V366, P2087, DOI 10.1016/s0140-6736(05)66544-0
[2]  
American Cancer Society, BREAST CANC FACTS FI
[3]  
Bergh L., BREATHING ADAPTED RA
[4]   Deep Inspiration Breath Hold-Based Radiation Therapy: A Clinical Review [J].
Boda-Heggemann, Judit ;
Knopf, Antje-Christin ;
Simeonova-Chergou, Anna ;
Wertz, Hansjoerg ;
Stieler, Florian ;
Jahnke, Anika ;
Jahnke, Lennart ;
Fleckenstein, Jens ;
Vogel, Lena ;
Arns, Anna ;
Blessing, Manuel ;
Wenz, Frederik ;
Lohr, Frank .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2016, 94 (03) :478-492
[5]   Cardiovascular Disease Risk in a Large, Population-Based Cohort of Breast Cancer Survivors [J].
Boekel, Naomi B. ;
Schaapveld, Michael ;
Gietema, Jourik A. ;
Russell, Nicola S. ;
Poortmans, Philip ;
Theuws, Jacqueline C. M. ;
Schinagl, Dominic A. X. ;
Rietveld, Derek H. F. ;
Versteegh, Michel I. M. ;
Visser, Otto ;
Rutgers, Emiel J. T. ;
Aleman, Berthe M. P. ;
van Leeuwen, Flora E. .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2016, 94 (05) :1061-1072
[6]   Modern Radiation Therapy and Cardiac Outcomes in Breast Cancer [J].
Boero, Isabel J. ;
Paravati, Anthony J. ;
Triplett, Daniel P. ;
Hwang, Lindsay ;
Matsuno, Rayna K. ;
Gillespie, Erin F. ;
Yashar, Catheryn M. ;
Moiseenko, Vitali ;
Einck, John P. ;
Mell, Loren K. ;
Parikh, Sahil A. ;
Murphy, James D. .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2016, 94 (04) :700-708
[7]   Using surface imaging and visual coaching to improve the reproducibility and stability of deep-inspiration breath hold for left-breast-cancer radiotherapy [J].
Cervino, Laura I. ;
Gupta, Sonia ;
Rose, Mary A. ;
Yashar, Catheryn ;
Jiang, Steve B. .
PHYSICS IN MEDICINE AND BIOLOGY, 2009, 54 (22) :6853-6865
[8]   Is There a Dose-Response Relationship for Heart Disease With Low-Dose Radiation Therapy? [J].
Chung, Eugene ;
Corbett, James R. ;
Moran, Jean M. ;
Griffith, Kent A. ;
Marsh, Robin B. ;
Feng, Mary ;
Jagsi, Reshma ;
Kessler, Marc L. ;
Ficaro, Edward C. ;
Pierce, Lori J. .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2013, 85 (04) :959-964
[9]   Reduced lung dose and improved inspiration level reproducibility in visually guided DIBH compared to audio coached EIG radiotherapy for breast cancer patients [J].
Damkjaer, Sidsel M. S. ;
Aznar, Marianne C. ;
Pedersen, Anders N. ;
Vogelius, Ivan R. ;
Bangsgaard, Jens Peter ;
Josipovic, Mirjana .
ACTA ONCOLOGICA, 2013, 52 (07) :1458-1463
[10]   Risk of Ischemic Heart Disease in Women after Radiotherapy for Breast Cancer [J].
Darby, Sarah C. ;
Ewertz, Marianne ;
McGale, Paul ;
Bennet, Anna M. ;
Blom-Goldman, Ulla ;
Bronnum, Dorthe ;
Correa, Candace ;
Cutter, David ;
Gagliardi, Giovanna ;
Gigante, Bruna ;
Jensen, Maj-Britt ;
Nisbet, Andrew ;
Peto, Richard ;
Rahimi, Kazem ;
Taylor, Carolyn ;
Hall, Per .
NEW ENGLAND JOURNAL OF MEDICINE, 2013, 368 (11) :987-998